| Literature DB >> 27423647 |
John Lemke1, Edmond Sardariani1, Joseph Bradley Phipps1, Niki Patel2, Loretta M Itri3, James Caravelli4, Eugene R Viscusi5.
Abstract
INTRODUCTION: Fentanyl iontophoretic transdermal system (fentanyl ITS, IONSYS(®)) is a patient-controlled analgesia system used for the management of acute postoperative pain, designed to be utilized in a hospital setting. The objective of the two studies was to determine if fentanyl ITS could be safely used with X-rays, computerized tomography (CT) scans and radiofrequency identification (RFID) devices.Entities:
Keywords: CT scan; Electromagnetic immunity; Fentanyl; IONSYS; Iontophoretic transdermal system; Patient-controlled analgesia; RFID; X-ray
Mesh:
Substances:
Year: 2016 PMID: 27423647 PMCID: PMC5020125 DOI: 10.1007/s12325-016-0381-y
Source DB: PubMed Journal: Adv Ther ISSN: 0741-238X Impact factor: 3.845
Description of each mode of operation of the fentanyl ITS
| Mode | LED | LCD | Audio | Current output |
|---|---|---|---|---|
| Ready Mode | Slow flashing green | Completed dose count | None | Off |
| Dose Mode | Fast flashing green | Completed dose count/progress indicator | Single long tone | Enabled |
| End-of-Use Mode | Off | Flashing completed dose count | None | Off |
| End-of-Life Mode | Flashing red | Completed dose count | Repeated short tones | Off |
ITS iontophoretic transdermal system, LCD liquid-crystal display, LED light-emitting diode
ITS placebo system X-radiation test results
| System # | System status during the exposure (ready) | Required effective dose (mSv) | Minimum current calculated during the exposure (µA) | Maximum current calculated during the exposure (µA) |
|---|---|---|---|---|
| 1 | Ready | 20 | 0 | 0 |
| 200 | 0 | 0 | ||
| 2 | Ready | 20 | 0 | 0 |
| 200 | 0 | 0 | ||
| 3 | Ready | 20 | 0 | 0 |
| 200 | 0 | 0 | ||
| 4 | Ready | 20 | 0 | 0 |
| 200 | 0 | 0 | ||
| 5 | Ready | 20 | 0 | 0 |
| 200 | 0 | 0 | ||
| 6 | Dose | 20 | 170.08 | 171.32 |
| 200 | 169.65 | 171.19 | ||
| 7 | Dose | 20 | 170.57 | 171.94 |
| 200 | 170.45 | 171.69 | ||
| 8 | Dose | 20 | 170.08 | 171.32 |
| 200 | 169.65 | 171.19 | ||
| 9 | Dose | 20 | 170.2 | 171.44 |
| 200 | 170.08 | 171.32 | ||
| 10 | Dose | 20 | 169.65 | 171.19 |
| 200 | 169.83 | 171.07 |
The exposure time to ensure 200 mSv was 17 min 15 s, the Dose Mode time was about 10 min, and at the end of the Dose Mode the system went back to the Ready Mode for the remaining test time and the voltage recorded was 0 V
ITS iontophoretic transdermal system
ITS Placebo system performance test results after X-radiation exposure
| System # | Ready Mode current (µA) | Entered into Dose Mode | Dose Mode current (µA) | Dose duration (min, s) |
|---|---|---|---|---|
| 1 | 0.000 | Pass | 169.70 | 9 min 59 s |
| 2 | 0.000 | Pass | 169.95 | 10 min 0 s |
