Literature DB >> 27389062

Rapid and sensitive detection of Mycoplasma synoviae by an insulated isothermal polymerase chain reaction-based assay on a field-deployable device.

Hung-Chih Kuo1, Dan-Yuan Lo1, Chiou-Lin Chen1, Yun-Long Tsai2, Jia-Fong Ping2, Chien-Hsien Lee2, Pei-Yu Alison Lee3, Hsiao-Fen Grace Chang2.   

Abstract

Mycoplasma synoviae (MS), causing respiratory diseases, arthritis, and eggshell apex abnormalities in avian species, is an important pathogen in the poultry industry. Implementation of a biosecurity plan is important in MS infection management. Working on a field-deployable POCKIT™ device, an insulated isothermal polymerase chain reaction (iiPCR) assay has a potential for timely MS detection on the farm. The MS iiPCR assay had limit of detection 95% of about 9 genome equivalents by testing serial dilutions of a standard DNA. The detection endpoint of the assay for detection of MS genomic DNA was comparable to a reference real-time PCR. The assay did not crossreact with other important avian pathogens, including avian reovirus, Mycoplasma gallisepticum, Staphylococcus aureus, Escherichia coli, Pasteurella multocida, and Salmonella Pullorum. When 92 synovial fluid and respiratory tract swab samples collected from chickens, turkeys, and geese suspected of MS infection were tested, the clinical performance of the MS iiPCR had 97.8% agreement (Cohen's kappa value, 0.95) with that of the reference real-time PCR. In conclusion, the MS iiPCR/POCKIT™ system, working with field-deployable manual or automatic nucleic acid extraction methods, has potential to serve as a rapid and sensitive on-site tool to facilitate timely detection of MS.
© The Author 2016. Published by Oxford University Press on behalf of Poultry Science Association.

Entities:  

Keywords:  Mycoplasma synoviae; insulated isothermal PCR; point-of-need detection

Mesh:

Substances:

Year:  2016        PMID: 27389062      PMCID: PMC5161023          DOI: 10.3382/ps/pew228

Source DB:  PubMed          Journal:  Poult Sci        ISSN: 0032-5791            Impact factor:   3.352


