Literature DB >> 27443223

Fracture of a spline of the femoral stem in an ingrown modular hip implant.

David Clinton McNabb1,2, Jason M Jennings1, Douglas A Dennis1,3,4,5.   

Abstract

INTRODUCTION: Physiologic strain patterns of the femur are altered by the bending stiffness of the implant as well as the loading pattern of the implant itself. The Savish Range of Motion (S-ROM) femoral stem has been designed with flutes (to decrease distal material) and with a coronal slot (to decrease stiffness). MATERIALS: The purpose of this case report is to briefly discuss design characteristics of femoral stems and their relationship to thigh pain as well as to describe the only report in the literature of a fracture of the anterior spline of a S-ROM prosthesis at the coronal slot.
RESULTS: The femoral stem fractured at the coronal slot due to multiple bending moments in a small diameter stem.
CONCLUSIONS: The proximal metaphyseal sleeve remains well fixed and the patient is functioning well without pain nearly a year after the identification of the fracture on routine follow-up.

Entities:  

Mesh:

Year:  2016        PMID: 27443223     DOI: 10.5301/hipint.5000399

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  2 in total

1.  Modular prosthesis fracture in a patient with developmental dysplasia of the hip: a case report and literature review.

Authors:  Yuan-Pei Cheng; Xiao-Kang Cheng; Yong-Bo Li; Qian-Ru Zhang; Hao Feng; Yi-Han Zhong; Yan-Bo Zhang; Han Wu
Journal:  BMC Musculoskelet Disord       Date:  2021-05-14       Impact factor: 2.362

2.  Fracture of a Sivash-Range of Motion Femoral Stem at the Diaphyseal Spline.

Authors:  Tyler J Humphrey; Daniel Marchwiany; Hany S Bedair; Christopher M Melnic
Journal:  Cureus       Date:  2021-12-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.