12-Month Coronary Angiography, Intravascular Ultrasound and Histology Evaluation of a Novel Fully Bioabsorbable Poly-L-Lactic Acid/Amorphous Calcium Phosphate Scaffolds in Porcine Coronary Arteries.

Gaoke Feng, Jianmin Xiao, Yuying Bi, Zhiyuan Lan, Xiaoxin Zheng, Zhao Lu, Jun Li, Kan Wu, Edward Kislauskis, Stephen McCarthy, Qiang Hu, Xuejun Jiang, Tim Wu, Roger Laham.   

Abstract

Our previous studies have confirmed the superior biocompatibility of the poly-L-lactic acid/amorphous calcium phosphate (PLLA/ACP) scaffolds (PowerScaffold) compared to PLLA scaffolds and their similar 6-month radial strength compared with TAXUS stents. In order to conduct further dynamic observations on the performance of the PowerScaffold after 12-month implantation compared with the TAXUS stents. Twenty PowerScaffold and 20 TAXUS were implanted in porcine coronary arteries. At 12-month follow-up, Quantitative Coronary Angiography showed that the stent reference vessel diameter (3.19 ± 0.25 mm vs. 2.75 ± 0.22 mm, p < 0.05), the mean lumen diameter (3.07 ± 0.22 mm vs. 2.70 ± 0.17 mm, p < 0.05) and the late lumen gain (0.45 ± 0.07 mm vs. 0.06 ± 0.06 mm, p < 0.01) were all significantly greater with the PowerScaffold than the TAXUS. As well, Intravascular Ultrasound showed the stent reference vessel area (7.74 ± 0.48 mm2 vs. 6.96 ± 0.51 mm2, p < 0.05), the mean stent area (7.49 ± 0.46 mm2 vs. 6.53 ± 0.47 mm2, p < 0.05) and the mean lumen area (7.22 ± 0.50 mm2 vs. 6.00 ± 0.48 mm2, p < 0.01) were all significantly greater with the PowerScaffold than the TAXUS. The luminal patency rate of the PowerScaffold significantly increased from 72.45 ± 6.84% at 1 month to 93.54 ± 8.15% at 12 months (p < 0.01) while the TAXUS stents were associated with a non-significant decreasing trend (89.44 ± 8.44% vs. 86.53 ± 8.22%). Pathology indicated the average thickness of the struts degraded by 14.25 ± 3.04 μm at 1 month, 23.39 ± 2.45 μm at 6 months and 35.54 ± 2.20 μm at 12 months. Immunohistochemical examination showed that the expression of inflammatory factors NF-κB gradually decreased from 1-month to 12-month (36.79 ± 4.78 vs. 5.79 ± 2.85, P < 0.01). As the late lumen gain of arteries implanted with the PowerScaffold increases over time with the growth of vessels, it effectively reverse the late vascular negative remodeling observed with the TAXUS stents, providing a better option for lumen restoration treatment in clinical practice.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27301200     DOI: 10.1166/jbn.2016.2241

Source DB:  PubMed          Journal:  J Biomed Nanotechnol        ISSN: 1550-7033            Impact factor:   4.099


  1 in total

1.  Effect of novel bioresorbable scaffold composed of poly-L-lactic acid and amorphous calcium phosphate nanoparticles on inflammation and calcification of surrounding tissues after implantation.

Authors:  Gaoke Feng; Chaoshi Qin; Xin Yi; Jinggang Xia; Jingjing Chen; Xiyu Chen; Tao Chen; Xuejun Jiang
Journal:  J Mater Sci Mater Med       Date:  2018-07-17       Impact factor: 3.896

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.