INTRODUCTION

Mycoplasma synoviae (MS), one of the important Mycoplasma species that infect avian species, has been listed as a notifiable mycoplasma by World Organization for Animal Health (OIE, http://www.oie.int/) (OIE, 2015). MS infection could lead to symptoms including sinusitis, airsacculitis, synovitis, and eggshell apex abnormalities and result in reduction in egg production and meat quality, leading to considerable economic losses in the poultry industry worldwide (Kleven, 2008; Landman, 2014). MS infection is found commonly in chickens, turkeys, and guinea fowl, but less frequently in ducks, geese, and pigeons (Kleven, 2008; OIE, 2015). The morbidity of about 5 to 15% in chicken and 1 to 20% in turkey has been associated directly with MS infection alone and indirectly through synergistic effects of co-infection with other pathogens (Raviv et al., 2007; Kleven, 2008). Transmission of MS is accomplished laterally via direct contact and respiratory aerosols, and vertically within eggs. Like other pathogenic Mycoplasmas, MS could colonize and persist in the host (Noormohammadi, 2007; Kleven, 2008). Since MS infection may be asymptomatic and easily overlooked, routine MS monitoring has been recommended for biosecurity to help reduce MS outbreaks in the poultry industry (Landman, 2014). Furthermore, the symptoms caused by MS infection could not be differentiated easily from those caused by other avian pathogens, such as avian reovirus, Mycoplasma gallisepticum (MG), Staphylococcus aureus, Escherichia coli, Pasteurella multocida, and Salmonella spp., which in poultry species could cause symptoms similar to MS infection (Kleven, 2008; OIE, 2015). A live vaccine based on the temperature-sensitive MS-H strain is commercially available only in several countries (Morrow et al., 1998; Kreizinger et al., 2015). Therefore, keeping a flock MS-free through efficient biosecurity management has been suggested to be the most effective method to help protect the poultry investment (Kleven, 2008; Landman, 2014). Differentiation of MS from other pathogens could help implementation of proper therapeutic and/or prevention measures to improve flock health management. A rapid method to detect MS sensitively and specifically to help the diagnosis and monitoring of MS infection in the field is thus much needed (Kleven, 2008; Landman, 2014). Currently, the methods recommended by OIE for MS detection are bacterial isolation, serological assays, and polymerase chain reaction (PCR) (OIE, 2015). MS isolation, the gold-standard method for MS detection, is a slow process and may take up to 28 d to complete (Feberwee et al., 2005). MS isolation may also be compromised by competition from other pathogens, especially in the cases of chronically infected animals, which in general have relatively low MS loads (Ewing et al., 1996; Kleven, 2008; Roussan et al., 2015). Serological tests, being less time-consuming than MS isolation, have been the most commonly used method for MS detection. Nevertheless, antibody development takes about 2 to 4 wk and the antibody levels in turkeys are relatively low (Kleven, 2008). Moreover, serological tests have relatively high risks of cross-reactivity (particularly with other Mycoplasmas spp.) and low sensitivity (Yoder, 1989; Ewing et al., 1996; Ewing et al., 1998; Kleven, 2008). Alternatively, being relatively rapid, specific, and sensitive, PCR has been accepted to detect MS in clinical samples (Hess et al., 2007; Sprygin et al., 2010; OIE, 2015). Analysis of the vlhA (hemagglutinin) gene allowed intraspecific sequence typing of MS in conventional PCR assays (Hong et al., 2004; Hammond et al., 2009; OIE, 2015). In addition, real-time PCR assays targeting the 16S rRNA or vlhA gene are available to differentiate MS from its close relatives, such as MG (Jarquin et al., 2009; Sprygin et al., 2010; Fraga et al., 2013; Huang et al., 2015). Compared to conventional PCR, real-time PCR has higher detection sensitivity and specificity and is easier and faster to perform. However, as it requires a specialized laboratory and well-trained technicians, real-time PCR has been used mainly in diagnostic laboratories. Therefore, a sensitive, specific, and user-friendly detection tool is still much needed for rapid MS detection in the field to facilitate timely diagnosis and monitoring of MS infection for the poultry industry. A fluorescent probe hydrolysis insulated isothermal PCR (iiPCR) methodology has been developed specifically for on-site applications (Tsai et al., 2012). The reaction worked on a portable POCKIT™ Nucleic Acid Analyzer (POCKIT™; GeneReach, Taichung, Taiwan), which detected and processed reaction signals automatically to generate easy final positive and negative readouts. Carried out in a capillary tube and driven by the Rayleigh-Bènard convention (Krishnan et al., 2002), the reaction could generate detectable signals in less than 1 h. The iiPCR/POCKIT™ system has been demonstrated to achieve specific and sensitive detection of several bacterial and viral infections of veterinary interest in companion animals, livestock animals, and aquaculture animals (Tsai et al., 2012; Tsen et al., 2013; Balasuriya et al., 2014; Tsai et al., 2014; Wilkes et al., 2014; Ambagala et al., 2015; Lung et al., 2015; Wilkes et al., 2015a,b). In addition, a nucleic acid extraction step is generally required before PCR to remove reaction inhibitors in the sample matrix (Schrader et al., 2012). The iiPCR/POCKIT™ system is promising for field application because field-deployable automatic and manual nucleic extraction methods have been available to provide samples suitable for the reaction. Recently, an iiPCR assay (POCKIT™ M. synoviae Detection Kit, GeneReach) targeting the vlhA gene of MS has been available commercially for MS detection on the POCKIT™ device. In this study, analytical sensitivity and specificity of the assay were evaluated. Furthermore, clinical performance of the assay was verified by comparison with that of a previously published MS real-time PCR using clinical samples collected from chickens, turkeys, and geese suspected of MS infection.

