Abstract

PURPOSE: To examine the hypothesis that alternative flush media could be used for lower extremity optical coherence tomography (OCT) imaging in long lesions that would normally require excessive use of contrast.
METHODS: The OPTical Imaging Measurement of Intravascular Solution Efficacy (OPTIMISE) trial was a single-center, prospective study (ClinicalTrials.gov identifier NCT01743872) that enrolled 23 patients (mean age 68±11 years; 14 men) undergoing endovascular intervention involving the superficial femoral artery. Four flush media (heparinized saline, dextran, carbon dioxide, and contrast) were used in succession in random order for each image pullback. Quality was defined as ≥270° visualization of vessel wall layers from each axial image. Mean proportions (± standard deviation) of image quality for each flush medium were assessed using 1-way analysis of variance and are reported with the 95% confidence intervals (CI).
RESULTS: Four OCT catheters failed, leaving 19 patients who completed the OCT imaging protocol; from this cohort, 51 highest quality runs were selected for analysis. Average vessel diameter was 3.99±1.01 mm. OCT imaging allowed 10- to 15-μm resolution of the lumen border, with diminishing quality as vessel diameter increased. Plaque characterization revealed fibrotic lesions. Mean proportions of image quality were dextran 87.2%±12% (95% CI 0.81 to 0.94), heparinized saline 74.3%±24.8% (95% CI 0.66 to 0.93), contrast 70.1%±30.5% (95% CI 0.52 to 0.88), and carbon dioxide 10.0%±10.4% (95% CI 0.00 to 0.26). Dextran, saline, and contrast provided better quality than carbon dioxide (p<0.001).
CONCLUSION: OCT is feasible in peripheral vessels <5 mm in diameter. Dextran or saline flush media can allow lesion characterization, avoiding iodinated contrast. Carbon dioxide is inadequate for peripheral OCT imaging. Axial imaging may aid in enhancing durability of peripheral endovascular interventions.
© The Author(s) 2016.

Entities:  

Keywords:  carbon dioxide; contrast media; dextran; imaging; lower limb; optical coherent tomography; peripheral vascular disease; plaque morphology; saline; superficial femoral artery

Mesh:

Substances:

Year:  2016        PMID: 27385151     DOI: 10.1177/1526602816657392

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  5 in total

1.  New developments for the detection and treatment of cardiac vasculopathy.

Authors:  Kevin J Clerkin; Ziad A Ali; Donna M Mancini
Journal:  Curr Opin Cardiol       Date:  2017-02-15       Impact factor: 2.161

2.  Electroanatomical Navigation to Minimize Contrast Medium or X-Rays During Stenting: Insights From an Experimental Model.

Authors:  Jean-François Dorval; Louis-Philippe Richer; Luc Soucie; Luke C McSpadden; Adam Hoopai; Stéphanie Tan; Nick E J West; E Marc Jolicoeur
Journal:  JACC Basic Transl Sci       Date:  2021-11-05

3.  Feasibility and Safety of Very-Low Contrast Combined Ringer's Solution in Optical Coherence Tomography Imaging.

Authors:  Tao Chen; Huai Yu; Lijia Ma; Chao Fang; Haibo Jia; Huimin Liu; Maoen Xu; Donghui Zhang; Guang Yang; Shuangyin Zhang; Jincheng Han; Guo Wei; Yanchao Liu; Jingbo Hou; Bo Yu
Journal:  Front Cardiovasc Med       Date:  2022-03-24

4.  A neurovascular high-frequency optical coherence tomography system enables in situ cerebrovascular volumetric microscopy.

Authors:  Giovanni J Ughi; Miklos G Marosfoi; Robert M King; Jildaz Caroff; Lindsy M Peterson; Benjamin H Duncan; Erin T Langan; Amanda Collins; Anita Leporati; Serge Rousselle; Demetrius K Lopes; Matthew J Gounis; Ajit S Puri
Journal:  Nat Commun       Date:  2020-07-31       Impact factor: 14.919

5.  Study of saline optical coherence tomography-guided percutaneous coronary intervention (SOCT-PCI Study).

Authors:  Nalin K Mahesh; Ankush Gupta; Parag Barward; Rajesh Vijayvergiya; Prafull Sharma; Ashwin Mahesh
Journal:  Indian Heart J       Date:  2020-04-13
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.