Zoltán Márta1, Balázs Bobály2, Jenő Fekete2, Balázs Magda3, Tímea Imre3, Katalin Viola Mészáros4, Pál Tamás Szabó5.
Abstract
Ultratrace analysis of sample components requires excellent analytical performance in terms of limits of quantitation (LoQ). Micro UHPLC coupling with sensitive tandem mass spectrometry provides state of the art solutions for such analytical problems. Decreased column volume in micro LC limits the injectable sample volume. However, if analyte concentration is extremely low, it might be necessary to inject high sample volumes. This is particularly critical for strong sample solvents and weakly retained analytes, which are often the case when preparing biological samples (protein precipitation, sample extraction, etc.). In that case, high injection volumes may cause band broadening, peak distortion or even elution in dead volume. In this study, we evaluated possibilities of high volume injection onto microbore RP-LC columns, when sample solvent is diluted. The presented micro RP-LC-MS/MS method was optimized for the analysis of steroid hormones from human plasma after protein precipitation with organic solvents. A proper sample dilution procedure helps to increase the injection volume without compromising peak shapes. Finally, due to increased injection volume, the limit of quantitation can be decreased by a factor of 2-5, depending on the analytes and the experimental conditions.Entities:
Keywords: High volume injection; LC–MS/MS; Microflow liquid chromatography; Steroid hormone
Mesh:
Substances:
Year: 2016 PMID: 27423010 DOI: 10.1016/j.jpba.2016.06.024