Tokihiro Yamamoto1, Michael S Kent2, Erik R Wisner2, Lynelle R Johnson3, Joshua A Stern3, Lihong Qi4, Yukio Fujita5, John M Boone6. 1. Department of Radiation Oncology, University of California Davis School of Medicine, Sacramento, California 95817. 2. Department of Surgical and Radiological Sciences, University of California Davis School of Veterinary Medicine, Davis, California 95616. 3. Department of Medicine and Epidemiology, University of California Davis School of Veterinary Medicine, Davis, California 95616. 4. Department of Public Health Sciences, University of California Davis, Davis, California 95616. 5. Department of Radiation Oncology, Tokai University, Isehara, Kanagawa 259-1193, Japan. 6. Department of Radiology, University of California Davis School of Medicine, Sacramento, California 95817.
Abstract
PURPOSE: Radiotherapy (RT) that selectively avoids irradiating highly functional lung regions may reduce pulmonary toxicity, which is substantial in lung cancer RT. Single-energy computed tomography (CT) pulmonary perfusion imaging has several advantages (e.g., higher resolution) over other modalities and has great potential for widespread clinical implementation, particularly in RT. The purpose of this study was to establish proof-of-principle for single-energy CT perfusion imaging. METHODS: Single-energy CT perfusion imaging is based on the following: (1) acquisition of end-inspiratory breath-hold CT scans before and after intravenous injection of iodinated contrast agents, (2) deformable image registration (DIR) for spatial mapping of those two CT image data sets, and (3) subtraction of the precontrast image data set from the postcontrast image data set, yielding a map of regional Hounsfield unit (HU) enhancement, a surrogate for regional perfusion. In a protocol approved by the institutional animal care and use committee, the authors acquired CT scans in the prone position for a total of 14 anesthetized canines (seven canines with normal lungs and seven canines with diseased lungs). The elastix algorithm was used for DIR. The accuracy of DIR was evaluated based on the target registration error (TRE) of 50 anatomic pulmonary landmarks per subject for 10 randomly selected subjects as well as on singularities (i.e., regions where the displacement vector field is not bijective). Prior to perfusion computation, HUs of the precontrast end-inspiratory image were corrected for variation in the lung inflation level between the precontrast and postcontrast end-inspiratory CT scans, using a model built from two additional precontrast CT scans at end-expiration and midinspiration. The authors also assessed spatial heterogeneity and gravitationally directed gradients of regional perfusion for normal lung subjects and diseased lung subjects using a two-sample two-tailed t-test. RESULTS: The mean TRE (and standard deviation) was 0.6 ± 0.7 mm (smaller than the voxel dimension) for DIR between pre contrast and postcontrast end-inspiratory CT image data sets. No singularities were observed in the displacement vector fields. The mean HU enhancement (and standard deviation) was 37.3 ± 10.5 HU for normal lung subjects and 30.7 ± 13.5 HU for diseased lung subjects. Spatial heterogeneity of regional perfusion was found to be higher for diseased lung subjects than for normal lung subjects, i.e., a mean coefficient of variation of 2.06 vs 1.59 (p = 0.07). The average gravitationally directed gradient was strong and significant (R(2) = 0.99, p < 0.01) for normal lung dogs, whereas it was moderate and nonsignificant (R(2) = 0.61, p = 0.12) for diseased lung dogs. CONCLUSIONS: This canine study demonstrated the accuracy of DIR with subvoxel TREs on average, higher spatial heterogeneity of regional perfusion for diseased lung subjects than for normal lung subjects, and a strong gravitationally directed gradient for normal lung subjects, providing proof-of-principle for single-energy CT pulmonary perfusion imaging. Further studies such as comparison with other perfusion imaging modalities will be necessary to validate the physiological significance.
