Claire C Bristow1,2, Linda Severe3, Jean William Pape3,4, Marjan Javanbakht5, Sung-Jae Lee5,6, Warren Scott Comulada6, Jeffrey D Klausner7,8,9.
Abstract
BACKGROUND: Dual rapid tests for HIV and syphilis infections allow for detection of HIV infection and syphilis at the point-of-care. Those tests have been evaluated in laboratory settings and show excellent performance but have not been evaluated in the field. We evaluated the field performance of the SD BIOLINE HIV/Syphilis Duo test in Port-au-Prince, Haiti using whole blood fingerprick specimens.Entities:
Keywords: Diagnosis; Point-of-care, HIV, Syphilis, Haiti, Dual test, Rapid test, Dual elimination, Haiti
Mesh:
Year: 2016 PMID: 27316352 PMCID: PMC4912739 DOI: 10.1186/s12879-016-1574-3
Source DB: PubMed Journal: BMC Infect Dis ISSN: 1471-2334 Impact factor: 3.090
Fig. 1Standard for band color intensity for SD BIOLINE HIV/Syphilis Duo dual rapid test. The color bands on this reference standard were compared to positive results on the Duo test and the percent intensity was recorded by two trained phlebotomists
Field performance of the SD BIOLINE HIV/Syphilis Duo dual rapid test in Haiti for detection of HIV antibodies using a dual HIV/syphilis test, 2014
| HIV reference test* | Total | Sensitivity (95 % CI) | Specificity (95 % CI) | Kappa Coefficient (95 % CI) | |||
|---|---|---|---|---|---|---|---|
| Pos | Neg | ||||||
| SD BIOLINE | Pos | 128 | 5 | 133 | 99.2 % (95.8 %, 100 %) | 97.0 % (93.2 %, 99.0 %) | .96 |
| Neg | 1 | 164 | 165 | (.93, .99) | |||
| Total | 129 | 169 | 298 | ||||
*Reference tests were Murex HIV-1.2.0 (DiaSorin S.p.A., Saluggia, Italy) or Alere Determine HIV (Alere) rapid testing. Positive results were confirmed with a colloidal gold test from KHB (Shanghai Kehua Bio-Engineering Co., LTD, China)
Field performance of the SD BIOLINE HIV/Syphilis Duo dual rapid test in Haiti for detection of Treponema pallidum antibodies using a dual HIV/syphilis test, 2014
|
| Total | Sensitivity (95 % CI) | Specificity (95 % CI) | Kappa Coefficient (95 % CI) | |||
|---|---|---|---|---|---|---|---|
| Pos | Neg | ||||||
| SD BIOLINE HIV/ | Pos | 109 | 17 | 126 | 96.5 % (91.2 %, 99.0 %) | 90.8 % (85.7 %, 94.6 %) | .85 |
| Neg | 4 | 168 | 172 | (.79, .91) | |||
| Total | 113 | 185 | 298 | ||||
*Reference test was conducted using Treponema Pallidum Hemaglutination Assay (TPHA) (Human Gesellschaft fur Biochemica und Diagnostica mbH, Wiesbaden, Germany). TPHA indeterminate results were retested using a Treponema pallidum enzyme-linked immunosorbent assay test (ELISA)(Architect Syphilis TP; Abbott, Wiesbaden, Germany) and were reclassified based on the ELISA result
Visual intensity of the color of the bands indicating positive results for the SD BIOLINE HIV/Syphilis Duo dual rapid test
| N | Mean (%) | Standard deviation (%) | Median (%) | Quartile range (%) | Minimum (%) | Maximum (%) | |
|---|---|---|---|---|---|---|---|
| HIV band color intensity | 133 | 88.3 | 28.0 | 100 | 0 | 0.3 | 100 |
| HIV true positives | 128 | 91.7 | 22.4 | 100 | 0 | 0.3 | 100 |
| HIV false positives | 5 | 1.3 | 2.1 | 0.3 | 0.4 | 0.3 | 5.0 |
|
| 126 | 30.3 | 31.9 | 20 | 37 | 0.3 | 100 |
|
| 109 | 34.3 | 32.4 | 23 | 40 | 0.3 | 100 |
|
| 17 | 4.8 | 7.1 | 2 | 4 | 0.3 | 26 |
*Note: Intensity was not recorded for three of the tests that were syphilis positive on the Duo test