Pulse Oximeter-Derived Pleth Variability Index is a Reliable Indicator of Cardiac Preload in Patients Undergoing Liver Transplantation.

H-C Lee1, Y-F Tsai1, H-I Tsai1, P C-H Chung2, H-P Yu1, W-C Lee3, C-C Lin4.   

Abstract

BACKGROUND: Accurate estimation of cardiac preload during liver transplantation is essential. The right ventricular end-diastolic volume index (RVEDVI) is recognized as a good preload indicator in patients undergoing liver transplantation. Recently, dynamic variation parameters including pleth variability index (PVI) have been used as predictors of fluid responsiveness. However, the correlation between PVI and preload status has not been well studied. We evaluated the relationship between PVI and RVEDVI during liver transplantation.
METHODS: Eighteen patients undergoing liver transplantation were enrolled in this study. Data of hemodynamic parameters including PVI derived by Masimo Rainbow SET Pulse CO-Oximeter, central venous pressure (CVP), pulmonary arterial occlusion pressure (PAOP), and RVEDI were obtained at 10 defined time points throughout liver transplantation. The correlation between RVEDVI and CVP, PAOP, and PVI was analyzed using Spearman rank test. We also investigated the ability of PVI to accurately differentiate RVEDVI <123 or >142 mL/m(2) using receiver operating characteristic (ROC) analysis.
RESULTS: There was fair to good correlation between PVI and RVEDVI (correlation coefficient = -0.492, P < .001). The correlation coefficient between CVP, PAOP, and RVEDVI was 0.345 and 0.463, respectively. A 13.5% cutoff value of PVI estimated the RVEDVI <123 mL/m(2) (area under the curve [AUC] = 0.762). A 12.5% cutoff value of PVI estimated the RVEDVI >142 mL/m(2) (AUC = 0.745).
CONCLUSIONS: PVI presented as a reliable estimate of preload status and may be a useful predictor of fluid responsiveness in patients undergoing liver transplantation.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27320555     DOI: 10.1016/j.transproceed.2015.12.106

Source DB:  PubMed          Journal:  Transplant Proc        ISSN: 0041-1345            Impact factor:   1.066


  1 in total

1.  Variation in Frequency of Intraoperative Arterial, Central Venous and Pulmonary Artery Catheter Placement During Kidney Transplantation: An Analysis of Invasive Monitoring Trends.

Authors:  Alexander Nagrebetsky; Richard P Dutton; Jesse M Ehrenfeld; Richard D Urman
Journal:  J Med Syst       Date:  2018-03-02       Impact factor: 4.460

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.