PURPOSE: Radiotherapy (RT) that selectively avoids irradiating highly functional lung regions may reduce pulmonary toxicity, which is substantial in lung cancer RT. Single-energy computed tomography (CT) pulmonary perfusion imaging has several advantages (e.g., higher resolution) over other modalities and has great potential for widespread clinical implementation, particularly in RT. The purpose of this study was to establish proof-of-principle for single-energy CT perfusion imaging. METHODS: Single-energy CT perfusion imaging is based on the following: (1) acquisition of end-inspiratory breath-hold CT scans before and after intravenous injection of iodinated contrast agents, (2) deformable image registration (DIR) for spatial mapping of those two CT image data sets, and (3) subtraction of the precontrast image data set from the postcontrast image data set, yielding a map of regional Hounsfield unit (HU) enhancement, a surrogate for regional perfusion. In a protocol approved by the institutional animal care and use committee, the authors acquired CT scans in the prone position for a total of 14 anesthetized canines (seven canines with normal lungs and seven canines with diseased lungs). The elastix algorithm was used for DIR. The accuracy of DIR was evaluated based on the target registration error (TRE) of 50 anatomic pulmonary landmarks per subject for 10 randomly selected subjects as well as on singularities (i.e., regions where the displacement vector field is not bijective). Prior to perfusion computation, HUs of the precontrast end-inspiratory image were corrected for variation in the lung inflation level between the precontrast and postcontrast end-inspiratory CT scans, using a model built from two additional precontrast CT scans at end-expiration and midinspiration. The authors also assessed spatial heterogeneity and gravitationally directed gradients of regional perfusion for normal lung subjects and diseased lung subjects using a two-sample two-tailed t-test. RESULTS: The mean TRE (and standard deviation) was 0.6 ± 0.7 mm (smaller than the voxel dimension) for DIR between pre contrast and postcontrast end-inspiratory CT image data sets. No singularities were observed in the displacement vector fields. The mean HU enhancement (and standard deviation) was 37.3 ± 10.5 HU for normal lung subjects and 30.7 ± 13.5 HU for diseased lung subjects. Spatial heterogeneity of regional perfusion was found to be higher for diseased lung subjects than for normal lung subjects, i.e., a mean coefficient of variation of 2.06 vs 1.59 (p = 0.07). The average gravitationally directed gradient was strong and significant (R(2) = 0.99, p < 0.01) for normal lung dogs, whereas it was moderate and nonsignificant (R(2) = 0.61, p = 0.12) for diseased lung dogs. CONCLUSIONS: This canine study demonstrated the accuracy of DIR with subvoxel TREs on average, higher spatial heterogeneity of regional perfusion for diseased lung subjects than for normal lung subjects, and a strong gravitationally directed gradient for normal lung subjects, providing proof-of-principle for single-energy CT pulmonary perfusion imaging. Further studies such as comparison with other perfusion imaging modalities will be necessary to validate the physiological significance.
Pulmonary toxicity is substantial in lung
cancer
radiotherapy (RT), particularly for locally advanced disease. Symptomatic (grade ≥ 2) radiation pneumonitis is a
common toxicity that occurs in approximately 30% of patients irradiated for
lung
cancer, with
fatal pneumonitis in about 2% of patients. The current paradigm of RT is based on anatomic
imaging and assumes a homogeneous radiation
dose–response of
normal tissues. RT that selectively avoids irradiating highly functional lung regions may reduce
pulmonary toxicity. This hypothesis is supported by several reports in the literature,
demonstrating that lung
dose-function
metrics improve predictive power for pulmonary toxicity compared to dose–volume metrics (the
current clinical standard). In
addition, information of regional lung function, such as perfusion defect, was found to be
beneficial in predicting toxicity.There are several modalities for pulmonary perfusion imaging,
including nuclear medicine
imaging, magnetic
resonance (MR) imaging, and
dual-energy computed
tomography
(CT)
imaging.
Perfusion images can also be acquired using single-energy CT scans with deformable
image
registration (DIR)-assisted image
subtraction, henceforth referred to as single-energy CT perfusion imaging.
Single-energy CT
perfusion imaging is based on the following: (1) breath-hold CT scans before and after
intravenous (IV) injection of iodinated contrast agents, (2) DIR, and (3) subtraction of the
precontrast image
data set from
the postcontrast image
data set,
yielding a map of regional Hounsfield unit (HU) enhancement as a surrogate for
perfusion. Single-energy CT perfusion imaging has a higher spatial resolution, shorter
scan time, and/or is potentially more cost-effective than other modalities. Moreover,
this method has great potential for widespread clinical implementation, particularly for
applications in RT, considering that single-energy CT is already available at
most RT centers and that many lung
cancerpatients
(especially those with centrally located tumors) receive CT scans with IV
contrast for
improved target delineation. Some of
these routinely acquired CT
image data could be used for perfusion imaging.
