Impact of Neointimal Calcifications on Acute Stent Performance during the Treatment of In-Stent Restenosis.

Emile Mehanna1, Guilherme Ferragut Attizzani1, Daisuke Nakamura1, Setsu Nishino1, Anas Fares1, Reem Aoun1, Marco Aurelio Costa1, Hiram Grando Bezerra1.   

Abstract

Optical coherence tomography (OCT) has become the invasive imaging modality of choice for coronary stent assessment due to its unmatched spatial resolution. Neointimal calcification (NC) is a rare finding, observed in 5-10% of in-stent restenosis (ISR) neointima. The impact of NC on percutaneous coronary intervention of ISR is unknown. We therefore present the outcome of six unique cases of ISR and NC in which OCT was used to evaluate the impact of NC on the quality of stent-in-stent deployment for the treatment of ISR. This series demonstrates for the first time the impact of NC on stent expansion, a finding which might help guiding percutaneous coronary intervention for ISR with NC.

Entities:  

Mesh:

Year:  2016        PMID: 27305286      PMCID: PMC4914007          DOI: 10.5935/abc.20160068

Source DB:  PubMed          Journal:  Arq Bras Cardiol        ISSN: 0066-782X            Impact factor:   2.000


Brief Communication

Neoatherosclerosis, defined as the presence of neointimal calcification (NC) or lipid-laden neointima,[1] has been reported as an important mechanism of late stent failure.[2] Intravascular imaging modalities enabled further elucidation of neoatherosclerosis´ pathophysiology in vivo.[3] Neointimal calcification is observed in 5-10% of in-stent restenosis (ISR),[4] but its impact on the acute performance of stents implanted in-stent for the treatment of ISR is unknown. Intravascular optical coherence tomography (OCT) enables precise assessment of calcified plaques, while dramatically reducing imaging artifacts compared with intravascular ultrasound.[5] We therefore used OCT to evaluate the impact of NC on the quality of stent-in-stent deployment for the treatment of ISR. Herewith we present 6 cases of ISR and NC from our institution's OCT registry. OCT (C7-XR OCT Intravascular Imaging System; St.Jude Medical, St. Paul, Minnesota) images were acquired pre- and post-stent-in-stent procedure using the integrated automated pullback device at 20 mm/s (frame interval of 0.2 mm). Neointimal calcification was defined as an area of low attenuation, low backscattering and clear borders within the stent neointima (Figure 1). Areas and diameters for the old (outer) and newly implanted (inner) stents were obtained; in addition, the mean distance and area between the stents were automatically obtained by 360° chords (Figure 1). Stent eccentricity was defined as minimum stent diameter/maximum stent diameter, while stent expansion was defined as the average stent area at the NC zone divided by the average stent reference [(average proximal reference + average distal reference)/2]. OCT analysis was undertaken offline by a Core Laboratory blinded to the procedure´s characteristics using commercially available software (Version C.0.4, St Jude Medical, St. Paul, MN). Analyses were concentrated in 3 consecutive frames at 3 different locations (i.e. 9 frames per OCT pullback): 1) NC region, 2) proximal and 3) distal to NC region (Figure 1).
Figure 1

OCT images of in-stent restenosis case with neointimal calcification (red arrow) before (A) and after (B) stent-in-stent implantation. Panel C: schematic representation of the effect of neointimal calcification (NC) on stent expansion. D: stent diameter; d: distance between stents.

OCT images of in-stent restenosis case with neointimal calcification (red arrow) before (A) and after (B) stent-in-stent implantation. Panel C: schematic representation of the effect of neointimal calcification (NC) on stent expansion. D: stent diameter; d: distance between stents. Stent areas, diameters and stent eccentricity were similar between the frames with NC and the frames distal to the NC (Table 1). When comparing the NC area to the ISR region proximal to the NC, there was a trend for smaller area (difference = 0.9 mm2, p = 0.09), and diameter (difference = 0.2mm, p = 0.09) of the inner (newly implanted) stent and bigger stent area (difference = 1.2 mm2, p = 0.06), and diameter (difference = 0.2mm, p = 0.06) of the external (older) stent at the location of NC compared to the proximal non-calcified ISR analyzed frames (Table 1).
Table 1

Quantitative effect of neointimal calcification on stent implantation

 Difference  Difference  
 Proximal – NCtp-valueDistal – NCtp-value
Stent 1 Area0.902.110.090.581.520.19
Stent 2 Area-1.20-2.490.06-0.76-1.350.23
Minimum d0.031.610.170.020.830.45
Mean d0.213.900.010.133.210.02
Maximum d0.438.300.010.272.060.09
Minimum D10.171.920.110.090.950.38
Mean D10.202.120.090.121.430.21
Maximum D10.242.320.070.151.510.19
SE D10.020.950.390.021.310.25
Minimum D2-0.20-2.430.06-0.14-1.310.25
Mean D2-0.20-2.350.06-0.14-1.310.25
Maximum D2-0.22-2.680.04-0.15-1.450.21
SE D2-0.01-0.880.42-0.004-0.380.72

Stent 1: inner (newly implanted stent); stent 2: outer (older stent; d: distance between stents; D: diameter of stent; SE: stent eccentricity (minD/MaxD); NC: neointimal calcification.