MATERIALS AND METHODS

Microorganisms and Plasmids

MS (WVU 1853 [NCTC 10124] strain; ATCC 25204), MG (NCTC 10115, PG 31, X95 strain, ATCC 19610), Salmonella enterica subsp. enterica serovar Pullorum (NRRL B-663 [52-16T, CDC 2490-60, R. Hugh 529] strain, ATCC 13036), S. aureus Rosenbach (Seattle 1945 strain, ATCC 25923), E. coli (CDC EDL 932 strain, ATCC 43894), P. multocida subsp. multocida (NCTC 10322 [W-9217] strain, ATCC 43137), and avian reovirus (ATCC VR-2449) were used in the analysis; the specificity assays were purchased from the American Type Culture Collection (Manassas, VA). The pMSvlha and pMGvlha plasmids contained a fragment of the MS vlhA gene (nt 292195 - 292694, GenBank accession number AE017245) and a fragment of the vlhA gene from MG (pMGA1.7 precursor or hemagglutinin gene; nt 2901 - 3101, GenBank accession number U90714), respectively. They were synthesized on the basis of the pUC57 plasmid (Shanghai Generay Biotech, Shanghai, China). Concentrations of the plasmids were determined by a NanoDrop 2000 (Thermo Scientific, Wilmington, DE).

Clinical Samples

A total of 92 samples, including 15 synovial fluids, 32 laryngeal swabs, 39 nasal swabs, and 6 tracheal swabs, were analyzed. The clinical samples were collected from chickens, turkeys, or geese suspected of MS infection at various poultry farms in Taiwan in 2014. The clinical signs included a greenish discoloration of dropping, pale comb, lameness, swelling joints, listlessness, dehydration, emaciation, rales, and sinusitis. The swabs were placed into 1 mL of minimal essential medium. Synovial fluid was collected by direct penetration of the joint space using a 23 to 25 G needle with a small (3 mL) syringe (Campbell, 1994). Samples were stored at –80°C before DNA extraction.

Nucleic Acid Extraction

The taco™ mini Nucleic Acid Automatic Extraction System (taco™ mini; GeneReach) is a field-deployable device and can work with the iiPCR assay for on-site pathogen detection. In this study, nucleic acids were isolated by the taco™ mini device using the taco™ DNA/RNA Extraction Kit (GeneReach) according to the manufacturer's instructions. In brief, the wells of the extraction plate were filled with the designated reagent buffers. After 200 μL of the samples were added into the sample wells, the plate was loaded into a taco™ mini device for automatic DNA extraction. The DNA extracts eluted in 200 μL ELUTION BUFFER were individually collected and stored at –80°C for later use.

M. synoviae Insulated Isothermal PCR

The MS iiPCR assay (POCKIT™ M. synoviae Detection Kit, GeneReach) designed to target the vlhA (hemagglutinin) gene of MS was provided in a lyophilized format. The reagent was reconstituted with 50 μL of Premix Buffer B right before use. After the addition of 5 μL of the nucleic acid sample, the reaction mixture was transferred to an R-tube™ (GeneReach), which was placed subsequently into a POCKIT™ device to complete the amplification, detection, and data interpretation steps. Positive/negative qualitative readouts were displayed on the screen of the device within 1 h.

