Andreas Wachtel1, Tycho Zimmermann1, Tobias Spintig1, Florian Beuer2, Wolf-Dieter Müller1, Andreas Dominik Schwitalla1.
Abstract
Bacterial leakage from the implant-abutment-interface (IAI) is suspected of contributing to the development of peri-implantitis. The aim of the study was to develop a straightforward test setup to evaluate the bacterial leakage of the IAI of 2-piece implant systems under laboratory conditions. A test suspension of Enterococcus faecium was injected into 7 implants (PerioType Rapid Implants) prior to abutment fixation. The IAI was covered by kanamycin aesculn azide agar (KAAA), which serves as an optical indicator for E. faecium . The specimens were cyclically loaded with a force of 120 N for up to 1 000 000 cycles in a universal testing machine in accordance with the ISO 14801:2007 standard. The color change of the KAAA was recorded. Three of the 7 implants showed bacterial leakage before the cyclic loading test started. The bacterial tightness of the IAIs of the 4 remaining implants lasted for 35 680 ± 22 467 cycles on average. The experimental setup at hand provides the means for a straightforward evaluation of the bacterial tightness of the IAI of 2-piece dental implants.Entities:
Keywords: Enterococcus faecium; bacterial leakage; chewing simulation; dental implant-abutment interface; dental implants; peri-implantitis
Mesh:
Substances:
Year: 2016 PMID: 27327361 DOI: 10.1563/aaid-joi-D-16-00065
Source DB: PubMed Journal: J Oral Implantol ISSN: 0160-6972 Impact factor: 1.779