Vitamin E-Coated and Heparin-Coated Dialyzer Membranes for Heparin-Free Hemodialysis: A Multicenter, Randomized, Crossover Trial.

Mohamed Shariful Islam1, Zarih Alcheikh Hassan2, Florence Chalmin3, Sandor Vido3, Mohamed Berrada3, David Verhelst2, Patrick Donnadieu2, Olivier Moranne3, Vincent L M Esnault4.   

Abstract

BACKGROUND: Hemodialysis requires effective anticoagulation to avoid blood circuit clotting. In patients at high risk for bleeding, several alternative methods have been developed. STUDY
DESIGN: Multicenter, prospective, randomized, crossover study evaluating the noninferiority of vitamin E-coated compared with heparin-coated dialyzers in a 4-hour heparin-free hemodialysis strategy. SETTINGS & PARTICIPANTS: 32 adult long-term hemodialysis patients from 2 French hemodialysis units with well-functioning fistulas or double-lumen catheters. INTERVENTION: Patients were randomly allocated to a first period using either vitamin E- or heparin-coated dialyzers. After a washout period of 2 hemodialysis sessions, each patient was switched to the alternative dialyzer for a second period. Each study period started with 2 hemodialysis sessions with reduced heparin dose (50% and 25% of usual heparin dose, respectively, for sessions 1 and 2) followed by 2 heparin-free sessions. OUTCOMES: The primary end point was the percentage of successful study periods, defined as no circuit-clotting event leading to premature interruption of any of the 4 dialysis sessions. Secondary end points included total number and cumulative duration of hemodialysis sessions without clotting, number of saline solution flushes, dialysis circuit bubble trap status and dialyzer membrane status by visual inspection, and dialysis adequacy.
RESULTS: The percentage of success with vitamin E-coated dialyzers (25/32 study periods [78%]) was not inferior to that with heparin-coated dialyzers (26/32 study periods [81%]). Visual inspection showed equal numbers of clean dialysis circuit bubble traps (vitamin E-coated, 34/121; heparin-coated, 32/120), whereas clean fiber bundles were more frequently noted with the vitamin E-coated compared with heparin-coated dialyzers (25/121 vs 2/120; P=0.002). LIMITATIONS: Results may not extrapolate to critically ill patients. Differences in dialyzer transparency may account for visual inspection scores.
CONCLUSIONS: The success rate of 4-hour heparin-free hemodialysis sessions is lower than that previously claimed in uncontrolled studies. Vitamin E-coated and heparin-coated dialyzers exposed patients to similar and unacceptable high failure rates. Further studies are required to improve heparin-free hemodialysis.
Copyright © 2016 National Kidney Foundation, Inc. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Heparin-free hemodialysis; anticoagulation; blood circuit clotting; end-stage renal disease (ESRD); fibrin ring; heparin-coated dialyzer; non-inferiority trial; randomized controlled trial; saline flushes; vitamin E-coated dialyzer

Mesh:

Substances:

Year:  2016        PMID: 27344212     DOI: 10.1053/j.ajkd.2016.05.013

Source DB:  PubMed          Journal:  Am J Kidney Dis        ISSN: 0272-6386            Impact factor:   8.860


  15 in total

Review 1.  Enhancing dialyser clearance-from target to development.

Authors:  Kamonwan Tangvoraphonkchai; Andrew Davenport
Journal:  Pediatr Nephrol       Date:  2017-04-12       Impact factor: 3.714

2.  Transient blood thinning during extracorporeal blood purification via the inactivation of coagulation factors by hydrogel microspheres.

Authors:  Xin Song; Haifeng Ji; Yupei Li; Yuqin Xiong; Li Qiu; Rui Zhong; Meng Tian; Jayachandran N Kizhakkedathu; Baihai Su; Qiang Wei; Weifeng Zhao; Changsheng Zhao
Journal:  Nat Biomed Eng       Date:  2021-01-25       Impact factor: 25.671

3.  Polyethyleneimine-treated polyacrylonitrile membrane hemofilter for critically ill patients receiving anticoagulant-free prolonged intermittent renal replacement therapy: a single-center, prospective, self-controlled pilot study.

Authors:  Tao Su; Qizhuang Jin; Zhongyuan Liu
Journal:  BMC Nephrol       Date:  2017-06-30       Impact factor: 2.388

4.  Reporting of "dialysis adequacy" as an outcome in randomised trials conducted in adults on haemodialysis.

Authors:  Sanne Steyaert; Els Holvoet; Evi Nagler; Simon Malfait; Wim Van Biesen
Journal:  PLoS One       Date:  2019-02-05       Impact factor: 3.240

5.  Clinical evaluation of performance, biocompatibility, and safety of vitamin E-bonded polysulfone membrane hemodialyzer compared to non-vitamin E-bonded hemodialyzer.

Authors:  Mercedeh Kiaii; Masaharu Aritomi; Mitsuyo Nagase; Myriam Farah; Beverly Jung
Journal:  J Artif Organs       Date:  2019-06-26       Impact factor: 1.731

6.  A randomized cross-over study with objective quantification of the performance of an asymmetric triacetate and a polysulfone dialysis membrane using different anticoagulation strategies.

Authors:  Floris Vanommeslaeghe; Iván Josipovic; Matthieu Boone; Annemie Dhondt; Wim Van Biesen; Sunny Eloot
Journal:  Clin Kidney J       Date:  2019-12-22

7.  Timed Controlled Repeated Rotation of the CAR-170-C NXSTAGE Chronic Cartridge Hemodialysis Filter: A Novel Approach to Enabling Heparin-Free Frequent Daily Home Hemodialysis.

Authors:  Adam Locke; Margaret A Bushey; Cynthia LaCroix; Patience Deardoff; Macaulay Amechi Chukwukadibia Onuigbo
Journal:  Mayo Clin Proc Innov Qual Outcomes       Date:  2020-11-06

8.  How biocompatible haemodialysers can conquer the need for systemic anticoagulation even in post-dilution haemodiafiltration: a cross-over study.

Authors:  Floris Vanommeslaeghe; Iván Josipovic; Matthieu Boone; Arjan van der Tol; Annemie Dhondt; Wim Van Biesen; Sunny Eloot
Journal:  Clin Kidney J       Date:  2020-12-08

9.  Prescribing Hemodialysis or Hemodiafiltration: When One Size Does Not Fit All the Proposal of a Personalized Approach Based on Comorbidity and Nutritional Status.

Authors:  Giorgina Barbara Piccoli; Louise Nielsen; Lurilyn Gendrot; Antioco Fois; Emanuela Cataldo; Gianfranca Cabiddu
Journal:  J Clin Med       Date:  2018-10-08       Impact factor: 4.241

10.  Evaluation of Different Dialyzers and the Impact of Predialysis Albumin Priming in Intermittent Hemodialysis With Reduced Anticoagulation.

Authors:  Floris Vanommeslaeghe; Filip De Somer; Iván Josipovic; Matthieu Boone; Wim Van Biesen; Sunny Eloot
Journal:  Kidney Int Rep       Date:  2019-07-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.