Literature DB >> 27321977

Evaluation of Collision Cross Section Calibrants for Structural Analysis of Lipids by Traveling Wave Ion Mobility-Mass Spectrometry.

Kelly M Hines1, Jody C May2, John A McLean2, Libin Xu1.   

Abstract

Collision cross section (CCS) measurement of lipids using traveling wave ion mobility-mass spectrometry (TWIM-MS) is of high interest to the lipidomics field. However, currently available calibrants for CCS measurement using TWIM are predominantly peptides that display quite different physical properties and gas-phase conformations from lipids, which could lead to large CCS calibration errors for lipids. Here we report the direct CCS measurement of a series of phosphatidylcholines (PCs) and phosphatidylethanolamines (PEs) in nitrogen using a drift tube ion mobility (DTIM) instrument and an evaluation of the accuracy and reproducibility of PCs and PEs as CCS calibrants for phospholipids against different classes of calibrants, including polyalanine (PolyAla), tetraalkylammonium salts (TAA), and hexakis(fluoroalkoxy)phosphazines (HFAP), in both positive and negative modes in TWIM-MS analysis. We demonstrate that structurally mismatched calibrants lead to larger errors in calibrated CCS values while the structurally matched calibrants, PCs and PEs, gave highly accurate and reproducible CCS values at different traveling wave parameters. Using the lipid calibrants, the majority of the CCS values of several classes of phospholipids measured by TWIM are within 2% error of the CCS values measured by DTIM. The development of phospholipid CCS calibrants will enable high-accuracy structural studies of lipids and add an additional level of validation in the assignment of identifications in untargeted lipidomics experiments.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27321977      PMCID: PMC4955523          DOI: 10.1021/acs.analchem.6b01728

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


Lipids are essential components of cell membranes and play important roles in normal physiology[1,2] as well as in human diseases such as atherosclerosis,[3] cancer,[4,5] and diabetes.[6] Over the past decade, the study of the complete profile and pathways of lipids in biological systems, or lipidomics, has emerged as an important new area of “omics” studies,[7−10] making up a major component of metabolomics that is complementary to genomics, transcriptomics, and proteomics. Advances in lipidomics strategies are largely driven by the advancement of the mass spectrometry (MS) techniques, which have led to two commonly used methodologies: shotgun lipidomics[11,12] and liquid chromatography (LC)–MS-based lipidomics.[8,13,14] However, one major challenge for lipidomics studies remains: the narrow mass-to-charge (m/z) window (m/z 600–900) in which lipids are observed leads to a large number of lipid ions having the same (i.e., isobaric) mass, which obscures the definitive identification of many lipids. This challenge can be addressed by ion mobility-mass spectrometry (IM-MS),[15−19] which provides an orthogonal dimension of separation on the basis of gas-phase structure and aids in the unambiguous identification of isobaric lipid species.[20−25] The separation of ions in the low-field IM experiment is based on differences in the ion-neutral collision cross section (CCS) as the ions drift through a neutral background gas, commonly helium or nitrogen, under the influence of a static electric field (as in drift tube IM, DTIM)[21,26,27] or a dynamic electric field (as in traveling wave IM, TWIM).[28,29] When coupled with MS, a two-dimensional separation is achieved, which provides important molecular information due to the relationship between mass (as indicated by the mass-to-charge, m/z) and CCS (as indicated by td) in the form of structural density.[15−18] As a result, different classes of biological molecules can be separated by IM-MS based on differences in their gas phase packing efficiencies, where densely packed biomolecules occur with a smaller CCS (shorter td) than loosely packed species of a similar mass. As such, each class of biomolecule occupies a discrete region of IM-MS space with lipids on average occupying the largest CCSs, followed by peptides, carbohydrates, and oligonucleotides.[21−23] Structural separations may also be observed within a biomolecular class, such as the separation of sphingolipids and glycerophospholipids subspecies within the lipid class.[21,24,26,30,31] Because of the wide availability of commercial TWIM instruments, CCS measurement on the TWIM platform is of high interest. While CCS values can be directly determined from DTIM analysis using the Mason–Schamp equation,[32,33] a precise relationship between CCS and td in TWIM has not yet been developed, although some progress has been made toward understanding the theory of TWIM.[34−36] Thus, in order to determine CCS values using TWIM, it is necessary to determine a calibration relationship between ions with known CCS values obtained on DTIM instrumentation and the td measured in TWIMS. Commonly, a power regression analysis is used for calibrating TWIM data to CCS as it has been observed that CCS and td follow an empirical power function relationship in TWIM,[30,37−42] although both linear[39,43] and binomial regressions have also been utilized for TWIM calibration purposes.[41] Major efforts have been devoted to developing TWIM calibration strategies for the analysis of peptides and proteins, such as the use of tryptic peptides,[30,39] denatured or native proteins,[37,38,40,44] and polyalanines (PolyAla).[41] Small-molecule calibrants for masses of up to 609 Da have also been reported.[45] In recent years, PolyAla have become the most widely used calibrant series due to their long-term stability, even distribution of ions over a wide range of m/z and CCS, and ability to form several charge states.[41] Calibration of TWIM CCSs using PolyAla has been reported in the literature for use with a range of molecules, including small metabolites and lipids in both positive and negative modes.[24,42,46,47] However, it has been found on several occasions that TWIM calibration using ions that are of different physical properties could lead to large errors in calibrated CCSs. For example, calibration of lipids using tryptic peptides led to errors up to 6.4%.[30] In addition, Bush et al. found that for CCS measurements of native proteins, native protein calibrants alone performed better than a combination of native and denatured proteins or denatured proteins alone.[44] These errors are the result of commonly observed differences between molecular classes: (i) there is a mismatch in the m/z range in which the calibrants and the analytes occur as observed in Bush et al.[44] and (ii) the calibrant and analyte experience different forces in the TWIM separation, such as the ion–dipole interactions due to polarizability of the N2 drift gas, which cause their drift times to scale differently with CCS.[48−50] To avoid errors in calibrated CCSs due to the above and other sources, calibration using ions with similar physical properties is desirable in order to achieve the highest CCS accuracy. Because of the large structural differences between the lipid ions and those of other classes of molecules in the gas phase,[21−23] designated lipid CCS calibrants are needed for TWIM analysis. In this work, we report the direct CCS measurements of a series of phosphatidylcholines (PCs) and phosphatidylethanolamines (PEs) (ranging from m/z 400 to 1000) in nitrogen using DTIM in both positive and negative-ion modes specifically for use as lipid CCS calibrants in other IM instrumentation. The PE and PC series were evaluated against different classes of calibrants, including PolyAla, tetraalkylammonium salts (TAA), and hexakis(fluoroalkoxy)phosphazines (HFAP), in both positive and negative modes to determine their accuracy and reproducibility for calibration of lipid CCSs from TWIM analysis. We demonstrate that structurally mismatched calibrants lead to larger error in calibrated CCS values and establish two sets of structurally matched lipid calibrants, PCs and PEs, for generating highly accurate CCS values of lipids in TWIM analysis. To our knowledge, this work represents the first systematic comparison of the effect of several structurally distinct classes of calibrants on TWIM CCS measurement, and the CCS of many of the lipid species presented here have not yet been measured by DTIM, particularly in the context of both positive and negative ion forms.