Single-energy CT
perfusion imaging was first proposed by Wildberger et al.
using a swine model with artificially induced pulmonary embolism. Their method was based on precontrast and postcontrast
CT scans
acquired during a single breath-hold (<30 s), and they successfully visualized
perfusion defects. However, a long breath-hold poses a challenge for clinical
implementation of this method, as many patients, especially lung
cancerpatients
with poor pulmonary function, cannot tolerate holding their breath for extended
periods.The purpose of this study was to establish proof-of-principle for single-energy
CT perfusion
imaging based on precontrast and postcontrast CT scans acquired with two
separate breath-holds using canine models of normal lungs and diseased
lungs. The
accuracy of DIR between the precontrast and postcontrast CT
image
data sets was
quantified, as this is the key to accurate computation of regional perfusion. Moreover,
we assessed spatial heterogeneity and gravitationally directed gradients of
regional perfusion for normal lung subjects and diseased lung subjects.
METHODS AND MATERIALS
Subjects
We used seven canines that were considered to have normal lungs (no clinical or
radiographic evidence of pulmonary disease) and seven canines with diseased
lungs
(including primary lung
tumor,
lung
metastasis,
and bronchointerstitial pneumonia) recruited from animals presenting to the William
R. Pritchard Veterinary Medical Teaching Hospital at University of California Davis
for veterinary care. Dogs had to be at least one year of age, weigh between 10 and 50
kg, and have adequate health and organ function as identified on a physical
examination, complete blood count, and a chemistry panel to allow safe anesthesia
administration. Owner informed consent was obtained. The protocol was approved by the
institutional clinical trials review board and the institutional animal care and use
committee.Dogs were premedicated with intramuscular administration of either butorphanol (0.3
mg/kg) or hydromorphone (0.05 mg/kg) and atropine (0.02 mg/kg). An IV catheter was
placed in a cephalic vein. Anesthesia was induced with propofol (2 mg/kg IV) and
midazolam (0.1 mg/kg IV) to effect in most cases. Dogs were intubated and maintained
on isoflurane gas anesthesia to effect. An arterial catheter was placed in the dorsal
pedal artery. Dogs were kept in ventral recumbency to limit the effects of
atelectasis. Heart rate, blood pressure, body temperature, depth of anesthesia, and
end-tidal carbon dioxide were monitored throughout the anesthetic procedure. The dogs
were then transported to the CT scanner and kept in ventral recumbency when placed on the
CT
table.
Overview of single-energy CT perfusion imaging
Figure 1 shows a schematic of image
acquisition, processing, and analysis for single-energy CT pulmonary perfusion
imaging. End-inspiratory breath-hold CT scans before and after
IV injection of iodinated contrast ( and , respectively) were acquired for perfusion
computation. In addition, two CT scans were also acquired at end-expiration
() and midinspiration () before contrast administration to build a model to correct
HUs of for variation in the lung inflation level
between the and scans. The main reason of using end-inspiratory
breath-hold CT
scans for perfusion computation was because this study was
designed based on a likely clinical scenario in RT applications. End-inspiratory
breath-hold CT
scans are sometimes preferred rather than end-expiratory
CT scans
because of a relatively long acquisition time for the whole lungs on CT scanners currently
available at most RT centers than on diagnostic CT scanners.
FIG. 1.
Schematic of image acquisition, processing and, analysis for single-energy CT
pulmonary perfusion imaging. Three precontrast breath-hold CT scans acquired at
end-expiration, midinspiration, and end-inspiration were used to correct HUs of the
precontrast end-inspiratory image for variation in the lung inflation level between
the precontrast and postcontrast end-inspiratory scans. A pair of the HU-corrected
precontrast image data set and deformed postcontrast image data set was used for
perfusion computation.