M. synoviae Real-Time PCR

The reference MS real-time PCR targeting the vlhA gene (Sprygin et al., 2010) was performed on an Applied Biosystems® 7500 Real-Time PCR System (Thermo Fisher Scientific, Carlsbad, CA). The 25-μL reaction mixture contained 1 x PCR Reaction Buffer (BioMi, Taichung, Taiwan), 0.5 mM dNTP, 0.8 μM forward primer (5′-CCAGGAGGTGGTACAGTTGAC-3′), 0.8 μM reverse primer (5′-TTAATGCTTCTTTAACT(G/A)AATCTGA-3′), 0.2 μM probe (FAM-5′-CTGCTAAAACAGAAGCTAAAAC(C/T)GCTAT-3′-BHQ1), 30 nM ROX, and 2.5 units of Taq DNA polymerase (BioMi) and 2 μL DNA sample. The cycling program included one cycle at 95°C for 5 min, followed by 40 cycles of 95°C for 25 sec, 58°C for 35 sec, and 72°C for 45 sec. The amplicon was 124 base pairs long. A typical example of the MS real-time PCR (data not shown) had a linearity range between 106 and 101 copies of the standard plasmid DNA (pMSvlha) with a slope of –3.10. The correlation coefficient was 0.99 with a y-intercept of 36.5. All reactions that had a recorded threshold cycle (C) number value were considered positive.

Analytical Sensitivity and Specificity

Replicates of reactions containing different copies of the plasmid (100, 50, 20, 10, 5, and 0 copies) and serial dilutions of the MS genomic DNA (10-, 102-, 103-, 104-, and 105-fold) were used to determine the analytical sensitivity of the MS iiPCR assay. Dilutions of the pMSvlha plasmid or genomic DNA of a MS culture were prepared in 40 ng/μL yeast tRNA. For the evaluation of analytical specificity, nucleic acids of the organisms in the exclusivity panel and the pMGvlha (106 copies) were tested.

Statistical Analysis

Limit of detection 95% (LOD95%) of the assay was determined by probit analysis at 95% confidence interval by using the SPSS v14 (SPSS, Chicago, IL). Kappa analysis of a 2 × 2 contingency table was used to assess the agreement between 2 assays.

RESULTS

Analytic Sensitivity of M. synoviae Insulated Isothermal PCR Assay

To evaluate the sensitivity of the MS iiPCR assay on the POCKIT™ device, serial dilutions of the standard plasmid were tested. All reactions containing more than 20 copies of the template were positive (Table 1), while 90% (18/20) of the 10-copy and 95% (19/20) of the 5-copy reactions were positive. The LOD95% of the assay calculated by the probit regression analysis was 9 genome equivalents per reaction.
Table 1.

Analytical sensitivity of MS iiPCR assay.

Plasmid DNA (copy)No. totalNo. positiveRate (%)
1001010100
502020100
202020100
10201890
5201995
02400
Analytical sensitivity of MS iiPCR assay. In addition, the analytical sensitivity of the MS iiPCR assay for MS genomic DNA was compared to that of the published MS real-time PCR, which also targeted the vlhA gene (Sprygin et al., 2010). The detection limits were found at the 103-fold dilution in both assays (Table 2), indicating that both assays had comparable analytical detection sensitivity for MS genomic DNA.
Table 2.

Detection limit of MS iiPCR assay: Comparison with the reference real-time PCR.

Delution foldReal-time PCR (Ct)iiPCR
1028.3228.7629.09+++
10231.9732.3232.24+++
10336.0035.8334.92+++
104ND38.0938.40++
105NDNDND

Ct, threshold cycle; ND, not detected.

Detection limit of MS iiPCR assay: Comparison with the reference real-time PCR. Ct, threshold cycle; ND, not detected.

Analytical Specificity of M. synoviae Insulated Isothermal PCR Assay

Analytical specificity of the MS iiPCR assay on the POCKIT™ device was verified with an exclusivity panel of 6 avian pathogens (MG, S. enterica, S. aureus, E. coli, P. multocida, and avian reovirus), which cause symptoms similar to those of the MS infection in poultry species (Jones, 2000; Kleven, 2008). The assay did not detect any of the 6 pathogens (data not shown). Furthermore, as the MG pMGA1.7 gene had been found to have >90% identity with the MS vlhA gene (Noormohammadi et al., 1998), the specificity of the MS iiPCR was further verified analytically with an artificial template of the homologous MG pMGA1.7 sequence. Negative test results were obtained, demonstrating its high specificity for the MS vlhA sequence (data not shown).