Experimental Section

Materials

A total of 10 phosphatidylcholine (PC) and 14 phosphatidylethanolamine (PE) lipid standards (Avanti Polar Lipids) with fatty acid tails ranging from 6 to 24 carbons were prepared as separate samples for analysis in positive and negative modes (see Table S1 for a list of lipid species and Avanti Polar Lipid catalog numbers). For positive mode analysis, mixtures of PCs and PEs were prepared at 5–10 μM (concentration of each species noted in Table S1) in methanol (Sigma-Aldrich) with 0.1% formic acid (Fisher Scientific). For negative mode analysis, mixtures of PEs and PCs were prepared at 5–10 μM (concentration of each species noted in Table S1) in methanol with 50 μM sodium hydroxide (Fisher Scientific) and 150 μM ammonium acetate (Fisher Scientific), respectively. Poly-dl-alanine (Sigma-Aldrich) was prepared at 25 μg/mL in 1:1 acetonitrile/water (Fisher Scientific) with 0.1% formic acid for positive mode analysis and without formic acid for negative mode analysis. A mixture of tetraalkylammonium bromide salts (TAA3, 5, 8 and 10, Acros Organics; TAA4, 6, 7, and 12, Sigma-Aldrich) was prepared at 1 ng/mL in methanol. A mixture of hexakis(fluoroalkoxy)phosphazines (HFAP, ESI-L low concentration tuning mixture, Agilent Technologies, Santa Clara, CA) was prepared by 1:4 dilution with 95% acetonitrile in water. Sphingomyelin (SM, porcine brain), phosphatidyl serines (PS, porcine brain), PE (porcine brain), and PC (chicken egg) mixtures (Avanti Polar Lipids) were prepared as 10 μM solutions in 95% acetonitrile in water.

DTIM CCS Measurements

CCS measurements for PolyAla, PEs, PCs, and HFAP were performed in positive and negative ESI modes using an Agilent 6560 IM-MS with nitrogen drift gas as described previously.[26] The IM spectrometer portion of the instrument was operated under conditions which optimize the IM resolving power between 50 and 60 (t/Δt).[51] PE and PC lipids were detected as protonated and sodium coordinated adducts in the positive mode analysis, whereas [M – H]− and [M + CH3COO]− species were observed in negative mode for PE and PC lipids, respectively. CCS values were determined from the Mason–Schamp equation as described previously.[26] For all ions, the CCS measurement precision was observed to be less than 1% RSD (see Tables S1–S3 for RSDs for each ion individually), and mass measurement accuracy was ≤5 ppm. Negative mode measurements were performed on 2 separate days, and the reported CCS values represent the weighted average of the two experiments (n = 9 per day).

