Everolimus-eluting bioresorbable vascular scaffolds implanted in coronary bifurcation lesions: Impact of polymeric wide struts on side-branch impairment.

Marcella De Paolis1, Cordula Felix1, Nienke van Ditzhuijzen1, Jiang Ming Fam1, Antonis Karanasos1, Sanneke de Boer1, Nicolas M van Mieghem1, Joost Daemen1, Francesco Costa1, Luis Carlos Bergoli1, Jurgen M R Ligthart1, Evelyn Regar1, Peter P de Jaegere1, Felix Zijlstra1, Robert Jan van Geuns1, Roberto Diletti2.   

Abstract

BACKGROUND: Limited data are available on bioresorbable vascular scaffolds (BVS) performance in bifurcations lesions and on the impact of BVS wider struts on side-branch impairment.
METHODS: Patients with at least one coronary bifurcation lesion involving a side-branch ≥2mm in diameter and treated with at least one BVS were examined. Procedural and angiographic data were collected and a dedicated methodology for off-line quantitative coronary angiography (QCA) in bifurcation was applied (eleven-segment model), to assess side-branch impairment occurring any time during the procedure. Two- and three-dimensional QCA were used. Optical coherence tomography (OCT) analysis was performed in a subgroup of patients and long-term clinical outcomes reported.
RESULTS: A total of 102 patients with 107 lesions, were evaluated. Device- and procedural-successes were 99.1% and 94.3%, respectively. Side-branch impairment occurring any time during the procedure was reported in 13 bifurcations (12.1%) and at the end of the procedure in 6.5%. Side-branch minimal lumen diameter (Pre: 1.45±0.41mm vs Final: 1.48±0.42mm, p=0.587) %diameter-stenosis (Pre: 26.93±16.89% vs Final: 27.80±15.57%, p=0.904) and minimal lumen area (Pre: 1.97±0.89mm(2) vs Final: 2.17±1.09mm(2), p=0.334), were not significantly affected by BVS implantation. Mean malapposed struts at the bifurcation polygon-of-confluence were 0.63±1.11.
CONCLUSIONS: The results of the present investigation suggest feasibility and relative safety of BVS implantation in coronary bifurcations. BVS wide struts have a low impact on side-branch impairment when considering bifurcations with side-branch diameter≥2mm.
Copyright © 2016. Published by Elsevier Ireland Ltd.

Entities:  

Keywords:  Bifurcations; Bioresorbable vascular scaffold; Optical coherence tomography; Quantitative coronary angiography; Side-branch

Mesh:

Substances:

Year:  2016        PMID: 27423087     DOI: 10.1016/j.ijcard.2016.06.153

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  1 in total

1.  Efficacy and safety of bioresorbable scaffolds in patients with coronary bifurcation lesions: A systematic review and meta-analysis.

Authors:  Xi-Ying Liang; Yan Li; Wen-Jiao Zhang; Xuan Qiao; Rong-Rong Yang; Zhi-Lu Wang
Journal:  Cardiol J       Date:  2021-04-12       Impact factor: 3.487

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.