Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA-Compact-Max analyser.

A Woolley1, J-L Golmard2, S Kitchen1.   

Abstract

INTRODUCTION: Haemolysis, icterus and lipaemia (HIL) may affect haemostasis test results. This may be influenced by the level of interfering substance and the reagents and end-point detection system used.
METHODS: We assessed the influence of HIL on prothrombin time, activated partial thromboplastin time and fibrinogen assay using a viscosity-based detection analyser.
RESULTS: Spontaneous haemolysis that occurred during sample collection and processing had no effect on PT with either a rabbit tissue factor extract or recombinant human tissue factor reagents. In contrast, addition of mechanically haemolysed cells impacted on PT for the highest haemoglobin concentration. For APTTs determined with STA®-Cephascreen® reagent, there was no significant difference between results in haemolysed and nonhaemolysed samples. For the other two reagents studied, APTTs were statistically significantly shorter in haemolysed samples compared with nonhaemolysed samples. This bias was clinically significant only for STA®-PTT Automate. For all three APTT reagents, the impact of haemolysis was sufficient to impact patient management decisions, and in some samples, the effects of lipaemia and icterus were not clinically significant.
CONCLUSION: Overall, our results confirm that PT and fibrinogen were not clinically significantly affected by HIL. The APTTs of some haemolysed samples were falsely normal. Haemolysed samples for APTT determination should be rejected.
© 2016 John Wiley & Sons Ltd.

Entities:  

Keywords:  APTT; Coagulation; HIL; PT; mechanical endpoint

Mesh:

Substances:

Year:  2016        PMID: 27306848     DOI: 10.1111/ijlh.12498

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  7 in total

1.  A collaborative study by the Working Group on Hemostasis and Thrombosis of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC) on the interference of haemolysis on five routine blood coagulation tests by evaluation of 269 paired haemolysed/non-haemolysed samples.

Authors:  Chiara Novelli; Matteo Vidali; Bruno Brando; Benedetto Morelli; Giovanna Andreani; Marina Arini; Paola Calzoni; Roberta Giacomello; Barbara Montaruli; Emanuela Muccini; Angela Papa; Paola Pradella; Lucia Ruocco; Fosca Siviero; Filomena Gemma Viola; Mario Zanchetta; Lorena Zardo; Giuseppe Lippi
Journal:  Biochem Med (Zagreb)       Date:  2018-10-15       Impact factor: 2.313

2.  European survey on preanalytical sample handling - Part 2: Practices of European laboratories on monitoring and processing haemolytic, icteric and lipemic samples. On behalf of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for the Preanalytical Phase (WG-PRE).

Authors:  Janne Cadamuro; Giuseppe Lippi; Alexander von Meyer; Mercedes Ibarz; Edmee van Dongen; Michael Cornes; Mads Nybo; Pieter Vermeersch; Kjell Grankvist; Joao Tiago Guimaraes; Gunn B B Kristensen; Barbara de la Salle; Ana-Maria Simundic
Journal:  Biochem Med (Zagreb)       Date:  2019-06-15       Impact factor: 2.313

Review 3.  Advances in laboratory assessment of thrombosis and hemostasis.

Authors:  Jaewoo Song
Journal:  Blood Res       Date:  2022-04-30

4.  Clinical Evaluation of the Pre-Analytical Capabilities of Hemostasis Instrument.

Authors:  Mingyu Yang; Xiaoning Gui; Run Wang; Shiju Jiang; Jing Zhou; Jian Chen; Meiling Wang; Jiwei Ning; Linzi Miao; Hongwei Liu; Xiaomei Tang; Chenxue Qu
Journal:  Clin Appl Thromb Hemost       Date:  2022 Jan-Dec       Impact factor: 3.512

5.  The missing slope: paradoxical shortening of activated partial thromboplastin time in a patient on unfractionated heparin therapy.

Authors:  Ivana Lapić; Ana Lončar Vrančić; Désirée Coen Herak; Dunja Rogić
Journal:  Biochem Med (Zagreb)       Date:  2021-06-15       Impact factor: 2.313

6.  A comparative evaluation of the CN-6000 haemostasis analyser using coagulation, amidolytic, immuno-turbidometric and light transmission aggregometry assays.

Authors:  Chris Gardiner; Philip Lane; Katy Langley; Hitesh Tailor; Samuel J Machin; Ian J Mackie
Journal:  Int J Lab Hematol       Date:  2020-06-17       Impact factor: 3.450

7.  A performance evaluation of chemiluminescence enzyme immunoassays on the Sysmex CN-6500 haemostasis analyser.

Authors:  Chris Gardiner; Philip Lane; Hitesh Tailor; Samuel J Machin; Ian J Mackie
Journal:  Int J Lab Hematol       Date:  2021-07-12       Impact factor: 3.450

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.