BACKGROUND: Historically it has been suggested that noise-induced hearing loss (NIHL) affects approximately 50% of the orthopedic surgery personnel. This noise may be partially caused by the use of powered saw systems that are used to make the bone cuts. The first goal was to quantify and compare the noise emission of these different saw systems during total knee arthroplasty (TKA) surgery. A second goal was to estimate the occupational NIHL risk for the orthopedic surgery personnel in TKA surgery by quantifying the total daily noise emission spectrum during TKA surgery and to compare this to the Dutch Occupational Health Organization guidelines.
METHODS: A conventional sagittal oscillating blade system with a full oscillating blade and 2 newer oscillating tip saw systems (handpiece and blade) were compared. Noise level measurements during TKA surgery were performed during cutting and hammering, additionally surgery noise profiles were made.
RESULTS: The noise level was significantly lower for the oscillating tip saw systems compared to the conventional saw system, but all were in a range that can cause NIHL. The conventional system handpiece produced a considerable higher noise level compared to oscillating tip handpiece.
CONCLUSION: NIHL is an underestimated problem in the orthopedic surgery. Solutions for decreasing the risk of hearing loss should be considered. The use of oscillating tip saw systems have a reduced noise emission in comparison with the conventional saw system. The use of these newer systems might be a first step in decreasing hearing loss among the orthopedic surgery personnel.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  TKA surgery; hearing loss; noise-induced hearing loss; orthopedic theater; saw blade

Mesh:

Year:  2016        PMID: 27394075     DOI: 10.1016/j.arth.2016.05.030

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  The risk of noise-induced hearing loss performing knee replacement surgery.

Authors:  Mark G Siegel
Journal:  Noise Health       Date:  2019 Sep-Oct       Impact factor: 0.867

2.  Primary total knee arthroplasty assisted by computed tomography-free navigation for secondary knee osteoarthritis following massive calcium phosphate cement packing for distal femoral giant-cell bone tumor treatment: a case report.

Authors:  Akihiko Takeuchi; Norio Yamamoto; Takaaki Ohmori; Katsuhiro Hayashi; Shinji Miwa; Kentaro Igarashi; Takashi Higuchi; Kensaku Abe; Hirotaka Yonezawa; Sei Morinaga; Yoshihiro Araki; Yohei Asano; Shiro Saito; Hiroyuki Tsuchiya
Journal:  BMC Musculoskelet Disord       Date:  2022-02-22       Impact factor: 2.362