Literature DB >> 27356555

Long term results of the titanium clip prosthesis.

Antoniu-Oreste Gostian1, Jean-Marc Kouame2, Martin Bremke2, Magdalene Ortmann3, Karl Bernd Hüttenbrink2, Dirk Beutner2.   

Abstract

The objective of the study is to evaluate the long term results of ossiculoplasty using the clip titanium partial ossicular replacement prosthesis. This study retrospectively reviews the partial ossiculoplasty conducted using clip titanium partial ossicular replacement prosthesis at a tertiary referral center. Audiometric outcomes and intraoperative findings were postoperatively measured from revision surgery of 47 ears (20 women, 27 men, mean age 43 years) averaging 6.5 years. The overall air-bone gap decreased from 25.7 dB preoperatively to 16.8 dB 6.5 years postoperatively (p ≤ 0.001, η p2  = 0.210). An air-bone gap of <20 dB was present in 28 % of ears preoperatively and increased to 72 % postoperatively. In revision (n = 30) and primary tympanoplasties (n = 17), the preoperative air-bone gaps were reduced from 28.9 and 20.1 to 18.7 dB (p ≤ 0.001, η p2  = 0.240) and 13.2 dB (p = 0.033, η p2  = 0.192), respectively. In canal wall down (n = 15) procedures and tympanoplasties with intact canal wall (n = 32), the mean air-bone gaps diminished from 28.9 to 18.1 dB (p = 0.02, η p2  = 0.245) and 24.2 to 16.1 dB (p ≤ 0.001, η p2  = 0.221), respectively. In our own revision tympanoplasties (n = 8) and second look operations (n = 6), we found that the prostheses were safe to remove without any deleterious effects. Two prostheses were dislocated from the stapes' head due to recurrent cholesteatoma. Therefore, it can be concluded that ossiculoplasty using the clip partial ossicular replacement prosthesis allows for good and reliable long term hearing results. Also, the flexible strips reveal no adverse effects on the encompassed stapes' head and do not complicate revision surgery.

Entities:  

Keywords:  Clip-prosthesis; Long-term results; Partial ossiculoplasty; Tympanoplasty

Mesh:

Substances:

Year:  2016        PMID: 27356555     DOI: 10.1007/s00405-016-4174-3

Source DB:  PubMed          Journal:  Eur Arch Otorhinolaryngol        ISSN: 0937-4477            Impact factor:   2.503


  21 in total

1.  Malleus-to-footplate prosthetic interposition: experience with 265 patients.

Authors:  V Colletti; F G Fiorino
Journal:  Otolaryngol Head Neck Surg       Date:  1999-03       Impact factor: 3.497

2.  The restoration of the function of the middle ear, in chronic otitis media.

Authors:  H WULLSTEIN
Journal:  Ann Otol Rhinol Laryngol       Date:  1956-12       Impact factor: 1.547

3.  Long-term results using ossicular grafts.

Authors:  J B Farrior; S W Nichols
Journal:  Am J Otol       Date:  1996-05

4.  [Replacement of ear ossicles with titanium prostheses].

Authors:  C H Stupp; H F Stupp; D Grün
Journal:  Laryngorhinootologie       Date:  1996-06       Impact factor: 1.057

Review 5.  [Surgical treatment of chronic otitis media. III: Middle ear reconstruction].

Authors:  K B Hüttenbrink
Journal:  HNO       Date:  1994-11       Impact factor: 1.284

6.  Osseointegrated titanium implants for the attachment of facial prostheses.

Authors:  H McComb
Journal:  Ann Plast Surg       Date:  1993-09       Impact factor: 1.539

7.  TORPs and PORPs: causes of failure--a report on 446 operations.

Authors:  J L Sheehy
Journal:  Otolaryngol Head Neck Surg       Date:  1984-10       Impact factor: 3.497

8.  Preliminary ossiculoplasty results using the Kurz titanium prostheses.

Authors:  Wesley W O Krueger; Joseph G Feghali; Clough Shelton; J Douglas Green; Charles W Beatty; David F Wilson; Bradley S Thedinger; David M Barrs; John T McElveen
Journal:  Otol Neurotol       Date:  2002-11       Impact factor: 2.311

9.  Titanium clip prosthesis.

Authors:  Karl-Bernd Hüttenbrink; Th Zahnert; E G Wüstenberg; G Hofmann
Journal:  Otol Neurotol       Date:  2004-07       Impact factor: 2.311

10.  Pathology of the ossicular chain in various chronic middle ear diseases.

Authors:  M Tos
Journal:  J Laryngol Otol       Date:  1979-08       Impact factor: 1.469

View more
  4 in total

1.  Applied Force during Piston Prosthesis Placement in a 3D-Printed Model: Freehand vs Robot-Assisted Techniques.

Authors:  Christopher R Razavi; Paul R Wilkening; Rui Yin; Nicolas Lamaison; Russell H Taylor; John P Carey; Francis X Creighton
Journal:  Otolaryngol Head Neck Surg       Date:  2018-12-04       Impact factor: 3.497

Review 2.  Biocompatible Materials in Otorhinolaryngology and Their Antibacterial Properties.

Authors:  Jakub Spałek; Przemysław Ociepa; Piotr Deptuła; Ewelina Piktel; Tamara Daniluk; Grzegorz Król; Stanisław Góźdź; Robert Bucki; Sławomir Okła
Journal:  Int J Mol Sci       Date:  2022-02-25       Impact factor: 5.923

3.  Study on the Prosthesis Structural Design and Vibration Characteristics Based on the Conduction Effect of Human Middle Ear.

Authors:  Wu Ren; Huijuan Yan; Yi Yu; Jinghong Ren; Jinlong Chang; Yidong Wang; Yibo Han
Journal:  Appl Bionics Biomech       Date:  2020-05-20       Impact factor: 1.781

4.  The Effect of the Status of the Ossicular Chain and Choice of Graft Material on Hearing Outcomes in Pediatric Cholesteatoma Surgery.

Authors:  Eiko Kimura; Colin G Leonard; Adrian L James
Journal:  J Int Adv Otol       Date:  2021-03       Impact factor: 1.316

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.