Breath-hold CT
imaging
Breath-hold CT
scans were acquired in the prone position with a LightSpeed
16-slice CT
scanner (GE Healthcare, Waukesha, WI). Assisted hyperventilation was performed before
CT scans
to facilitate breath-hold. Scan parameters were as follows: 120 kV, 150 mA, 1.25 mm
slice thickness, 1 s rotation time, and 1.375 pitch. For scans, iodinated contrast (3 ml/kg;
Isovue, 370 mg of iodine/ml, Bracco Diagnostics, Cranbury Township, NJ) was
administered intravenously at a flow rate of 4 ml/s through a peripheral IV line
inserted into the cephalic vein. The delay time between the start of contrast administration
and start of CT
scan ranged from 13 to 20 s, depending on the volume of
contrast
agent administered. CT
images were reconstructed using a filtered backprojection
algorithm with a sharp reconstruction kernel (GE Lung).
Deformable image
registration
DIR for spatial mapping of (moving) to (fixed) was performed for perfusion computation (Fig.
1). DIR of (moving) to (fixed) and then (moving) to (fixed) was also performed to correct HUs of
for variation in the lung inflation level (see
Sec. 2.E). elastix, an open source software
package for medical image
registration,
was used along with parameter settings used by Metz et al. B-spline DIR was performed in a
multigrid setting, which was driven by a similarity function and a transform bending
energy penalty (set to 0.05 for all conditions). For the similarity function, normalized cross correlation
and mutual information were compared to select the one that showed better performance
based on visual assessment. Normalized cross correlation was selected for DIR between
and , and mutual information for DIR between
and , and between and . For the image data, Gaussian pyramids were used
for down-sampling to increase robustness. For the B-spline transformation, a
multigrid approach was used. Five resolution levels were used with the following
parameter settings: down-sampling factor 16, 8, 4, 2, and 1; B-spline grid spacing
80, 40, 20, 10, and 5 mm. DIR was performed with lung masks that were
generated by segmenting the lungs on both the fixed and moving CT
images and merging those segmented lungs together. The DIR
method and parameter settings used in this study are similar to Staring et
al.The accuracy of DIR was evaluated based on the target registration error (TRE) of
anatomic pulmonary landmarks and on singularities in the displacement vector field
(DVF) in a manner similar to the MICCAI EMPIRE10 Challenge. The TRE is defined as the distance between
landmarks in the fixed image mapped from the moving
image by DIR and those mapped manually as the reference. Manual
annotation of landmarks was performed by a medical physicist. We used the iX
software to generate 50
landmarks per subject distributed throughout the lungs. The DVF was also
analyzed for singularities, i.e., regions where the DVF is not bijective. The
Jacobian determinant of the DVF was calculated for each voxel and analyzed to examine
whether there was any voxel with a negative Jacobian value. The following two
scenarios of DIR were evaluated for 10 randomly selected subjects out of 14 subjects:
(1) (moving) to (fixed), and (2) (moving) to (fixed).
HU correction for variation in lung inflation level
Lung HUs vary
with the lung
inflation level (depth of breathing), i.e., lower HUs for deeper breathing. Thus, the
lung
inflation variation between and hampers accurate computation of postcontrast HU
enhancement. To address this issue, we corrected HUs of for the lung inflation variation based on a relationship
between the HU for each voxel and total lung volume as a measure of the lung inflation level.
This relationship was determined using three precontrast CT scans
(, , and ) as follows. First, DIR was performed to establish
the spatial
correspondence between the three image
data sets for
each voxel. Second, a relationship between the HU and total lung volume (measured by
segmented lungs in the CT
image) was modeled using linear regression for each voxel. Third,
this relationship was used to estimate a HU at the lung inflation level of
by linear interpolation or extrapolation. Finally,
the original HU of was replaced with the estimated HU for each voxel,
yielding an image with precontrast HUs at the same lung inflation level as
. For perfusion computation, the resulting
HU-corrected
image
data set was
subtracted from the deformed
image
data set.
Analysis of CT perfusion images
Spatial
heterogeneity of regional perfusion was evaluated to determine whether diseased
lung
subjects demonstrated higher heterogeneity than normal lung subjects. Several
investigators reported significantly higher heterogeneity of regional perfusion or
ventilation for subjects with diseased lungs than healthy subjects. The coefficient of variation (CoV), i.e.,
standard deviation (SD)/mean of HU enhancement was used to quantify overall
heterogeneity of regional perfusion. The CoV has been used by several investigators
in the literature.
