Literature DB >> 27373394

Outcomes of implantation of modified capsule tension rings with multiple black occluder paddles for eyes with congenital and acquired iris defects: Report 3.

Kevin M Miller1, C Manuel Nicoli2, Michael D Olson2, Manali Shah2, Samuel Masket2.   

Abstract

PURPOSE: To evaluate the safety and efficacy of Morcher 50F iris diaphragm implantation to manage moderate to large defects of the human iris.
SETTING: Stein Eye Institute, UCLA, Los Angeles, California, USA.
DESIGN: Prospective nonrandomized interventional case series.
METHODS: The demographic, preoperative, and postoperative data of patients who had implantation of modified capsular tension rings and followed to 1 year were reviewed. Safety measures included loss of corrected distance visual acuity (CDVA), surgical complications, adverse events, and secondary surgical interventions. Efficacy measures included CDVA with glare, daytime and nighttime glare symptom scores, and subjective cosmesis scores.
RESULTS: The study comprised 12 patients. The median CDVA was 20/70 before surgery and 20/20 after surgery. There were no lost lines of CDVA and no intraoperative complications. The most common postoperative complication was posterior capsule opacification. Two adverse events were unrelated to the device. Four patients had secondary surgical interventions, the most common of which was laser capsulotomy. The median CDVA with glare improved from less than 20/400 before surgery to 20/50 after surgery. One patient worsened. The median subjective daytime glare symptom score improved from 9 to 3 on a 10-point scale (P = .001). The median nighttime subjective glare symptom score improved from 8 to 2 (P = .001). The subjective cosmetic appearance of the eye stayed the same or improved for all patients (P = .031).
CONCLUSION: Iris diaphragm implantation was relatively safe and effective for reducing light and glare sensitivity in eyes with iris defects when combined with cataract extraction and intraocular lens implantation. FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27373394     DOI: 10.1016/j.jcrs.2016.03.035

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  Lessons Learned from Implantation of Morcher 50D and 96S Artificial Iris Diaphragms.

Authors:  Shawn R Lin; Kevin M Miller
Journal:  Case Rep Ophthalmol       Date:  2017-11-23

2.  Traumatic Aniridia and Aphakia Management with Iris Reconstruction Lens Using Gore-Tex Sutures, an Ab-Externo Approach.

Authors:  Efstratios Parikakis; Georgios Batsos; Loukas Kontomichos; Vasileios Peponis; Eleni Christodoulou; Dimitrios Karagiannis
Journal:  Am J Case Rep       Date:  2020-09-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.