TWIM Analysis and CCS Calibration

TWIM experiments were performed over 3 days, with three data acquisitions (positive mode, 1 min; negative mode, 2 min) per day for a total of nine data points. Samples were directly infused into the ESI source of a Waters Synapt G2-Si HDMS (Waters Corp., Milford, MA) at 10 μL/min for positive mode and 30 μL/min for negative mode. Data was acquired over m/z 50–1200 (to 1400 for HFAPs) in resolution mode (m/Δm = 20 000) with 550 m/s TWIM wave velocity and 40 V wave height (see the Supporting Information for source and IM parameters). Additional experiments for validation were performed at varied wave heights and velocities (30, 35, 40 V and 450, 500, 550 m/s, respectively) and data was acquired over m/z 50–5000 to prevent roll over into the next IM cycle at lower wave heights. Additionally, data was acquired using a ramped wave velocity of 900–300 m/s (tuned to prevent roll over into the next IM cycle) and 40 V over m/z 50 to 1200 (to 1400 for HFAPs). For the analysis of lipid extracts, CCS calibration was performed against the PC sodium adducts using a 500 m/s wave velocity and 40 V wave height. Drift times were obtained for each analyte by generating an extracted ion chromatogram (XIC) from the arrival time distribution function in MassLynx v4.1 using the monoisotopic mass and a mass window of ±0.075 Da. XICs were smoothed using the mean method (1 smooth; window size, ±1 scan) prior to reading the drift time from the peak apex. TWIM drift times (td, ms) and DTIM CCS (Ω, Å2) values were corrected as described previously.[38] The calibration curves were generated in GraphPad Prism 5 by fitting the corrected drift time, td, and the corrected CCS, Ω′, to an equation of the formwhere A′, t0, and B are fit parameters and z is the ion charge state. This equation was used to account for ion flight time in areas outside the IM cell, t0, and this method has been demonstrated to result in better correlation to the regression model and more accurate CCS values than the simple power regression model (without the t0 term).[42] In both positive and negative modes, the three intraday replicates were treated independently during the CCS calibration process (i.e., the number of points analyzed in negative mode for PEs and PCs were 42 and 30, respectively). Both species (+ H and + Na) observed in positive mode were similarly treated independently, such that the total number of points analyzed for PEs and PCs in positive mode were 85 and 60, respectively. Reported CCSs represent the average of nine experiments and interday RSDs were below 0.5% for all TWIM CCSs (Tables S4–S17).

Results and Discussion

CCS Calibrants

The primary factors taken into consideration for choosing CCS calibrants to evaluate for use with phospholipids were: (i) ability to ionize in both positive and negative modes and/or (ii) structural similarity to phospholipids. Figure a summarizes the four types of calibrants evaluated: polyalanine (PolyAla, n = 12(+)/14(−)),[41] tetralkylammonium (TAA, n = 8) salts,[26,45] hexakis(fluoroalkoxy)phosphazines (HFAP, n = 4), and glycerophospholipids (PC, n = 10; PE, n = 14) (see Figure S1 for the CCS plot in negative mode). All the CCS values of the calibrants were measured on the same DTIM platform to achieve a fair comparison between the different calibrants (see Experimental Section). Collectively, the calibrants cover a range of m/z and CCS from approximately 200–1250 m/z and 125–350 Å2, respectively, in both positive and negative modes (Figures b and S1; see Tables S1–S3 for DTIM CCS values). The PC, PolyAla, and HFAP calibrants were evaluated in both positive and negative ionization modes, whereas the TAA salts were analyzed in positive mode only, as TAA does not readily generate an anion form. Although larger analytes (m/z ≥ 2000) are present in the HFAP mix, only the four ions covering the range m/z 300–1250 (+ESI) were used for CCS calibration purposes. Notably, the nonlipid calibrants exhibit both larger and smaller CCS values than those of the lipids within the same mass range. The conformational distribution of the calibrants in the CCS vs m/z 2D plot (Figure and Figure S1) suggests that the molecular packing efficiencies, or gas-phase densities, of these ions in the gas phase increase in the order of TAA ≪ PCPE < PolyAlaHFAP in the positive mode and PC < PE < PolyAlaHFAP in the negative mode. These trends can be reasonably understood based on their structures: each TAA contains four flexible hydrocarbon chains; each phospholipid contains two flexible fatty ester chains; each PolyAla is comprised of the same monomer (i.e., alanine) and can potentially form compact structures via intramolecular hydrogen bonding; and each HFAP is made of a rigid aromatic ring and heavier atoms such as fluorine and phosphorus that greatly increase the density of the structure relative to other calibrants of similar m/z.
Figure 1

(A) Structures of select or exemplary species in each class of TWIM CCS calibrant (from left to right): polyalanine (PolyAla; trialanine shown); hexakis(fluoroalkoxy)phospazines (HFAP; hexakis(2,2-difluoroethoxy)phosphazine shown); phosphatidylethanolamine (PE; PE 4:0/4:0 shown); phosphatidylcholine (PC; PC 4:0/4:0 shown); tetralkylammonium salts (TAA; tetrapropylammonium shown). (B) IM-MS conformational space plot showing the trends in CCS-m/z for each of the five calibrants from + ESI DTIM-MS measurements in N2. For clarity, only the protonated adducts of the PE and PC calibrants are shown.