| 3 | 0.000 | Pass | 170.32 | 10 min 0 s |
| 4 | 0.000 | Pass | 170.08 | 9 min 58 s |
| 5 | 0.000 | Pass | 170.94 | 9 min 59 s |
| 6 | 0.000 | Pass | 170.20 | 9 min 58 s |
| 7 | 0.000 | Pass | 170.82 | 9 min 58 s |
| 8 | 0.000 | Pass | 170.20 | 9 min 59 s |
| 9 | 0.000 | Pass | 170.20 | 9 min 59 s |
| 10 | 0.000 | Pass | 169.95 | 10 min 0 s |
ITS iontophoretic transdermal system
ITS Placebo system RFID test results
| System # | RF exposure condition | Minimum current calculated during the exposure (µA) | Maximum current calculated during the exposure (µA) |
|---|---|---|---|
| 11 | ISO 14223 134.2 kHz at 65 A/m | 169.78 | 172.43 |
| 12 | ISO 14223 134.2 kHz at 65 A/m | 170.62 | 173.28 |
| 13 | ISO 14223 134.2 kHz at 65 A/m | 169.74 | 172.42 |
| 14 | IEC 14443 Type A (13.56 MHz) 7.5 A/m | 170.38 | 172.29 |
| 15 | IEC 14443 Type A (13.56 MHz) 7.5 A/m | 170.91 | 173.66 |
| 16 | IEC 14443 Type A (13.56 MHz) 7.5 A/m | 170.37 | 172.99 |
| 17 | IEC 14443 Type B (13.56 MHz) 7.5 A/m | 169.87 | 172.55 |
| 18 | IEC 14443 Type B (13.56 MHz) 7.5 A/m | 170.01 | 172.72 |
| 19 | IEC 14443 Type B (13.56 MHz) 7.5 A/m | 170.45 | 173.21 |
| 20 | IEC 15693 (13.56 MHz) 5 A/m | 170.32 | 172.94 |
| 21 | IEC 15693 (13.56 MHz) 5 A/m | 170.29 | 172.99 |
| 22 | IEC 15693 (13.56 MHz) 5 A/m | 170.05 | 170.19 |
| 23 | IEC 18000-6 Type C (900 MHz) 54 V/m | 170.00 | 170.14 |
| 24 | IEC 18000-6 Type C (900 MHz) 54 V/m | 169.96 | 170.08 |
| 25 | IEC 18000-6 Type C (900 MHz) 54 V/m | 170.10 | 170.22 |
| 26 | IEC 18000-4 Mode 1 (2.45 GHz) 54 V/m | 170.42 | 172.92 |
| 27 | IEC 18000-4 Mode 1 (2.45 GHz) 54 V/m | 170.60 | 173.25 |
| 28 | IEC 18000-4 Mode 1 (2.45 GHz) 54 V/m | 170.10 | 172.26 |
ITS iontophoretic transdermal system, RFID radiofrequency identification, RF radiofrequency
ITS Placebo system performance test results after RFID exposure
| System # | Ready Mode current (µA) | Entered into Dose Mode | Dose Mode current (µA) | Dose duration (min, s) |
|---|---|---|---|---|
| 11 | 0 | Pass | 169.96 | 9 min 58 s |
| 12 | 0 | Pass | 170.65 | 9 min 59 s |
| 13 | 0 | Pass | 169.79 | 9 min 58 s |
| 14 | 0 | Pass | 170.29 | 9 min 59 s |
| 15 | 0 | Pass | 170.90 | 10 min 01 s |
| 16 | 0 | Pass | 170.54 | 10 min 01 s |
| 17 | 0 | Pass | 169.87 | 10 min 01 s |
| 18 | 0 | Pass | 170.00 | 9 min 57 s |
| 19 | 0 | Pass | 170.49 | 9 min 59 s |
| 20 | 0 | Pass | 170.40 | 9 min 58 s |
| 21 | 0 | Pass | 170.28 | 9 min 59 s |
| 22 | 0 | Pass | 170.16 | 9 min 59 s |
| 23 | 0 | Pass | 170.08 | 9 min 58 s |
| 24 | 0 | Pass | 169.95 | 9 min 58 s |
| 25 | 0 | Pass | 170.20 | 9 min 59 s |
| 26 | 0 | Pass | 170.60 | 9 min 59 s |
| 27 | 0 | Pass | 170.40 | 10 min 00 s |
| 28 | 0 | Pass | 170.16 | 10 min 00 s |
ITS iontophoretic transdermal system, RFID radiofrequency identification