Alexander Vladimirovich Ivanov1, Edward Danielevich Virus2, Ksenya Alexandrovna Nikiforova2, Nicolai Evgenevich Kushlinskii3, Boris Petrovich Luzyanin2, Marina Yurievna Maksimova4, Mikhail Aleksanrovich Piradov4, Aslan Amirkhanovich Kubatiev2.
Abstract
An approach that allows direct analysis of the ratio of S-adenosylmethionine (SAM) and S-adenosylhomocysteine (SAH) by using CE is presented. The analytes were extracted on phenylboronic acid phase and eluted with 100 mmol/L HCl. CE separation of the analytes took place in the transient isotachophoresis mode with addition of NaCl and meglumine to the samples. The sensitivity (S/N = 3) and quantification limit (S/N = 10) of the method were 0.07 and 0.2 μmol/L, respectively, using a silica capillary with 50 μm internal diameter and 30.5 cm total length. The BGE was 0.02 mol/L Tris with 1 mol/L HCOOH (pH 2.2), and the separation voltage was 15-17 kV. Accuracy of SAM and SAH analysis in urine was 96 and 105%, respectively; interday precision for the SAM/SAH ratio was within 6%. The theoretical plate number exceeded a million. Total analysis time was 8.5 min.Entities:
Keywords: Capillary electrophoresis; S-adenosylhomocysteine; S-adenosylmethionine; Urine