Lung masks
were generated by delineating voxels with less than −250 HU within lung outlines. A
two-sample two-tailed t-test was used to compare the CoVs of
diseased lung
subjects with those of normal lung subjects.Moreover, gravitationally directed gradients of regional perfusion were evaluated to
determine whether single-energy CT perfusion imaging
demonstrated the known effect of gravity on regional perfusion, i.e., greater
perfusion in gravity-dependent (ventral) regions than in nondependent (dorsal)
regions. This effect has been demonstrated with other imaging
modalities. The
slope (regression coefficient) was quantified from linear regression for the
relationship between the relative ventral-to-dorsal distance and HU enhancement. The
total lung
was divided into five coronal section regions of interest (ROIs), equally spaced
along the ventral-to-dorsal direction. The mean HU enhancement was calculated for
each ROI. Statistical analysis was performed to test whether the slope was
significantly different from zero (p < 0.05) using a two-sample
two-tailed t-test.
RESULTS
Accuracy of deformable image registration
Across the entire data
set, the mean TRE (and SD) was 0.6 ± 0.6 mm for DIR between
and , and 0.6 ± 0.7 mm for DIR between
and . The mean TRE was smaller than the voxel dimension
(approximately 1 × 1 × 1.25 mm3) in both DIR scenarios. Figure 2 shows a box plot of TREs for individual subjects.
For most cases, there were only several landmarks with a TRE exceeding the voxel
dimension. Subject C1 and D1 showed a relatively large number (range 6–13) of such
TREs in both DIR scenarios. In particular, subject D1 showed large TREs of up to 6.5
mm, many of which occurred around dark streaks of beam hardening artifacts from dense
iodinated contrast in great vessels. Table I shows weight and lung volumes of four breath-hold CT scans for individual
subjects. The % differences in the lung volume ranged from 11.2% to 67.1% for the pair
of and , and 0.2%–16.0% for the pair of
and . There were only weak correlations between the
maximum TREs and % differences in the lung volume for both pairs: and (r = 0.27),
and (r = 0.50). No singularities were
observed in the displacement vector fields for both DIR scenarios.
FIG. 2.
Box plots of target registration errors (TREs) for DIR (a) between end-expiratory
() and midinspiratory CT image data sets
(), and (b) between precontrast end-inspiratory
() and postcontrast end-inspiratory CT image data sets
() for five normal lung subjects and five diseased lung
subjects (50 pulmonary landmarks/subject). Each box is composed of three horizontal
lines corresponding to the 25th, 50th (median), and 75th percentile. Whiskers extend
to 2.7 times the standard deviation. Outliers are plotted as individual points.
TABLE I.
Weight and lung volumes of four breath-hold CT scans for seven normal lung subjects
and seven diseased lung subjects.
Lung volume (cm3)
% difference
Subject
Weight (kg)
CTend_exppre
CTmid_inspre
CTend_inspre
CTend_inspost
CTmid_inspre−CTend_exppre/CTend_exppre
CTend_inspost−CTend_inspre/CTend_inspre
Normal
C1
42.5
2217
3193
3532
3888
44.0
10.1
C2
13.1
302
468
514
481
54.7
−6.3
C3
25.5
963
1610
1732
1671
67.1
−3.5
C4
22.0
1592
1847
2059
2029
16.0
−1.4
C5
34.4
2090
2524
3075
3085
20.8
0.3
C6
36.4
1534
2330
2422
2316
51.9
−4.4
C7
24.4
792
1084
1137
1032
37.0
−9.3
Mean
28.3
1356
1865
2067
2072
41.7
−2.1
Diseased
D1
40.4
920
1296
1764
2046
40.9
16.0
D2
13.7
906
1212
1202
1295
33.8
7.7
D3
33.6
1533
2035
2294
2247
32.7
−2.0
D4
13.5
616
726
824
801
17.9
−2.7
D5
23.0
1468
1677
1959
1847
14.2
−5.7
D6
26.6
1489
1922
2060
2184
29.1
6.0
D7
21.2
968
1076
1171
1169
11.2
−0.2
Mean
24.6
1128
1420
1610
1656
25.7
2.7
Overall HU enhancement
Figure 3 shows the overall postcontrast HU
enhancement without and with HU correction of the precontrast CT
images for individual subjects. With HU correction, the mean
enhancement (and SD) was 37.3 ± 10.5 HU for normal lung subjects and 30.7 ±
13.5 HU for diseased lung subjects. There was no significant difference between the
normal and diseased lung groups (p = 0.33). The HU enhancement
varied widely among subjects, ranging from 17.9 HU (C4) to 60.3 HU (D1).
