| Literature DB >> 27405364 |
Maria Nieddu1, Lucia Burrai2, Elena Baralla3, Valeria Pasciu3, Maria Vittoria Varoni3, Irene Briguglio1, Maria Piera Demontis3, Gianpiero Boatto1.
Abstract
Amphetamine designer drugs are central nervous system stimulants that are widely disseminated in the illegal market. Generally, in forensic laboratories, immunoassay methods are the first line of screening for these types of drugs in a biological specimen (typically blood, urine or oral fluid). In this article, we describe the cross-reactivity profiles of 30 new amphetamine designer drugs, using the Neogen(®) [Amphetamine Specific and Methamphetamine/3,4-Methylenedioxymethamphetamine (MDMA) assays] drug tests. To assess the potential matrix influence on the response, each assay was tested on whole blood, urine and oral fluid. Concentrations of 10,000 ng/mL were not sufficient to produce a positive response for the majority of the analyzed amphetamines. This clearly demonstrates that, although these kits are extremely effective for the target drugs for which they are intended (amphetamine, methamphetamine and MDMA), they cannot be used to reliably identify the tested designer drugs in real cases, as these concentrations greatly exceed those expected to be found in forensic samples.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27405364 DOI: 10.1093/jat/bkw060
Source DB: PubMed Journal: J Anal Toxicol ISSN: 0146-4760 Impact factor: 3.367