(A) Structures of select or exemplary species in each class of TWIM CCS calibrant (from left to right): polyalanine (PolyAla; trialanine shown); hexakis(fluoroalkoxy)phospazines (HFAP; hexakis(2,2-difluoroethoxy)phosphazine shown); phosphatidylethanolamine (PE; PE 4:0/4:0 shown); phosphatidylcholine (PC; PC 4:0/4:0 shown); tetralkylammonium salts (TAA; tetrapropylammonium shown). (B) IM-MS conformational space plot showing the trends in CCS-m/z for each of the five calibrants from + ESI DTIM-MS measurements in N2. For clarity, only the protonated adducts of the PE and PC calibrants are shown. In positive mode, the majority of the CCS calibrants demonstrate strong power relationships between CCS and m/z, with R2 > 0.99 for TAA, PolyAla, PCs (+ H and + Na), and HFAPs when fit to the equation y = ax.[26] The TAA salts display the largest exponent value being 0.625, followed by PCs (0.575), PEs (0.529), PolyAla (0.496), and HFAPs (0.451). The PEs had the poorest fit (R2 = 0.988) due to the number of PEs with one or two unsaturated fatty acids (n = 5) versus those with fully saturated fatty acids (n = 9). Excluding PEs with unsaturated fatty acids improved the fit of the PE mobility-mass correlation for both the protonated and sodium coordinated species (R2 = 0.9995), while the unsaturated PEs alone have an R2 = 0.806 (Figure S2). In negative mode, linear regression analysis provided a better fit to the data (all R2 > 0.99) than the power fitting for all calibrants (HFAPs, PolyAla, PCs, PEs), suggesting that the relationship between CCS and m/z is more linear in negative mode.

Comparison of Different Calibrants on Lipid CCS Measurement

A principal goal of this work is to evaluate the cross-calibration of each class of phospholipids (i.e., PCs to calibrate PEs and PEs to calibrate PCs), in comparison with other classes of calibrants: PolyAla, HFAP, and TAA. For the TWIM analysis, we chose to carry out initial studies at a wave velocity of 550 m/s and wave height of 40 V since these conditions were observed to give maximum separation of the lipid ions without carrying over the ions to the next IM cycle. Figure contains representative TWIM CCS calibration curves for PC, PE, PolyAla, and HFAP from both positive (open gray circles) and negative (filled black squares) mode analyses, which were obtained using eq . For all analytes investigated, the resulting calibration curves yielded regression values greater than 0.995 with averaged calibration fit errors from 0.1 to 0.5% (Tables S4–S9). The most notable difference between the calibration curves for the different calibrants is in the degree of curvature in the fit. For example, the calibration curves for PolyAla (Figure c) and HFAP (Figure d), as well as TAA (Figure S3), are slightly curved. However, calibration curves for PE and PC are distinctly more linear as indicated by the values of exponent B in eq being closer to one. Comparison of the fit parameter for the phospholipids relative to the other calibrants shows that PEs and PCs yielded smaller values for fit parameter A′ and positive values greater than 1 for t0, whereas the other calibrants had larger values of A′ and negative values below 1 for t0. These differences in fitting parameters provide further evidence that lipid-specific calibrants are needed in order to achieve the most accurate measurement of lipid CCS using TWIM.
Figure 2

TWIM CCS calibration plots for A) PC, B) PE, C) PolyAla, and (D) HFAP CCS calibrants, where Ω′ is the corrected drift tube CCS and td′ is the mass-independent drift time. Calibration was performed with wave settings of 40 V and 550 m/s. Plots from positive (open gray circles) and negative (filled black squares) mode analyses are overlaid. The calibration equations for each mode are also shown. All calibration fits had R2 ≥ 0.995.

TWIM CCS calibration plots for A) PC, B) PE, C) PolyAla, and (D) HFAP CCS calibrants, where Ω′ is the corrected drift tube CCS and td′ is the mass-independent drift time. Calibration was performed with wave settings of 40 V and 550 m/s. Plots from positive (open gray circles) and negative (filled black squares) mode analyses are overlaid. The calibration equations for each mode are also shown. All calibration fits had R2 ≥ 0.995. The effects of using calibration curves such as those for PolyAla, TAA, HFAP, and the phospholipids to calculate CCS values for PEs and PCs can be seen in Figure (detailed calibration errors in Tables S10–S17), which presents the calibrated CCSs and the ±2% CCS bands (black dashed lines) based on the DTIM CCSs. For positive mode analysis of PCs (Figure a), calibration with PolyAla and TAAs yield some cross sections that deviate significantly (≥2%) from CCS values measured on the DTIM instrumentation, notably within the intermediate mass range. A similar trend is observed for PCs in negative mode (Figure b), where PolyAla yields calibrated CCSs that have errors greater than +2% CCS for intermediate masses. Calibration of PCs with HFAP and PEs provided the most accurate results with errors within ±2% CCS in both positive and negative modes.
Figure 3