FIG. 3.
Overall postcontrast HU enhancement (average in the lung parenchyma) without and with
HU correction of precontrast CT images for seven normal lung subjects and seven
diseased lung subjects.
CT
perfusion images
Figure 4 shows example images of
, deformed , and perfusion for two representative subjects:
normal (C6) and diseased (D4). The
images showed considerable HU enhancement in great vessels and
slight enhancement in the lung parenchyma. The perfusion image of
subject C6 showed a more homogeneous distribution (CoV = 1.66) and more sharply
peaked histogram compared to subject D4 showing poorly perfused regions near the
tumors
(CoV = 2.10). Subject C6 showed higher perfusion in gravity-dependent ventral regions
than in nondependent regions.
FIG. 4.
Example images of the precontrast () (HU-corrected), postcontrast
() (deformed to precontrast) and perfusion for two
representative subjects: normal (C6) and diseased (D4). Histograms of perfusion (HU
enhancement) are also shown. Subject C6 showed a more homogeneous perfusion
distribution with a more sharp-peaked histogram and a lower coefficient of variation
(CoV) (1.66) compared to subject D4 (2.10).
Table II shows CoVs for individual subjects.
The mean CoV of the normal lung group was 1.59, which was lower (i.e., more homogeneous)
than the diseased lung group (mean CoV = 2.06) (p = 0.07).
Although the difference was not statistically significant at the 0.05 level, this
result suggests that spatial heterogeneity of regional perfusion is higher for
diseased lung
subjects than for normal lung subjects.
TABLE II.
CoVs as a measure of overall heterogeneity of regional perfusion for seven normal
lung dogs and seven diseased lung dogs.
Subject
CoV
Normal
C1
1.48
C2
1.14
C3
1.35
C4
2.44
C5
1.73
C6
1.66
C7
1.34
Mean
1.59
Diseased
D1
1.22
D2
2.36
D3
2.02
D4
2.10
D5
2.06
D6
1.97
D7
2.67
Mean
2.06
Figure 5 shows ventral-to-dorsal gradients of
regional perfusion for individual subjects as well as an average for each group. The
average ventral-to-dorsal gradient of the normal lung group was found to
be strong and significant (R2 = 0.99, p
< 0.01), indicating higher perfusion in gravity-dependent ventral regions than in
nondependent regions. In contrast, the diseased lung group showed a moderate, nonsignificant gradient
(R2 = 0.61, p = 0.12). The
R2 values ranged from 0.57 (subject C4) to 0.96
(subject C2) for normal lung subjects, and 0.12 (subject D4) to 0.94 (subject D7) for
diseased lung
subjects.
FIG. 5.
Ventral-to-dorsal gradients of regional perfusion for seven normal lung dogs, seven
diseased lung dogs, and average for each group. Each data point represents the mean
HU enhancement in a coronal section ROI. The equations from linear regression are
also shown for the average gradients.
