Accuracy of chest radiography for positioning of the umbilical venous catheter.

Adriana F M Guimarães1, Aline A C G de Souza2, Maria Cândida F Bouzada3, Zilda M A Meira4.   

Abstract

OBJECTIVES: To evaluate the accuracy of the simultaneous analysis of three radiographic anatomical landmarks - diaphragm, cardiac silhouette, and vertebral bodies - in determining the position of the umbilical venous catheter distal end using echocardiography as a reference standard.
METHODS: This was a cross-sectional, observational study, with the prospective inclusion of data from all neonates born in a public reference hospital, between April 2012 and September 2013, submitted to umbilical venous catheter insertion as part of their medical care. The position of the catheter distal end, determined by the simultaneous analysis of three radiographic anatomical landmarks, was compared with the anatomical position obtained by echocardiography; sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were calculated.
RESULTS: Of the 162 newborns assessed by echocardiography, only 44 (27.16%) had the catheter in optimal position, in the thoracic portion of the inferior vena cava or at the junction of the inferior vena cava with the right atrium. The catheters were located in the left atrium and interatrial septum in 54 (33.33%) newborns, in the right atrium in 26 (16.05%), intra-hepatic in 37 (22.84%), and intra-aortic in-one newborn (0.62%). The sensitivity, specificity and accuracy of the radiography to detect the catheter in the target area were 56%, 71%, and 67.28%, respectively.
CONCLUSION: Anteroposterior radiography of the chest alone is not able to safely define the umbilical venous catheter position. Echocardiography allows direct visualization of the catheter tip in relation to vascular structures and, whenever possible, should be considered to identify the location of the umbilical venous catheter.
Copyright © 2016 Sociedade Brasileira de Pediatria. Published by Elsevier Editora Ltda. All rights reserved.

Entities:  

Keywords:  Cateterismo venoso central; Central venous catheter; Chest radiography; Echocardiography; Ecocardiografia; Newborn; Radiografia torácica; Recém-nascido; Umbilical veins; Veias umbilicais

Mesh:

Year:  2016        PMID: 27424226     DOI: 10.1016/j.jped.2016.05.004

Source DB:  PubMed          Journal:  J Pediatr (Rio J)        ISSN: 0021-7557            Impact factor:   2.197


  2 in total

1.  Incidence of umbilical vein catheter-associated thrombosis of the portal system: A systematic review and meta-analysis.

Authors:  Iliana Bersani; Fiammetta Piersigilli; Giulia Iacona; Immacolata Savarese; Francesca Campi; Andrea Dotta; Cinzia Auriti; Enrico Di Stasio; Matteo Garcovich
Journal:  World J Hepatol       Date:  2021-11-27

2.  Effects of different concentrations of intraluminal sodium chloride solution on intracavitary ECG used for arm infusion port implantation.

Authors:  Lei Dong; Chen-Yang Guan; Ying Zhang; Ai-Xia Wang; Ming-Hua Liu; Chen Guo; Xiao-Li Hao; Qi Zhang
Journal:  Sci Rep       Date:  2022-08-15       Impact factor: 4.996

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.