Diagnostic Performance of M. synoviae Insulated Isothermal PCR Assay

The diagnostic performance of the MS iiPCR/POCKIT™ assay to detect MS in avian clinical samples was compared to that of the reference real-time PCR (Sprygin et al., 2010). Among the 92 samples (15 synovial fluids, 32 laryngeal swabs, 39 nasal swabs, and 6 tracheal swabs) collected from chickens, turkeys, and geese suspected of MS infection (Table 3), 29 tested positive (Ct = 21.54 to 38.88) and 63 were negative by the real-time PCR. The MS iiPCR assay also reacted positively and negatively with 29 and 63 samples, respectively. One of the 29 MS real-time PCR-positive samples was negative by the MS iiPCR, and 1 of the 63 real-time PCR-negative samples was positive by the MS iiPCR assay. The agreement between the 2 methods was 97.8% (95% confidence interval, 93.78 to 100%) with a Cohen's kappa value of 0.95 (Table 4), suggesting that the MS iiPCR assay on POCKIT™ system was comparable to the real-time PCR to detect MS in avian clinical samples.
Table 3.

Detection of MS by MS real-time PCR and MS iiPCR assay: testing synovial fluid, and laryngeal, nasal, and tracheal swab samples from chickens, turkeys, and geese suspected of MS infection.

Sample typeSample IDReal-time PCR (Ct)iiPCR
Chicken/synovial fluidCJ1ND
CJ2ND
CJ328.64+
CJ431.58+
CJ521.54+
CJ623.08+
CJ724.18+
CJ8ND+
CJ9ND
CJ10ND
CJ11ND
CJ12ND
CJ13ND
CJ14ND
CJ15ND
Chicken/swabCL130.67+
CL236.75+
CL328.99+
CL433.67+
CL538.11
CL636.11+
CL7ND
CL8ND
CL9ND
CL1036.08+
CL1135.67+
CL12ND
CL13ND
CL14ND
CL1533.07+
CL16ND
CN138.88+
CN227.86+
CN333.10+
CN427.62+
CN5ND
CN6ND
CN7ND
CN8ND
CN9ND
CN10ND
CN1135.24+
CN12ND
CN13ND
CN14ND
CN1535.33+
CN16ND
CN17ND
CN18ND
CN19ND
CN20ND
CN21ND
CN22ND
CT1ND
CT2ND
CT3ND
CT4ND
CT5ND
CT6ND
Turkey/swabTL129.27+
TL229.93+
TL330.68+
TL427.12+
TL526.21+
TL635.35+
TN1ND
TN230.28+
TN335.54+
TN4ND
TN533.06+
TN6ND
Goose/swabGL1ND
GL2ND
GL3ND
GL4ND
GL5ND
GL6ND
GL7ND
GL8ND
GL9ND
GL10ND
GN1ND
GN2ND
GN3ND
GN4ND
GN5ND
GN6ND
GN7ND
GN8ND
GN9ND
GN10ND
GN11ND

CJ, chicken synovial/joint fluid; CL, chicken laryngeal swab; CN, chicken nasal swab; CT, chicken tracheal swab; TL, turkey laryngeal swab; TN, turkey nasal swab; GL, goose laryngeal swab; GN, goose nasal swab; Ct, threshold cycle; ND, not detected.

Table 4.

A 2 × 2 performance comparison between MS iiPCR and real-time PCR: detection of MS in clinical samples from animals suspected of MS infection.

Real-time PCR
PositiveNegative
iiPCRPositive281
Negative162
Detection of MS by MS real-time PCR and MS iiPCR assay: testing synovial fluid, and laryngeal, nasal, and tracheal swab samples from chickens, turkeys, and geese suspected of MS infection. CJ, chicken synovial/joint fluid; CL, chicken laryngeal swab; CN, chicken nasal swab; CT, chicken tracheal swab; TL, turkey laryngeal swab; TN, turkey nasal swab; GL, goose laryngeal swab; GN, goose nasal swab; Ct, threshold cycle; ND, not detected. A 2 × 2 performance comparison between MS iiPCR and real-time PCR: detection of MS in clinical samples from animals suspected of MS infection.

