A fine pointed glucose oxidase immobilized electrode for low-invasive amperometric glucose monitoring.

Jiang Li1, Pankaj Koinkar2, Yusuke Fuchiwaki3, Mikito Yasuzawa4.   

Abstract

A low invasive type glucose sensor, which has a sensing region at the tip of a fine pointed electrode, was developed for continuous glucose monitoring. Platinum-iridium alloy electrode with a surface area of 0.045mm(2) was settled at the middle of pointed PEEK (Polyetheretherketone) tubing and was employed as sensing electrode. Electrodeposition of glucose oxidase in the presence of surfactant, Triton X-100, was performed for high-density enzyme immobilization followed by the electropolymerization of o-phenylenediamine for the formation of functional entrapping and permselective polymer membrane. Ag/AgCl film was coated on the surface of PEEK tubing as reference electrode. Amperometric responses of the prepared sensors to glucose were measured at a potential of 0.60V (vs. Ag/AgCl). The prepared electrode showed the sensitivity of 2.55μA/cm(2) mM with high linearity of 0.9986, within the glucose concentration range up to 21mM. The detection limit (S/N=3) was determined to be 0.11mM. The glucose sensor properties were evaluated in phosphate buffer solution and in vivo monitoring by the implantation of the sensors in rabbit, while conventional needle type sensors as a reference were used. The results showed that change in output current of the proposed sensor fluctuated similar with one in output current of the conventional needle type sensors, which was also in similar accordance with actual blood sugar level measured by commercially glucose meter. One-point calibration method was used to calibrate the sensor output current.
Copyright © 2016 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Continuous glucose monitoring; In vivo measurement; Low-invasive device, Glucose oxidase

Mesh:

Substances:

Year:  2016        PMID: 27336616     DOI: 10.1016/j.bios.2016.06.037

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  5 in total

Review 1.  Blending Electronics with the Human Body: A Pathway toward a Cybernetic Future.

Authors:  Mehdi Mehrali; Sara Bagherifard; Mohsen Akbari; Ashish Thakur; Bahram Mirani; Mohammad Mehrali; Masoud Hasany; Gorka Orive; Paramita Das; Jenny Emneus; Thomas L Andresen; Alireza Dolatshahi-Pirouz
Journal:  Adv Sci (Weinh)       Date:  2018-08-01       Impact factor: 16.806

2.  Analysis and Testing of a Suitable Compatible Electrode's Material for Continuous Measurement of Glucose Concentration.

Authors:  Nikola Slaninova; Klara Fiedorova; Ali Selamat; Karolina Danisova; Jan Kubicek; Ewaryst Tkacz; Martin Augustynek
Journal:  Sensors (Basel)       Date:  2020-06-30       Impact factor: 3.576

3.  Temperature-Corrected Fluidic Glucose Sensor Based on Microwave Resonator.

Authors:  Chorom Jang; Jin-Kwan Park; Hee-Jo Lee; Gi-Ho Yun; Jong-Gwan Yook
Journal:  Sensors (Basel)       Date:  2018-11-09       Impact factor: 3.576

Review 4.  Non-Invasive Blood Glucose Monitoring Technology: A Review.

Authors:  Liu Tang; Shwu Jen Chang; Ching-Jung Chen; Jen-Tsai Liu
Journal:  Sensors (Basel)       Date:  2020-12-04       Impact factor: 3.576

5.  On-Chip Glucose Detection Based on Glucose Oxidase Immobilized on a Platinum-Modified, Gold Microband Electrode.

Authors:  Julia Madden; Colm Barrett; Fathima R Laffir; Michael Thompson; Paul Galvin; Alan O' Riordan
Journal:  Biosensors (Basel)       Date:  2021-07-25
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.