Comparison of the accuracies of TWIM CCS calibrants for PCs (A and B) and PEs (C and D) at 550 m/s and 40 V. Each symbol represents averaged (n = 9) calibrated CCS values from a different calibrant. The dashed lines represent ±2% error from the drift tube (DT) CCS values (black circles), fit to power functions in parts A and C and linear functions in parts B and D. Calibrated CCSs for PE 20:4/20:4 are shown as open markers in parts C and D as the DTIM CCS of PE 20:4/20:4 was excluded from the fit of the ±2% error lines. For clarity, only protonated adducts are shown in parts A and C.

Comparison of the accuracies of TWIM CCS calibrants for PCs (A and B) and PEs (C and D) at 550 m/s and 40 V. Each symbol represents averaged (n = 9) calibrated CCS values from a different calibrant. The dashed lines represent ±2% error from the drift tube (DT) CCS values (black circles), fit to power functions in parts A and C and linear functions in parts B and D. Calibrated CCSs for PE 20:4/20:4 are shown as open markers in parts C and D as the DTIM CCS of PE 20:4/20:4 was excluded from the fit of the ±2% error lines. For clarity, only protonated adducts are shown in parts A and C. For positive mode analysis of PEs (Figure c), PolyAla and TAA calibrated CCS values increasingly deviate from the DTIM CCSs as the PE species increase in mass and size. The PolyAla calibration errors are predominantly within the +2% CCS band, but calibration errors for TAA-calibrated PEs exceed +2% starting with PE 10:0/10:0 at m/z 524.3. Alternatively, calibration of PEs with HFAP yields CCSs with negative error in the low mass region and positive error in the high mass region. The use of PCs for CCS calibration of PEs works well in positive mode, with all calibrated CCSs within ±2% of the DTIM CCSs. Calibration of PEs with PolyAla in negative mode (Figure d) has a trend similar to that observed in positive mode; however, the errors are greater than 2% CCS for all PE species. As with the calibrated PC results, the HFAP-calibrated PE CCS values in negative mode are well within 2% error from the DTIM values. Calibration of PEs with PCs in negative mode also yielded favorable results, with only PE 6:0/6:0 and 8:0/8:0 displaying errors great than 2%. Overall, at wave velocity of 550 m/s and wave height of 40 V, PCs and PEs performed the best as the lipid CCS calibrants (see Table S4 for a summary of precision and accuracy of the calibrants). In positive mode, average calibration errors comparable to the fit errors (i.e., 0.1–0.5% error) observed from the PC and PE calibration curves themselves. Average errors were higher in negative mode but remained below 1% error. Calibration with HFAP also provided accurate CCSs for PEs and PCs in both ionization modes, with average errors ≤1%.

Effect of Wave Velocity and Wave Height on Lipid CCS Calibration

To evaluate the performance of PEs, PCs, and HFAP as lipid CCS calibrants under different traveling wave parameters, the above experiments were repeated using varied combinations of wave height (40, 35, and 30 V) and wave velocity (550, 500, and 450 m/s). Calibration of PEs and PCs with HFAP yielded the largest average percent errors in positive mode (Tables S18 and S20) due to dependencies on both mass and traveling wave parameters. The error of the CCS calibration decreased with decreasing wave height for both PEs and PCs (Figures S4 and S6), while calibration error increased with decreasing wave velocity (i.e., lowest errors at 550 m/s and 30 V). The effects of traveling wave parameters on calibration error were small in negative mode, where errors across all species and combinations of wave height and velocity were within ±2% CCS. For HFAP-calibration of PEs, the errors were relatively consistent over the mass range with the exception of PE 20:4/20:4, which showed greater average calibration errors in positive and negative modes. HFAP-calibrated PC species, on the other hand, exhibited larger errors in the middle of the calibration range which was most pronounced in positive mode, where the absolute error ranged from −0.5% for PC 6:0/6:0 to +2.5% for the intermediate PC species (i.e., PC 14:0/14:0–18:0/18:0). Trends were also observed in positive and negative modes for CCS calibration with phospholipids at varied wave heights and wave velocities. Calibration of PCs with PEs showed clear bias toward smaller PCs in positive mode, with consistently small calibration errors up to PC 18:0/18:0 (<0.5%), and larger errors for PC 20:0/20:0 to 24:0/24:0 (−0.5% to −2.5%) (Figure A and Figure S7). The reverse effect was observed for CCS calibration of PEs with PCs in negative mode, where calibration errors were +0.5% to +3.5% for PEs 6:0/6:0 to 12:0/12:0 over all combinations of wave height and velocity (Figure D and Figure S5). Calibration errors for PC-calibrated PE CCSs in positive mode and PE-calibrated PC CCSs in negative mode were consistently low across all TWIM parameters with only PC 24:0/24:0 displaying errors greater than 1% (Figure B,C and Figures S5 and S7). In general, the calibration bias appears to become more evident as the wave velocity decreases, i.e., CCSs with large calibration errors tend to display even larger errors at smaller wave velocity while the CCSs with smaller errors are relatively consistent at all wave height and velocity (Figure and Figures S5 and S7). These observations may be due to the fact that in TWIM, under the conditions of lower wave heights and higher wave velocities, the ions experience more structurally selective roll-over events which better describe differences in their gas-phase structure,[36] and the corresponding longer drift times would also serve to reduce the magnitude of time measurement errors, such as bin averaging and IM peak centroiding.
Figure 4

