Literature DB >> 27305517

Shaping ability of rotatory or reciprocating instruments in curved canals: a micro-computed tomographic study.

Alexander Pompermayer Jardine1, Ricardo Abreu da Rosa1, Manuela Favarin Santini1, Ivana Maria Zaccara1, Marcus Vinicius Reis Só1, Patrícia Maria Poli Kopper1.   

Abstract

This study aimed to compare apical transportation, centering ratio, and working time during root canal preparation with Wizard Navigator (WN), WaveOne (WO), or ProTaper Universal (PT) and to describe deformation and fracture of these instruments. Thirty-six mesiobuccal roots from maxillary molars were micro-computed tomography (micro-CT) scanned and then sorted into three groups (n = 12): Wizard Navigator (WN), WaveOne (WO), and ProTaper Universal (PT). The root canals were prepared using WN, WO, or PT; the time of each canal preparation was timed; and the specimens were micro-CT re-scanned. The instruments were submitted to scanning electron microscopy (SEM) before and after their use. The data on canal transportation at 1, 2, 3, 4 and 5 mm and preparation time were analyzed by the Kruskal-Wallis test, followed by the Mann-Whitney U test. The centering ratio was analyzed by ANOVA, followed by Tukey's test. Both instrument deformation and fracture were described. Apical transportation was similar among groups at 1, 2, 3 and 4 mm. The WO group showed higher canal transportation at 5mm than the other groups (p = 0.03). There was no significant difference in centering ratio among the groups. Preparation time in the WO group was significantly lower than in the other groups (p = 0.004). Small differences were observed in the surface area of all instruments. The WN, WO, and PT groups had a similar centering ratio without procedural errors or significant structural changes. At 5 mm from the apex, the WO group showed the largest canal transportation toward the furcation and root canal preparation was faster than in the WN and PT groups.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27305517     DOI: 10.1590/1807-3107BOR-2016.vol30.0086

Source DB:  PubMed          Journal:  Braz Oral Res        ISSN: 1806-8324


  1 in total

1.  Centering Ability of ProTaper Next and WaveOne Classic in J-Shape Simulated Root Canals.

Authors:  Giuseppe Troiano; Mario Dioguardi; Armando Cocco; Michele Giuliani; Cristiano Fabiani; Alfonso D'Alessandro; Domenico Ciavarella; Lorenzo Lo Muzio
Journal:  ScientificWorldJournal       Date:  2016-12-08