Effects of Intraocular Lens Opacification on Light Scatter, Stray Light, and Overall Optical Quality/Performance.

Liliana Werner1, John C Stover2, Jim Schwiegerling3, Kamal K Das4.   

Abstract

PURPOSE: To evaluate light scatter and stray light in intraocular lenses (IOLs) explanted because of postoperative opacification (13 calcified hydrophilic acrylic, 1 calcified silicone, and 4 polymethylmethacrylate [PMMA] lenses with snowflake degeneration), as well as effect of opacification on other optical quality/performance indicators, in comparison with controls.
METHODS: The Complete Angle Scatter Instrument (CASI) scatterometer was used to measure the forward light scattering (FLS) of the IOLs, and the stray light values at various angles were calculated from the measured FLS. Modulation transfer function (MTF) was obtained with an optical bench, and a Badal optometer was used to obtain letter chart images through the lenses. Back light scatter and light transmittance were also measured.
RESULTS: Average stray light values (Log (s)) at a scattered angle of 10° were 1.79 ± 0.37 for hydrophilic acrylic IOLs (controls 0.36 ± 0.05), 1.53 for the silicone lens (control 0.41), and 1.62 ± 0.46 for PMMA IOLs (control 0.25). Stray light was significantly higher for explanted opacified lenses (N = 18) in comparison with controls (N = 7; two-tail P < 0.001 at 10°). Modulation transfer function and Badal image contrast were drastically reduced in lenses with calcification and snowflake degeneration.
CONCLUSIONS: Different studies described the impact of stray light in human vision, with serious hindrance above 1.47 Log (s). Lenses explanted from patients because of clinically significant opacification are associated with a considerable increase in light scatter and stray light, as well as with a decline of other optical quality/performance indicators.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27315539     DOI: 10.1167/iovs.16-19514

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  3 in total

1.  Validation of a spectral light scattering method to differentiate large from small particles in intraocular lenses.

Authors:  Grzegorz Łabuz; Eleni Papadatou; Fernando Vargas-Martín; Norberto López-Gil; Nicolaas J Reus; Thomas J T P van den Berg
Journal:  Biomed Opt Express       Date:  2017-02-27       Impact factor: 3.732

2.  Light scattering in intraocular lenses explanted 15 to 40 years after surgery.

Authors:  Jorge L Alio; Augusto Arias; Francesco D'Oria; Francesca Toto; Jorge Alio Del Barrio; Raul Duarte; Pablo Artal
Journal:  Biomed Opt Express       Date:  2021-05-18       Impact factor: 3.732

3.  Comparison of visual outcomes after bilateral implantation of a diffractive trifocal intraocular lens and blended implantation of an extended depth of focus intraocular lens with a diffractive bifocal intraocular lens.

Authors:  André Lins de Medeiros; André Gustavo de Araújo Rolim; Antonio Francisco Pimenta Motta; Bruna Vieira Ventura; César Vilar; Mário Augusto Pereira Dias Chaves; Pedro Carlos Carricondo; Wilson Takashi Hida
Journal:  Clin Ophthalmol       Date:  2017-10-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.