Effects of traveling wave velocity on CCS calibration accuracy: (A) calibration of protonated PCs with PE CCSs in positive mode, (B) calibration of PCs with PE CCS in negative mode, (C) calibration of protonated PEs with PC CCSs in positive mode, (D) calibration of PEs with PC CCSs in negative mode. A traveling wave height of 40 V was used for all measurements.

Effects of traveling wave velocity on CCS calibration accuracy: (A) calibration of protonated PCs with PE CCSs in positive mode, (B) calibration of PCs with PE CCS in negative mode, (C) calibration of protonated PEs with PC CCSs in positive mode, (D) calibration of PEs with PC CCSs in negative mode. A traveling wave height of 40 V was used for all measurements. Overall, PCs and PEs outperformed HFAP as the calibrants for lipids even though all three CCS calibrants displayed some calibration errors that were dependent upon mass, size, or traveling wave parameters. Additional experiments performed with a ramped wave velocity also demonstrate the inconsistency of HFAP as CCS calibrants. Using a linearly ramped wave velocity range of 900–300 m/s and a fixed wave height of 40 V, the performance of PE and PC calibrants was consistent with results obtained under static wave velocities whereas the performance of HFAP diminished greatly when using the ramped wave velocity (Tables S22–S25). The inconsistencies in the performance of HFAP as generalized CCS calibrants for TWIM is likely the result of the low number of HFAP species (n = 4) used in these experiments. Although the HFAP ions chosen adequately bracket the mass range typically observed for lipids, the limited number of HFAP calibrants within the lipid m/z range was not sufficient to provide robust and accurate calibration of lipid CCSs. While the use of phospholipids as TWIM CCS calibrants yielded the most accurate calibrated CCSs for both PEs and PCs, the mass-dependent calibration errors observed for PE-calibrated PC CCSs in positive mode and PC-calibrated PE CCSs in negative mode merit further discussion (Figure A,D). For PC-calibrated PE CCSs in negative mode, the increase in calibration errors with decreasing PE mass can likely be attributed to the different negative ion species formed between PEs and PC lipids, specifically deprotonated and acetate-adducted, respectively. While the PC headgroup is 44 Da larger than the PE headgroup for the neutral species, this mass difference becomes even larger when analyzing the negative ions due to the need to form adducts for negative mode ionization of PC species. The formation of acetate adducts (+ 59 Da) was chosen for this analysis due to the frequent use of ammonium acetate as a buffer in LC solvents, particularly for reverse phase and HILIC methods. However, the formation of acetate adducts led to larger proportion of structural changes to the PCs with shorter acyl chains (10:0/10:0 or smaller) than those with longer acyl chains. As a result, acetate adduction decreases the structural similarity of PC negative ions relative to PE negative ions with similar fatty acid compositions. It has been reported that different metal ion adducts of PCs significantly affect their CCS mobility-mass correlations in the positive mode, with the bigger metal ions displaying a larger effect.[21] On the other hand, the increased calibration errors observed in positive mode for PE-calibrated PC CCSs with acyl chains of 20 carbons or greater likely results from the lack of PE calibrant species above 800 Da. Unsaturation in the acyl chains was found to affect the CCS and the calibration accuracy. CCS measurements of PEs possessing the same number of carbons and increasing degrees of unsaturation (i.e., PE 18:0/18:0, 18:0/18:1, and 18:1/18:1 in Table S1) demonstrate that the addition of double bonds leads to a reduction in the CCS. Kim et al. have reported similar observations for the addition of double bonds to the acyl chains of PCs, where the first double bond led to a 5% reduction in drift time with a further 1% reduction for each additional double bond.[39] The reduction in CCS observed for additional double bonds in PEs was 0.3 and 0.9% for acyl chains lengths of 18 and 16, respectively, but no bias was observed in the CCS calibration of saturated versus singly- or doubly unsaturated PEs. The single highly unsaturated PE evaluated in this study, PE 20:4, deviated significantly (approximately 6%) from the PE and PC trendline in the CCS vs. m/z plot (Figure b) and tended to have greater calibration errors than PEs with fewer or no double bonds (Tables S5 and S11), although all errors were within ±2% when calibration was performed with PCs. Thus, a set of highly unsaturated phospholipid CCS calibrants may be of benefit if TWIM CCS information for highly unsaturated phospholipids is desired.

