Effect of load on the bone around bone-anchored amputation prostheses.

Patrik Stenlund1,2,3, Margarita Trobos1,3, Jukka Lausmaa1,2, Rickard Brånemark1,4, Peter Thomsen1,3, Anders Palmquist1,3.   

Abstract

Osseointegrated transfemoral amputation prostheses have proven successful as an alternative method to the conventional socket-type prostheses. The method improves prosthetic use and thus increases the demands imposed on the bone-implant system. The hypothesis of the present study was that the loads applied to the bone-anchored implant system of amputees would result in locations of high stress and strain transfer to the bone tissue and thus contribute to complications such as unfavourable bone remodeling and/or elevated inflammatory response and/or compromised sealing function at the tissue-abutment interface. In the study, site-specific loading measurements were made on amputees and used as input data in finite element analyses to predict the stress and strain distribution in the bone tissue. Furthermore, a tissue sample retrieved from a patient undergoing implant revision was characterized in order to evaluate the long-term tissue response around the abutment. Within the limit of the evaluated bone properties in the present experiments, it is concluded that the loads applied to the implant system may compromise the sealing function between the bone and the abutment, contributing to resorption of the bone in direct contact with the abutment at the most distal end. This was supported by observations in the retrieved clinical sample of bone resorption and the formation of a soft tissue lining along the abutment interface.
© 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 35:1113-1122, 2017. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  biomechanics; finite element analysis; interface; osseointegration; titanium implant

Mesh:

Year:  2016        PMID: 27341064     DOI: 10.1002/jor.23352

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  5 in total

Review 1.  [Transcutaneous osseointegrated prosthesis (TOP) after limb amputation : Status quo and perspectives].

Authors:  Christian Willy; Christian Krettek
Journal:  Unfallchirurg       Date:  2017-05       Impact factor: 1.000

2.  A preliminary, observational study using whole-blood RNA sequencing reveals differential expression of inflammatory and bone markers post-implantation of percutaneous osseointegrated prostheses.

Authors:  Andrew Miller; Sujee Jeyapalina; Jay Agarwal; Mitchell Mansel; James Peter Beck
Journal:  PLoS One       Date:  2022-05-26       Impact factor: 3.752

Review 3.  Biomechanical Characterisation of Bone-anchored Implant Systems for Amputation Limb Prostheses: A Systematic Review.

Authors:  Alexander Thesleff; Rickard Brånemark; Bo Håkansson; Max Ortiz-Catalan
Journal:  Ann Biomed Eng       Date:  2018-01-11       Impact factor: 3.934

4.  Finite Element Analysis of Transhumeral and Transtibial Percutaneous Osseointegrated Endoprosthesis Implantation.

Authors:  Carolyn E Taylor; Heath B Henninger; Kent N Bachus
Journal:  Front Rehabil Sci       Date:  2021-11-23

5.  Noncontact Strain Monitoring of Osseointegrated Prostheses.

Authors:  Sumit Gupta; Han-Joo Lee; Kenneth J Loh; Michael D Todd; Joseph Reed; A Drew Barnett
Journal:  Sensors (Basel)       Date:  2018-09-09       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.