DISCUSSION
This canine study on single-energy CT pulmonary perfusion imaging
demonstrated the accuracy of DIR with subvoxel TREs on average, higher spatial heterogeneity of
regional perfusion for diseased lung subjects than for normal lung subjects, and a strong
gravitationally directed gradient for normal lung subjects. These results provide proof-of-principle
for single-energy CT perfusion imaging. This is the first investigation
to quantify spatial heterogeneity and a gravitationally directed gradient using
single-energy CT
perfusion imaging. Our results were consistent with the previous studies based
on other imaging modalities. Vidal Melo et al. demonstrated
higher heterogeneity of regional perfusion measured by 13N positron emission
tomography (PET) for patients with chronic obstructive pulmonary disease (COPD) compared
to healthy subjects. Similarly,
Tzeng et al. showed higher ventilation heterogeneity measured by
hyperpolarized 3He MRI for asthmatic patients than healthy subjects. The gravitationally directed gradient
of regional perfusion has also been reported by many investigators with several
different imaging modalities including 99mTc-labeled
macroaggregated albumin (MAA) single-photon emission CT (SPECT), Fourier decomposition MRI, and arterial spin labeling MRI. High heterogeneity in diseased
lungs and a
ventral-to-dorsal gradient in normal lungs are both necessary but not sufficient conditions
for accurate perfusion imaging. Further studies such as a
comparison with other perfusion imaging modalities will be necessary for
validation in the future.The proposed method of single-energy CT perfusion imaging is considered a form of parametric
response mapping (PRM), an emerging method of classifying and measuring tissue changes
voxel by voxel. PRM has been used in oncology for assessment of response to therapy for
gliomas, head and neck
cancer, breast cancer, and metastatic prostate cancer to the bone, and moreover for assessment of COPD
phenotypes. For example, a PRM
parameter, i.e., % of voxels within the tumor yielding increased apparent diffusion coefficient
(ADC) values determined by pretreatment and midtreatment diffusion-weighted MRI, was
found to be predictive of clinical progression for head and neck cancerpatients, whereas %
changes in tumor
volume and whole-tumor ADC were not predictive. Similar analysis to investigate the predictive power of
single-energy CT
perfusion imaging for pulmonary toxicity after RT or respiratory diseases
would be a fascinating topic for future work.A pair of the end-expiration and midinspiration or end-inspiration breath-hold
CT scans
acquired for HU correction can also be used to compute regional volume change, a
surrogate for ventilation. Under normal conditions, regional ventilation and perfusion
are tightly matched to each other, yielding efficient gas exchange. However, normal
distributions of regional ventilation and perfusion change dramatically under
pathological conditions, leading to ventilation/perfusion mismatch and inefficient gas
exchange. For example, Vidal Melo
et al. reported significantly higher heterogeneity of
ventilation/perfusion ratios measured by 13N PET for sheep with pulmonary
embolism, saline lung lavage or bronchoconstriction compared to control sheep. Investigations into ventilation and
perfusion, e.g., comparison of ventilation/perfusion heterogeneity in diseased
lungs and in
normal lungs are
currently underway to provide further validation of CT perfusion imaging.There are several limitations in this study. First, dense concentration of iodinated
contrast in
great vessels caused dark streaks (beam hardening artifacts), which could lead to DIR
errors and negative HU enhancement due to lower HUs in the postcontrast CT
image. Voxels with negative HU enhancement appeared to be
distributed mainly in the proximity of great vessels as shown in Fig. 6. There were a considerable number of voxels with
negative HU enhancement, perhaps because of beam hardening artifacts and also residual
DIR errors and/or uncertainty in the HU correction method for precontrast CT
images. For future studies, strategies that reduce beam hardening
artifacts such as an iterative reconstruction algorithm might increase the accuracy of DIR and reduce negative HU
enhancement. Second, the HU enhancement was relatively subtle (34 HU on average) and
varied widely among subjects (range 17.9–60.3 HU), suggesting that there is room for
improvement. The average HU enhancement of 34 HU in this study is similar to the work of
Groell et al., which observed an enhancement of 30 HU in the normal
lung segment
by comparing precontrast and postcontrast CT scans. Also the 13–20 s delay time between contrast administration and
CT
image acquisition in this study is similar to other CT perfusion studies, e.g.,
13 s by Wildberger et al. and 15 s by Fuld et al., Given that this study included dogs with varying
breeds (e.g., Pembroke Welsh Corgi and Shepherd), weight (range 13.1–42.5 kg), and
lung
pathologies (normal, primary lung
tumor,
lung
metastasis, or
interstitial lung disease), there might be intersubject variability in the
hemodynamic profile, which could influence the optimal delay time. Optimizing the delay
time using a more specific patient population would be an important subject of future
work. Finally, different disease models were used in this study, which might have
contributed to a nonsignificant difference in the mean CoV between the normal and
diseased lung
groups (p = 0.07). Some dogs had relatively small lung
tumors that
might have little effect on regional perfusion. For future studies, respiratory diseases
that result in large perfusion defects (e.g., pulmonary embolism) may provide further
insights into the physiological significance of CT perfusion imaging. Also
other methods to quantify heterogeneity, including texture analysis, fractal techniques,
and Minkowski functionals, should also be considered.