Validation of the Performance of Lipid CCS Calibrants

To evaluate the performance of the lipid CCS calibrants in an untargeted lipidomics experiment, TWIM CCSs were obtained for the components in SM, PS, PE, and PC extracts and compared against their DTIM CCS values.[26,52] The results from triplicate measurements of the most abundant SMs and PSs (n = 5 each) are shown in Table (additional results in Table S26). Notably, the absolute errors between DTIM and TWIM CCSs were below 2% with the exception of only two lipid species (PS 38:01 and PC 36:1) at 2.2%. The outstanding performance of these lipid calibrants on lipid classes other than PE and PC suggest that these calibrants may have broad application in measurement of CCS of polar lipids using TWIM.
Table 1

Validation of Lipid CCS Calibration against DTIM Measurements of SM and PS Extracts

identificationam/z observedTWIM CCS (Å2) [adduct]DTIM CCS (Å2) [adduct]CCS error (%)RSD (%) TWIM CCS
SM 34:01703.58278.7 [H]280.1 [H]–0.50.2
SM 36:01731.61284.7 [H]286.2 [H]–0.50.1
SM 40:02785.65297.6 [H]295.7 [H]0.60.1
SM 42:02813.69304.0 [H]301.4 [H]0.90.0
SM 42:01815.70305.2 [H]303.2 [H]0.60.0
PS 38:04812.54281.3 [H]284.4 [H]–1.10.2
PS 38:02816.58282.6 [H]287.7 [H]–1.80.4
PS 38:01818.59283.3 [H]289.6 [H]–2.20.5
PS 40:06836.55284.3 [H]287.7 [H]–1.20.2
PS 40:05838.56286.1 [H]289.6 [H]–1.20.2

All identifications made within 10 ppm of the observed m/z.

All identifications made within 10 ppm of the observed m/z.

Conclusions

While several sets of calibrants have been proposed for generating calibrated CCSs from TWIM-MS platforms, we have found that the use of structurally matched calibrants provides greater accuracy of calibrated CCS for phospholipids than PolyAla and TAAs and greater consistency over multiple traveling wave parameters than HFAPs. On the basis of the observations described above, the best implementation of the proposed phospholipid CCS calibrants was determined to be the use of PCs for CCS calibration of phospholipids in positive mode and the use of PEs for CCS calibration of phospholipids in negative mode. We believe this work will equally benefit those laboratories who may choose to use PolyAla for its ease-of-use and cost-effectiveness, in that the inherent errors of using PolyAla to calibrate lipid CCSs are demonstrated here for the first time. Although the cost of purchasing individual standards is greater than that of premixed calibrants, the expense is justifiable given the improved accuracy of calibrated CCSs provided by the lipid calibrants relative to PolyAla or HFAP. The development of phospholipid CCS calibrants will enable high-accuracy structural studies of lipids and add an additional level of validation in the assignment of identifications in untargeted lipidomics experiments.
  44 in total

Review 1.  Shotgun lipidomics: electrospray ionization mass spectrometric analysis and quantitation of cellular lipidomes directly from crude extracts of biological samples.

Authors:  Xianlin Han; Richard W Gross
Journal:  Mass Spectrom Rev       Date:  2005 May-Jun       Impact factor: 10.946

2.  Structural characterization of drug-like compounds by ion mobility mass spectrometry: comparison of theoretical and experimentally derived nitrogen collision cross sections.

Authors:  Iain Campuzano; Matthew F Bush; Carol V Robinson; Claire Beaumont; Keith Richardson; Hyungjun Kim; Hugh I Kim
Journal:  Anal Chem       Date:  2011-12-27       Impact factor: 6.986

Review 3.  Ion mobility-mass spectrometry.

Authors:  Abu B Kanu; Prabha Dwivedi; Maggie Tam; Laura Matz; Herbert H Hill
Journal:  J Mass Spectrom       Date:  2008-01       Impact factor: 1.982

4.  Ion mobility-mass spectrometry analysis of large protein complexes.

Authors:  Brandon T Ruotolo; Justin L P Benesch; Alan M Sandercock; Suk-Joon Hyung; Carol V Robinson
Journal:  Nat Protoc       Date:  2008-06-19       Impact factor: 13.491

5.  The Influence of Drift Gas Composition on the Separation Mechanism in Traveling Wave Ion Mobility Spectrometry: Insight from Electrodynamic Simulations.

Authors:  Jody C May; John A McLean
Journal:  Int J Ion Mobil Spectrom       Date:  2003-06-01

Review 6.  Lipid regulation of cell membrane structure and function.

Authors:  P L Yeagle
Journal:  FASEB J       Date:  1989-05       Impact factor: 5.191

7.  Sphingolipidomics: high-throughput, structure-specific, and quantitative analysis of sphingolipids by liquid chromatography tandem mass spectrometry.

