Literature DB >> 27390567

Biomechanical Analysis of Internal Fixation Methods for Distal Interphalangeal Joint Arthrodesis.

Stephanie K Rigot1, Rafael Diaz-Garcia2, Richard E Debski2, John Fowler2.   

Abstract

BACKGROUND: The biomechanical performance of distal interphalangeal joint (DIPJ) fixation techniques in response to cyclic and load-to-failure bending loads is generally unknown. The purpose of this study is to assess the performance of 4 commonly used techniques for DIP arthrodesis.
METHODS: Phalanges were fixed by one of the following techniques: (1) 2 parallel Kirschner wires (K-wires), (2) 1 longitudinal K-wire and 1 oblique K-wire, (3) an Acutrak® 2 Micro headless compression screw, or (4) 90-90 wiring. The phalanges were loaded in cantilever bending on the palmar side of the phalange for 3000 cycles, then loaded-to-failure.
RESULTS: The 90-90 wiring was significantly worse than the other 3 methods in 9 of 10 parameters analyzed. Both K-wire methods and the screw produced similar results in nearly all parameters.
CONCLUSIONS: Given the lack of difference in biomechanical performance between K-wires and compression screws, consideration should be given to other factors such as cost and complication profiles when choosing an implant for DIPJ fusion.

Entities:  

Keywords:  biomechanical; distal interphalangeal arthrodesis; headless compression screws; k-wires

Year:  2016        PMID: 27390567      PMCID: PMC4920526          DOI: 10.1177/1558944715627211

Source DB:  PubMed          Journal:  Hand (N Y)        ISSN: 1558-9447


  10 in total

1.  Identification of key pinch forces required to complete functional tasks.

Authors:  Niels Smaby; Mary Elise Johanson; Brian Baker; Deborah Ellen Kenney; Wendy Marie Murray; Vincent Rod Hentz
Journal:  J Rehabil Res Dev       Date:  2004-03

Review 2.  Distal interphalangeal joint arthrodesis.

Authors:  John R Fowler; Mark E Baratz
Journal:  J Hand Surg Am       Date:  2013-09-12       Impact factor: 2.230

3.  Distal interphalangeal joint arthrodesis comparing tension-band wire and Herbert screw: a biomechanical and dimensional analysis.

Authors:  B Wyrsch; J Dawson; S Aufranc; D Weikert; M Milek
Journal:  J Hand Surg Am       Date:  1996-05       Impact factor: 2.230

4.  Biomechanical analysis of internal fixation techniques for proximal interphalangeal joint arthrodesis.

Authors:  J C Kovach; F W Werner; A K Palmer; S Greenkey; D J Murphy
Journal:  J Hand Surg Am       Date:  1986-07       Impact factor: 2.230

5.  A compression arthrodesis device for small joints of the hands.

Authors:  J W Tupper
Journal:  Hand       Date:  1972-02

Review 6.  Small joint arthrodesis.

Authors:  A T Bishop
Journal:  Hand Clin       Date:  1993-11       Impact factor: 1.907

7.  Distal interphalangeal joint arthrodesis in extension using a headless compressive screw.

Authors:  Sujith Konan; Aditi Das; Emma Taylor; Elliot Sorene
Journal:  Acta Orthop Belg       Date:  2013-04       Impact factor: 0.500

8.  Use of a headless compressive screw for distal interphalangeal joint arthrodesis in digits: clinical outcome and review of complications.

Authors:  Jean-Paul Brutus; Andrew K Palmer; John F Mosher; Brian J Harley; Jon B Loftus
Journal:  J Hand Surg Am       Date:  2006-01       Impact factor: 2.230

Review 9.  Instructional Course Lecture. Arthrodesis of the interphalangeal joints with headless compression screws.

Authors:  Stephen J Leibovic
Journal:  J Hand Surg Am       Date:  2007-09       Impact factor: 2.230

10.  Distal interphalangeal joint arthrodesis: an analysis of complications.

Authors:  P J Stern; D B Fulton
Journal:  J Hand Surg Am       Date:  1992-11       Impact factor: 2.230

  10 in total
  2 in total

1.  Distal interphalangeal joint arthrodesis with nonaxial multiple small screws: a biomechanical analysis with axial headless compression screw and clinical result of 15 consecutive cases.

Authors:  Seung Hun Woo; Sang Ho Kwak; Hyo Seok Jang; Dong Hee Kim; Jang Hyeon Seo; Sang Hyun Lee
Journal:  BMC Musculoskelet Disord       Date:  2022-05-27       Impact factor: 2.562

2.  Comparison of Screw Quantity and Placement of Metacarpal Fracture Fixation: A Biomechanical Study.

Authors:  Stephen P Canton; Srujan Dadi; Austin Anthony; Ryan T Black; Michael Clancy; John R Fowler
Journal:  Hand (N Y)       Date:  2020-12-21
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.