Quantitative effect of neointimal calcification on stent implantation Stent 1: inner (newly implanted stent); stent 2: outer (older stent; d: distance between stents; D: diameter of stent; SE: stent eccentricity (minD/MaxD); NC: neointimal calcification. The mean distance between the stents was always longer at the area of calcification: difference between the NC area and the distal area was 0.13mm (p = 0.02) and the NC area and the proximal ISR region as 0.21 (p = 0.01). The average stent expansion at the area of calcified neointima was 81.4%. Stent underexpansion has been linked to clinical adverse events, notably stent thrombosis and restenosis.[6,7] We were able to demonstrate that the presence of NC led to underexpansion of the newly implanted stent compared with adjacent segments. Further investigation is required to determine whether these findings have an impact on clinical events. The mechanisms leading to stent ISR have been divided into technical (barotrauma outside stented segment, stent gap, residual uncovered atherosclerotic plaques), mechanical (stent underexpansion, non-uniform stent strut distribution, stent fracture, non-uniform drug elution/ deposition, polymer peeling) and biological (drug resistance, hypersensitivity).[8] The advancement of intravascular imaging, notably OCT, is expected to allow a better understanding of the ISR process and will likely influence the therapeutic strategies (i.e., customized therapy) utilized in this scenario. While current alternatives for ISR therapy (i.e., plain balloon angioplasty, drug-eluting balloon, in-stent DES) are mostly based on the type of restenosis (focal in-stent, focal at stent edge, diffuse in-stent, proliferative),[9] they do not take neointimal qualitative assessment into account. We believe information provided by OCT imaging could, therefore, further improve therapeutic decisions in ISR. For example, in cases of ISR with NC as herewith described, more aggressive in-stent pre-dilations or use of debulking devices could potentially help improve the expansion of the newly implanted stent. The effect of neointimal atherosclerosis characterization on therapeutic choices for ISR therapy and its effect on clinical outcomes are yet to be determined in future prospective studies.
  9 in total

1.  Treatment of drug-eluting stent restenosis.

Authors:  Marco A Costa
Journal:  Am Heart J       Date:  2007-04       Impact factor: 4.749

2.  Optical coherence tomographic analysis of in-stent neoatherosclerosis after drug-eluting stent implantation.

Authors:  Soo-Jin Kang; Gary S Mintz; Takashi Akasaka; Duk-Woo Park; Jong-Young Lee; Won-Jang Kim; Seung-Whan Lee; Young-Hak Kim; Cheol Whan Lee; Seong-Wook Park; Seung-Jung Park
Journal:  Circulation       Date:  2011-06-06       Impact factor: 29.690

3.  Stent-related defects in patients presenting with stent thrombosis: differences at optical coherence tomography between subacute and late/very late thrombosis in the Mechanism Of Stent Thrombosis (MOST) study.

Authors:  Guido Parodi; Alessio La Manna; Luca Di Vito; Marco Valgimigli; Massimo Fineschi; Benedetta Bellandi; Giampaolo Niccoli; Betti Giusti; Renato Valenti; Alberto Cremonesi; Giuseppe Biondi-Zoccai; Francesco Prati
Journal:  EuroIntervention       Date:  2013-12       Impact factor: 6.534

4.  The pathology of neoatherosclerosis in human coronary implants bare-metal and drug-eluting stents.

Authors:  Gaku Nakazawa; Fumiyuki Otsuka; Masataka Nakano; Marc Vorpahl; Saami K Yazdani; Elena Ladich; Frank D Kolodgie; Aloke V Finn; Renu Virmani
Journal:  J Am Coll Cardiol       Date:  2011-03-15       Impact factor: 24.094

Review 5.  In-stent restenosis in the drug-eluting stent era.

Authors:  George D Dangas; Bimmer E Claessen; Adriano Caixeta; Elias A Sanidas; Gary S Mintz; Roxana Mehran
Journal:  J Am Coll Cardiol       Date:  2010-11-30       Impact factor: 24.094

Review 6.  In-stent neoatherosclerosis: a final common pathway of late stent failure.

Authors:  Seung-Jung Park; Soo-Jin Kang; Renu Virmani; Masataka Nakano; Yasunori Ueda
Journal:  J Am Coll Cardiol       Date:  2012-06-05       Impact factor: 24.094

7.  Spatial heterogeneity of neoatherosclerosis and its relationship with neovascularization and adjacent plaque characteristics: optical coherence tomography study.

Authors:  Jinwei Tian; Xuefeng Ren; Shiro Uemura; Harold Dauerman; Abhiram Prasad; Catalin Toma; Haibo Jia; Farhad Abtahian; Rocco Vergallo; Sining Hu; Iris McNulty; Hang Lee; Stephen Lee; Bo Yu; Ik-Kyung Jang
Journal:  Am Heart J       Date:  2014-04-03       Impact factor: 4.749

8.  A volumetric intravascular ultrasound comparison of early drug-eluting stent thrombosis versus restenosis.

Authors:  Xuebo Liu; Hiroshi Doi; Akiko Maehara; Gary S Mintz; Jose de Ribamar Costa; Koichi Sano; Giora Weisz; George D Dangas; Alexandra J Lansky; Edward M Kreps; Michael Collins; Martin Fahy; Gregg W Stone; Jeffrey W Moses; Martin B Leon; Roxana Mehran
Journal:  JACC Cardiovasc Interv       Date:  2009-05       Impact factor: 11.195

9.  Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention.

Authors:  Hiram G Bezerra; Guilherme F Attizzani; Vasile Sirbu; Giuseppe Musumeci; Nikoloz Lortkipanidze; Yusuke Fujino; Wei Wang; Sunao Nakamura; Andrej Erglis; Giulio Guagliumi; Marco A Costa
Journal:  JACC Cardiovasc Interv       Date:  2013-03       Impact factor: 11.195

  9 in total
  1 in total

1.  Severely Circumferentially Calcified Neointima as a New Cause of Undilatable In-Stent Restenosis.

Authors:  Manabu Kashiwagi; Takashi Tanimoto; Hironori Kitabata
Journal:  Case Rep Cardiol       Date:  2018-05-29
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.