The influence of different non-articular proximal forearm orthoses (brace) widths in the wrist extensors muscle activity, range of motion and grip strength in healthy volunteers.

Alexandre Márcio Marcolino1,2, Marisa de Cássia Registro Fonseca2, Naiara Tais Leonardi3, Rafael Inácio Barbosa1,2, Lais Mara Siqueira das Neves3, Rinaldo Roberto de Jesus Guirro2.   

Abstract

PURPOSE: The purpose this study was perform a biomechanical evaluation to compare the influence of commercial models of different non-articular proximal forearm orthoses widths (2.5 cm, 5.5 cm, 7.5 cm and 12.0 cm) in the extensor muscle activation, range of motion and grip strength in healthy subjects.
METHODS: Was analyzed data from extensor carpi radialis, extensor carpi ulnares and extensor digitorum comunis using surface electromyography, simultaneous with a wrist electrogoniometer MiotecTM and a hydraulic dynamometer JamarTM. The sequence of tests with all the commercial orthoses models was randomized. Statistics analyses were performed by linear model with mixed effects.
RESULTS: According to our findings the non-articular proximal forearm orthoses (2.5 cm - narrowest) positioned close to lateral epicondyle provided lesser muscle activation on extensor carpi radialis brevis/longus and extensor digitorum comunis, decreased wrist extension and grip strength during submaximal grip task (p< 0.01).
CONCLUSIONS: A narrow non-articular proximal forearm orthosis positioned close to the lateral epicondyle might decrease the extensor muscle activation and therefore could reduce mechanical stress on its insertion, based on this sample. Clinical studies must be conducted to confirm these findings.

Keywords:  Orthoses; brace; electromyography; lateral epicondylitis; wrist biomechanics

Year:  2016        PMID: 27392841     DOI: 10.3233/BMR-160727

Source DB:  PubMed          Journal:  J Back Musculoskelet Rehabil        ISSN: 1053-8127            Impact factor:   1.398


  1 in total

1.  Location of Reference Electrode Does Not Interfere on Electromyographic Parameters in the Domains of Time and Frequency.

Authors:  Rinaldo Roberto de Jesus Guirro; Alcimar Barbosa Soares; Amanda Caldeira Guirro; Camila Simieli; Alessandra Vairo Peres Boratino; Gabriela de Carvalho; Aline Gobbi; Elaine Caldeira de Oliveira Guirro
Journal:  J Med Syst       Date:  2018-08-11       Impact factor: 4.460