Abstract
The use of pulsed photon radiation in medical, industrial and security sectors has vastly increased during the recent years. The length of pulse from different X-ray flash generators that are commonly used as either portable, battery-operated or fixed systems can be as low as a few femtoseconds. However, the majority of radiation protection instruments, especially various active electronic dosemeters, have limitations when operated in pulsed fields. This study presents measurements that were performed using the dose-integration mode of the RI-02 and Ram Ion survey metres, which are based on ventilated ionisation chamber, when exposed to different pulsed X-ray sources. An intercomparison between the results that were obtained with the survey metres and those that were obtained with thermoluminescence dosemeter, reference passive dosemeter, show good agreement (deviation lies within 10%).Mesh:
Year: 2017 PMID: 27342448 DOI: 10.1093/rpd/ncw145
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972