DISCUSSION

In this study, the MS iiPCR assay was shown to have excellent analytical sensitivity and specificity to detect MS nucleic acids. The MS vlhA gene, also the target sequence of several PCR assays verified previously to detect MS specifically (Sprygin et al., 2010; Huang et al., 2015), was highly conserved among all sequences (n = 180) available in the GenBank. Notably, no cross-reactivity was found with the MG organism as well as with the plasmid containing the highly homologous region from the MG pMGA1.7 gene in this study, demonstrating excellent specificity of the MS iiPCR assay. Specificity of the assay was shown further by the fact that no signals were detected with the 6 important avian pathogens tested in this study. Moreover, no sequences with significant homology to the MS vlhA gene was found from nucleotide BLAST analysis (http://www.ncbi.nlm.nih.gov/BLAST) of the GenBank database, which included complete genomes of 3 Mycoplasma and Acholeplasma species that were likely to be encountered in the chicken trachea (M. gallinarum [GenBank accession no. JHZE00000000], and M. gallinaceum [GenBank accession no. CP011021], and Acholeplasma laidlawii [GenBank accession no. NC_010163]), suggesting that the MS iiPCR assay was not likely to have cross-reactivity with these species. Test results of the 92 poultry samples from chickens, turkeys, and geese suspected of MS infection (Table 3) demonstrated that the MS iiPCR assay had clinical performance comparable to that of the reference real-time PCR (kappa = 0.95), which was demonstrated previously to have a detection limit for MS at 1 CFU equivalent/mL or 29 DNA copies per reaction (Sprygin et al., 2010). The one real-time PCR positive/iiPCR negative sample (CL5) had a Ct of 38.11, suggesting that the titers of MS in this sample were extremely low. Possible reasons for the one real-time PCR negative/iiPCR positive result include different effects of the sample matrix and the nucleic acid extraction method of the 2 PCR assays, as well as low MS titers in the sample. Overall, synovial fluids, laryngeal, nasal, and tracheal swab samples were shown to be suitable to work with the MS iiPCR assay for MS detection. In addition, the assay could be used for MS detection not only in chickens, but also in turkey and goose samples. The iiPCR/POCKIT™ system, designed to be field-deployable to allow point-of-need use (Tsai et al., 2014), could be a potentially useful tool for on-site MS monitoring. The high specificity and sensitivity of the MS iiPCR assay shown in this report was facilitated by the TaqMan probe-based PCR amplification and detection technology. The MS iiPCR/POCKIT™ assay could be completed easily by a user with basic operation training because only several reaction assembly steps are needed before the reaction and no data interpretation afterwards was needed. The POCKIT™ device collected fluorescent signals before and after the reaction to calculated the signal-to-noise (S/N) ratio (Tsai et al., 2012). Based on the default thresholds, the default algorithms converted automatically the S/N ratio into “+” (positive), “–” (negative), or “?” (inconclusive) results which were shown on the screen. The “?” results indicated that the signals were ambiguous and the sample needed to be retested. Eight reactions could be completed per run in the compact device (28 × 25 × 8.5 cm, W × D × H; 2.1 kg). Besides being rugged, the device could operate on a rechargeable or car battery. The lyophilized reagents could be shipped to distant places without a cold chain and stay stable for up to 2 years in a refrigerator, eliminating the need to ship samples to a diagnostic laboratory. Thus, the relatively inexpensive iiPCR/POCKIT™ system could serve as a tool to greatly reduce test turn-around time and costs. For on-site MS detection, the development of a loop-mediated isothermal amplification (LAMP)-based method targeting the MS vlhA gene was recently reported (Kursa et al., 2015). Relying on SYBR Green-based amplicon detection, the MS LAMP assay showed great analytical sensitivity and specificity. However, non-specific products may easily lead to false-positive results and the clinical performance of the MS LAMP assay remains to be verified. Because the Taq DNA polymerase is sensitive to inhibitors present in different clinical samples (Schrader et al., 2012), it is often necessary to purify nucleic acids from samples before the PCR amplification step. In general, automatic nucleic acid extraction methods could reduce labor costs and improve performance consistency in PCR-based molecular diagnosis. Basic requirements for an easy nucleic acid extraction system to work at points of need include light weight, user-friendliness, and being car-battery chargeable. A field-deployable automatic nucleic acid extraction device (taco™ mini) that could fulfill these criteria is available to work with the iiPCR/POCKIT™ system. Combining the iiPCR/POCKIT™ system with the automatic extraction device, minimal hands-on time was required at the sample preparation step for taco™ mini and at the reaction assembly step for iiPCR. Although being less user-friendly than the automatic methods, manual nucleic acid extraction methods are usually cheaper and well accepted in situations where resources are limited. A manual nucleic acid extraction method (PetNAD™ Nucleic Acid Co-prep kit, GeneReach) is also available to prepare nucleic acids for the MS iiPCR assay. With both nucleic extraction methods, the time required to obtain results from samples could be within 2 h. In conclusion, the rapid and user-friendly MS iiPCR/POCKIT™ system is a potential tool to help timely identification of MS infection at points of need in the poultry industry. With the high detection sensitivity and specificity demonstrated herein, the MS iiPCR reagent set could also be used to help diagnose MS infection at early and chronic stages.
  31 in total

