Spatial relationship of i-gel® and Ambu® AuraOnceTM on pediatric airway: a randomized comparison based on three dimensional magnetic resonance imaging.

Mansoor Aqil1, Bilal Delvi2, Abdullah Abujamea3, Tariq Alzahrani2, Abdulrahman Alzahem4, Saara Mansoor5, Ayman Aljazaeri4.   

Abstract

BACKGROUND: Gross morphological differences exist among different brands of pediatric supraglottic devices (SGDs). The aim of this study is to compare the spatial relationship of i-gel® and Ambu® AuraOnce (AO)TM on pediatric airway based on three dimensional (3-D) magnetic resonance imaging (MRI) measurements.
METHODS: Sixty patients up to 12 years of age were enrolled and assigned in two groups, i-gel® or Ambu® AOTM. After confirmation of proper placement of these SGDs, 3-D MRI scans of head and neck were performed. Another native scan was also obtained after removal of the SGD for comparison.
RESULTS: i-gel® produced significant degree of compression of the tongue (P<0.001) while Ambu® AOTM significantly reduced the axial diameter of glottis (P=0.033) compared to their native values. Both i-gel® and Ambu® AOTM significantly reduced the area of the glottic opening (P<0.001 for each device) and the distance between the arytenoids (P<0.001 and P=0.007 respectively); and increased the distance between the hyoid bone and cervical spine (P<0.001 and P=0.001 respectively) in comparison to their corresponding native values. Bowl of i-gel® produced greater dilation of the upper esophageal sphincter at all levels of measurement- upper (P<0.001), middle (P=0.001) and lower (P=0.015) in comparison to Ambu® AOTM.
CONCLUSIONS: Based on 3-D MRI measurements done on living patients, both SGDs distorted the anatomy of pediatric airway compared to their respective native values to variable extent. The relevance of these effects needs further studies on larger patient group in order to reduce morbidity on pediatric airway.

Entities:  

Mesh:

Year:  2016        PMID: 27314596     DOI: 10.23736/S0375-9393.16.11135-6

Source DB:  PubMed          Journal:  Minerva Anestesiol        ISSN: 0375-9393            Impact factor:   3.051


  3 in total

1.  Comparing i-gel and Ambu AuraOnce laryngeal mask airway in pediatric patients.

Authors:  Gui-Zhen Yang; Fu-Shan Xue; Hui-Xian Li; Ya-Yang Liu
Journal:  Saudi Med J       Date:  2017-12       Impact factor: 1.484

2.  Comparison of the clinical performance of i-gel and Ambu laryngeal masks in anaesthetised paediatric patients: A meta-analysis.

Authors:  Di Bao; Yun Yu; Wei Xiong; Ya-Xin Wang; Yi Liang; Lu Li; Bin Liu; Xu Jin
Journal:  World J Clin Cases       Date:  2022-02-06       Impact factor: 1.337

3.  Magnetic resonance imaging of the larynx in the pediatric population: A systematic review.

Authors:  Bernadette B L J Elders; Sergei M Hermelijn; Harm A W M Tiddens; Bas Pullens; Pjotr A Wielopolski; Pierluigi Ciet
Journal:  Pediatr Pulmonol       Date:  2019-01-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.