FIG. 6.
Example images of the postcontrast CT (left) and voxels with negative HU enhancement
denoted in green (right) for subject D1. The number of voxels with negative HU
enhancement accounted for 15% of the lung voxels. (See color online version.)
Table III shows advantages and disadvantages of
pulmonary perfusion imaging modalities including single-energy
CT,
dual-energy CT,
SPECT, and MRI. One of the major disadvantages of single-energy CT perfusion imaging is
that multiple CT
acquisitions (i.e., precontrast and postcontrast CT, and CT at different
lung
inflation levels for HU correction) are required for accurate perfusion computation,
leading to a higher radiation
dose than other
modalities. This disadvantage may not pose a major challenge for applications in RT.
Most patients treated with modern RT receive CT scans for treatment planning. Also many
lung
cancerpatients
(especially centrally located tumors) receive
CT scans with
IV contrast for
improved target delineation on a routine basis. These treatment planning CT scans could also serve as
a precontrast or postcontrast CT for perfusion imaging. Furthermore, four-dimensional
(4D) CT acquired
for motion management is estimated to be used at approximately 70% of RT centers in the
US. A HU correction model could
also be built from the 4D CT
image data. Nevertheless, strategies for dose reduction and
optimization should be explored prior to clinical applications.
TABLE III.
Advantages and disadvantages of pulmonary perfusion imaging modalities.
Modality
Advantages
Disadvantages
Single-energy CT
Excellent availability of single-energy CT scanners in
RT centers, great potential for widespread implementation, high resolution, and
applicability to RT (large bore; electron density for dose calculation)
Multiple acquisitions required, DIR optimization and
validation required, radiation dose, and potential side effects of iodinated
contrast agents
Dual-energy CT
High resolution, applicability to RT (electron density
for dose calculation)
Limited availability in RT centers, virtual
noncontrast imaging may not satisfactory remove the need for multiple
acquisitions (Ref. 40), small FOV for
dual-source CT, typically ineffective in obese patients, radiation dose,
potential side effects of iodinated contrast agents, and typically small
bore
SPECT
Well-established
Limited resolution, limited availability in RT
centers, and radiation dose
MRI
No radiation dose and excellent soft tissue
contrast
Limited availability in RT centers and potential side
effects of gadolinium-based contrast agents (dynamic contrast-enhanced
MRI)
CONCLUSIONS
This canine study demonstrated the accuracy of DIR with subvoxel TREs on average, higher
spatial
heterogeneity of regional perfusion for diseased lung subjects than normal
lung
subjects, and a strong ventral-to-dorsal gradient for normal lung subjects, providing
proof-of-principle for single-energy CT pulmonary perfusion imaging.
Further studies such as comparison with other perfusion imaging
modalities (e.g., 99mTc − MAA SPECT) will be necessary to validate the
physiological significance.
Authors: Joachim Ernst Wildberger; Ernst Klotz; Hendrik Ditt; Elmar Spüntrup; Andreas H Mahnken; Rolf W Günther Journal: Eur Radiol Date: 2005-03-18 Impact factor: 5.315
Authors: Marcos F Vidal Melo; Dominick Layfield; R Scott Harris; Kevin O'Neill; Guido Musch; Torsten Richter; Tilo Winkler; Alan J Fischman; Jose G Venegas Journal: J Nucl Med Date: 2003-12 Impact factor: 10.057
Authors: Yvette Seppenwoolde; Katrien De Jaeger; Liesbeth J Boersma; José S A Belderbos; Joos V Lebesque Journal: Int J Radiat Oncol Biol Phys Date: 2004-11-01 Impact factor: 7.038
Authors: Katherina P Farr; Jesper F Kallehauge; Ditte S Møller; Azza A Khalil; Stine Kramer; Henrik Bluhme; Anni Morsing; Cai Grau Journal: Radiother Oncol Date: 2015-08-21 Impact factor: 6.280