Authors:  Alfred H Merrill; M Cameron Sullards; Jeremy C Allegood; Samuel Kelly; Elaine Wang
Journal:  Methods       Date:  2005-06       Impact factor: 3.608

8.  Structural characterization of unsaturated phosphatidylcholines using traveling wave ion mobility spectrometry.

Authors:  Hugh I Kim; Hyungjun Kim; Eric S Pang; Ernest K Ryu; Luther W Beegle; Joseph A Loo; William A Goddard; Isik Kanik
Journal:  Anal Chem       Date:  2009-10-15       Impact factor: 6.986

9.  Characterizing ion mobility-mass spectrometry conformation space for the analysis of complex biological samples.

Authors:  Larissa S Fenn; Michal Kliman; Ablatt Mahsut; Sophie R Zhao; John A McLean
Journal:  Anal Bioanal Chem       Date:  2009-02-27       Impact factor: 4.142

10.  Ion mobility derived collision cross sections to support metabolomics applications.

Authors:  Giuseppe Paglia; Jonathan P Williams; Lochana Menikarachchi; J Will Thompson; Richard Tyldesley-Worster; Skarphédinn Halldórsson; Ottar Rolfsson; Arthur Moseley; David Grant; James Langridge; Bernhard O Palsson; Giuseppe Astarita
Journal:  Anal Chem       Date:  2014-03-28       Impact factor: 6.986

View more
  35 in total

1.  New frontiers for mass spectrometry based upon structures for lossless ion manipulations.

Authors:  Yehia M Ibrahim; Ahmed M Hamid; Liulin Deng; Sandilya V B Garimella; Ian K Webb; Erin S Baker; Richard D Smith
Journal:  Analyst       Date:  2017-03-27       Impact factor: 4.616

2.  Characterization of the Impact of Drug Metabolism on the Gas-Phase Structures of Drugs Using Ion Mobility-Mass Spectrometry.

Authors:  Dylan H Ross; Ryan P Seguin; Libin Xu
Journal:  Anal Chem       Date:  2019-10-29       Impact factor: 6.986

3.  Ion Mobility Spectrometry-Mass Spectrometry Coupled with Gas-Phase Hydrogen/Deuterium Exchange for Metabolomics Analyses.

Authors:  Hossein Maleki; Ahmad K Karanji; Sandra Majuta; Megan M Maurer; Stephen J Valentine
Journal:  J Am Soc Mass Spectrom       Date:  2017-09-27       Impact factor: 3.109

4.  ISiCLE: A Quantum Chemistry Pipeline for Establishing in Silico Collision Cross Section Libraries.

Authors:  Sean M Colby; Dennis G Thomas; Jamie R Nuñez; Douglas J Baxter; Kurt R Glaesemann; Joseph M Brown; Meg A Pirrung; Niranjan Govind; Justin G Teeguarden; Thomas O Metz; Ryan S Renslow
Journal:  Anal Chem       Date:  2019-03-06       Impact factor: 6.986

5.  Structural Characterization of Methylenedianiline Regioisomers by Ion Mobility-Mass Spectrometry, Tandem Mass Spectrometry, and Computational Strategies. 3. MALDI Spectra of 2-Ring Isomers.

Authors:  Sarah M Stow; Tiffany M Crescentini; Jay G Forsythe; Jody C May; John A McLean; David M Hercules
Journal:  Anal Chem       Date:  2017-08-30       Impact factor: 6.986

6.  Conditions for Analysis of Native Protein Structures Using Uniform Field Drift Tube Ion Mobility Mass Spectrometry and Characterization of Stable Calibrants for TWIM-MS.

Authors:  Julian A Harrison; Celine Kelso; Tara L Pukala; Jennifer L Beck
Journal:  J Am Soc Mass Spectrom       Date:  2018-10-15       Impact factor: 3.109

7.  High-Precision, Gas-Phase Hydrogen/Deuterium-Exchange Kinetics by Mass Spectrometry Enabled by Exchange Standards.

Authors:  Sanjit S Uppal; Abhigya Mookherjee; Rick Harkewicz; Sarah E Beasley; Matthew F Bush; Miklos Guttman
Journal:  Anal Chem       Date:  2020-05-18       Impact factor: 6.986

8.  Evaluating Separation Selectivity and Collision Cross Section Measurement Reproducibility in Helium, Nitrogen, Argon, and Carbon Dioxide Drift Gases for Drift Tube Ion Mobility-Mass Spectrometry.

Authors:  Caleb B Morris; Jody C May; Katrina L Leaptrot; John A McLean
Journal:  J Am Soc Mass Spectrom       Date:  2019-03-18       Impact factor: 3.109

Review 9.  Challenges in Identifying the Dark Molecules of Life.

Authors:  María Eugenia Monge; James N Dodds; Erin S Baker; Arthur S Edison; Facundo M Fernández
Journal:  Annu Rev Anal Chem (Palo Alto Calif)       Date:  2019-03-18       Impact factor: 10.745

10.  Application of Group I Metal Adduction to the Separation of Steroids by Traveling Wave Ion Mobility Spectrometry.

Authors:  Alana L Rister; Tiana L Martin; Eric D Dodds
Journal:  J Am Soc Mass Spectrom       Date:  2018-11-09       Impact factor: 3.109

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.