| Literature DB >> 27352025 |
Shuai Na1, Albert I H Chen1, Lawrence L P Wong1, Zhenhao Li1, Mirek Macecek2, John T W Yeow3.
Abstract
A novel design of an air-coupled capacitive micromachined ultrasonic transducer (CMUT) with annular cell geometry (annular CMUT) is proposed. Finite element analysis shows that an annular cell has a ratio of average-to-maximum displacement (RAMD) of 0.52-0.58 which is 58-76% higher than that of a conventional circular cell. The increased RAMD leads to a larger volume displacement which results in a 48.4% improved transmit sensitivity and 127.3% improved power intensity. Single-cell annular CMUTs were fabricated with 20-μm silicon plates on 13.7-μm deep and 1.35-mm wide annular cavities using the wafer bonding technique. The measured RAMD of the fabricated CMUTs is 0.54. The resonance frequency was measured to be 94.5kHz at 170-V DC bias. The transmit sensitivity was measured to be 33.83Pa/V and 25.85Pa/V when the CMUT was excited by a continuous wave and a 20-cycle burst, respectively. The receive sensitivity at 170-V DC bias was measured to be 7.7mV/Pa for a 20-cycle burst, and 15.0mV/Pa for a continuous incident wave. The proposed annular CMUT design demonstrates a significant improvement in transmit efficiency, which is an important parameter for air-coupled ultrasonic transducers.Entities:
Keywords: Air-coupled ultrasound; Annular cells; Capacitive micromachined ultrasonic transducers (CMUTs)
Year: 2016 PMID: 27352025 DOI: 10.1016/j.ultras.2016.06.008
Source DB: PubMed Journal: Ultrasonics ISSN: 0041-624X Impact factor: 2.890