1.  Specific detection and typing of Mycoplasma synoviae strains in poultry with PCR and DNA sequence analysis targeting the hemagglutinin encoding gene vlhA.

Authors:  Yang Hong; Maricarmen García; Victoria Leiting; Dulan Bencina; Louise Dufour-Zavala; Guillermo Zavala; Stanley H Kleven
Journal:  Avian Dis       Date:  2004-09       Impact factor: 1.577

2.  A survey of Mycoplasma gallisepticum and Mycoplasma synovaie with avian influenza H9 subtype in meat-type chicken in Jordan between 2011-2015.

Authors:  Dergham Ahmad Roussan; Ghassan Khawaldeh; Ibrahim Ali Shaheen
Journal:  Poult Sci       Date:  2015-05-12       Impact factor: 3.352

3.  Is Mycoplasma synoviae outrunning Mycoplasma gallisepticum? A viewpoint from the Netherlands.

Authors:  Wil J M Landman
Journal:  Avian Pathol       Date:  2014       Impact factor: 3.378

4.  Rapid detection of equine influenza virus H3N8 subtype by insulated isothermal RT-PCR (iiRT-PCR) assay using the POCKIT™ Nucleic Acid Analyzer.

Authors:  Udeni B R Balasuriya; Pei-Yu Alison Lee; Ashish Tiwari; Ashley Skillman; Bora Nam; Thomas M Chambers; Yun-Long Tsai; Li-Juan Ma; Pai-Chun Yang; Hsiao-Fen Grace Chang; Hwa-Tang Thomas Wang
Journal:  J Virol Methods       Date:  2014-06-30       Impact factor: 2.014

5.  A duplex real-time PCR assay for the detection and quantification of avian reovirus and Mycoplasma synoviae.

Authors:  Li Huang; Zhixun Xie; Liji Xie; Xianwen Deng; Zhiqin Xie; Sisi Luo; Jiaoling Huang; Tingting Zeng; Jiaxun Feng
Journal:  Virol J       Date:  2015-02-12       Impact factor: 4.099

6.  Rapid, Simple and Cost-Effective Molecular Method to Differentiate the Temperature Sensitive (ts+) MS-H Vaccine Strain and Wild-Type Mycoplasma synoviae Isolates.

