The Effects of Ambulatory Accelerations on the Stability of a Magnetically Suspended Impeller for an Implantable Blood Pump.

Gordon Paul1, Mohammed Amin Rezaienia1, Akbar Rahideh2, Ante Munjiza1, Theodosios Korakianitis3.   

Abstract

This article describes the effects of ambulatory accelerations on the stability of a magnetically suspended impeller for use in implantable blood pumps. A magnetic suspension system is developed to control the radial position of a magnetic impeller using coils in the pump casing. The magnitude and periodicity of ambulatory accelerations at the torso are measured. A test rig is then designed to apply appropriate accelerations to the suspension system. Accelerations from 0 to 1 g are applied to the suspended impeller with ambulatory periodicity while the radial position of the impeller and power consumption of the suspension system are monitored. The test is carried out with the impeller suspended in air, water, and a glycerol solution to simulate the viscosity of blood. A model is developed to investigate the effects of the radial magnetic suspension system and fluid damping during ambulatory accelerations. The suspension system reduces the average displacement of the impeller suspended in aqueous solutions within its casing to 100 µm with a power consumption of below 2 W during higher magnitude ambulatory accelerations (RMS magnitude 0.3 g). The damping effect of the fluid is also examined and it is shown that buoyancy, rather than drag, is the primary cause of the damping at the low displacement oscillations that occur during the application of ambulatory accelerations to such a suspension system.
© 2016 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  Ambulatory; Magnetic bearing; Third generation; Ventricular assist device

Mesh:

Year:  2016        PMID: 27401117     DOI: 10.1111/aor.12749

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

1.  Computational Parametric Study of the Axial and Radial Clearances in a Centrifugal Rotary Blood Pump.

Authors:  Mohammad Amin Rezaienia; Gordon Paul; Eldad Avital; Martin Rothman; Theodosios Korakianitis
Journal:  ASAIO J       Date:  2018 Sep/Oct       Impact factor: 2.872

2.  The airborne inertially stabilized platform suspend by an axial-radial integrated active magnetic actuator system.

Authors:  Tong Wen; Biao Xiang
Journal:  J Adv Res       Date:  2021-01-11       Impact factor: 10.479

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.