Authors:  Zsuzsa Kreizinger; Kinga Mária Sulyok; Alexandra Pásztor; Károly Erdélyi; Orsolya Felde; János Povazsán; László Kőrösi; Miklós Gyuranecz
Journal:  PLoS One       Date:  2015-07-24       Impact factor: 3.240

7.  Development of TaqMan probe-based insulated isothermal PCR (iiPCR) for sensitive and specific on-site pathogen detection.

Authors:  Yun-Long Tsai; Hwa-Tang Thomas Wang; Hsiao-Fen Grace Chang; Chuan-Fu Tsai; Ching-Ko Lin; Ping-Hua Teng; Chen Su; Chien-Chung Jeng; Pei-Yu Lee
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

8.  Validation of a commercial insulated isothermal PCR-based POCKIT test for rapid and easy detection of white spot syndrome virus infection in Litopenaeus vannamei.

Authors:  Yun-Long Tsai; Han-Ching Wang; Chu-Fang Lo; Kathy Tang-Nelson; Donald Lightner; Bor-Rung Ou; Ai-Ling Hour; Chuan-Fu Tsai; Cheng-Chi Yen; Hsiao-Fen Grace Chang; Ping-Hua Teng; Pei-Yu Lee
Journal:  PLoS One       Date:  2014-03-13       Impact factor: 3.240

9.  A Multiplex real-time PCR for detection of Mycoplasma gallisepticum and Mycoplasma synoviae in clinical samples from Brazilian commercial poultry flocks.

Authors:  Aline Padilha Fraga; Tatiana de Vargas; Nilo Ikuta; André Salvador Kazantzi Fonseca; Álvaro José Celmer; Edmundo Kanan Marques; Vagner Ricardo Lunge
Journal:  Braz J Microbiol       Date:  2013-10-30       Impact factor: 2.476

10.  Rapid and sensitive detection of canine distemper virus by one-tube reverse transcription-insulated isothermal polymerase chain reaction.

Authors:  Rebecca P Wilkes; Yun-Long Tsai; Pei-Yu Lee; Fu-Chun Lee; Hsiao-Fen Grace Chang; Hwa-Tang Thomas Wang
Journal:  BMC Vet Res       Date:  2014-09-09       Impact factor: 2.741

View more
  4 in total

1.  Development and validation of an insulated isothermal PCR assay for the rapid detection of Mannheimia haemolytica.

Authors:  Xiaofang Dao; Chien-Che Hung; Youwen Yang; Juan Wang; Falong Yang
Journal:  J Vet Diagn Invest       Date:  2022-02-10       Impact factor: 1.279

2.  Rapid and visual detection of Mycoplasma synoviae by recombinase-aided amplification assay combined with a lateral flow dipstick.

Authors:  Wenlong Xia; Ke Chen; Wensong Liu; Yan Yin; Qian Yao; Yu Ban; Yiwen Pu; Xingmin Zhan; Hongchun Bian; Shupei Yu; Kunpeng Han; Ling Yang; Huanli Wang; Zhongjun Fan
Journal:  Poult Sci       Date:  2022-03-15       Impact factor: 4.014

3.  Evaluation and Clinical Validation of Two Field-Deployable Reverse Transcription-Insulated Isothermal PCR Assays for the Detection of the Middle East Respiratory Syndrome-Coronavirus.

Authors:  Yun Young Go; Yeon-Sook Kim; Shinhye Cheon; Sangwoo Nam; Keun Bon Ku; Meehyein Kim; Nam Hyuk Cho; Hyun Park; Pei-Yu Alison Lee; Yu-Chun Lin; Yun-Long Tsai; Hwa-Tang Thomas Wang; Udeni B R Balasuriya
Journal:  J Mol Diagn       Date:  2017-08-12       Impact factor: 5.568

4.  Rapid Diagnosis of Babesia gibsoni by Point-of-Need Testing by Insulated Isothermal PCR in Dogs at High Risk of Infection.

Authors:  K L Cooke; P Frenzer; S J Tucker; P C Crawford; S K Kirk; J K Levy
Journal:  J Vet Intern Med       Date:  2018-01       Impact factor: 3.333

  4 in total