The Icare HOME (TA022) Study: Performance of an Intraocular Pressure Measuring Device for Self-Tonometry by Glaucoma Patients.

Lucy I Mudie1, Sophie LaBarre1, Varshini Varadaraj1, Sezen Karakus1, Jouni Onnela2, Beatriz Munoz1, David S Friedman3.   

Abstract

PURPOSE: To evaluate the Icare HOME (TA022) device (Icare Oy, Vanda, Finland) for use by glaucoma patients for self-tonometry.
DESIGN: Prospective performance evaluation of a medical device. PARTICIPANTS: One hundred eighty-nine participants with glaucoma or suspected glaucoma were recruited from the Wilmer Eye Institute, Baltimore, Maryland, between July 2014 and April 2015.
METHODS: Participants had standardized training and had to be able use the Icare HOME device independently. Subjects also had to be able to obtain the first intraocular pressure (IOP) measurement within 5 mmHg of Goldmann applanation tonometry (GAT). Those certified obtained 3 IOP measures using the HOME device, and these were compared with Icare TA01i and GAT IOP measurements. MAIN OUTCOME MEASURES: The agreement between Icare HOME and reference tonometers was used to assess precision. The intraclass correlation coefficient was used to assess within-patient reliability for the HOME device.
RESULTS: Eighteen of 189 recruited patients were ineligible to take part in the study. Forty-four of 171 patients (25.7%) started but failed to complete the study: 7 stopped because of time concerns, 10 of 171 patients (6%) stopped because of difficulty using the device during certification, and 27 of 171 patients (16%) failed to be certified based on IOP. The HOME and GAT measurements agreed within 5 mmHg in 116 of 127 participants (91.3%); 2 participants (1.6%) had a difference of more than 7 mmHg. The mean difference between the Icare HOME and GAT measurements was -0.33 mmHg (standard deviation, 3.11 mmHg). The overall intraclass correlation coefficient for the HOME device was 0.92 (95% confidence interval, 0.89-0.95).
CONCLUSIONS: Not all participants could learn how to use the Icare HOME device, but for those who could, most were able to obtain measurements similar to those obtained by GAT. The Icare HOME device is safe and reliable for self-tonometry, but nearly 1 in 6 individuals may fail to certify in use of the device based on large differences in IOP when comparing GAT with the Icare HOME measurements. The device has the potential to address an unmet need by providing more frequent IOP measurements in a patient's day to day life.
Copyright © 2016 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27289178     DOI: 10.1016/j.ophtha.2016.04.044

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  13 in total

1.  Critical pathway for primary open angle glaucoma diagnosis.

Authors:  Alejandro R Allocco; Julia A Ponce; Maria J Riera; Mauricio G Magurno
Journal:  Int J Ophthalmol       Date:  2017-06-18       Impact factor: 1.779

Review 2.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

3.  Association Between 24-Hour Intraocular Pressure Monitored With Contact Lens Sensor and Visual Field Progression in Older Adults With Glaucoma.

Authors:  Carlos Gustavo De Moraes; Kaweh Mansouri; Jeffrey M Liebmann; Robert Ritch
Journal:  JAMA Ophthalmol       Date:  2018-07-01       Impact factor: 7.389

4.  Measurement of Intraocular Pressure by Patients With Glaucoma.

Authors:  Savva Pronin; Lyndsay Brown; Roly Megaw; Andrew J Tatham
Journal:  JAMA Ophthalmol       Date:  2017-10-01       Impact factor: 7.389

5.  Usefulness of Icare Home in Telemedicine Workflow to Detect Real-World Intraocular Pressure Response to Glaucoma Medication Change.

Authors:  Carin D Rojas; David M Reed; Sayoko E Moroi
Journal:  Ophthalmol Glaucoma       Date:  2020-05-11

6.  Home Self-tonometry Trials Compared with Clinic Tonometry in Patients with Glaucoma.

Authors:  Elyse J McGlumphy; Aleksandra Mihailovic; Pradeep Y Ramulu; Thomas V Johnson
Journal:  Ophthalmol Glaucoma       Date:  2021-04-09

7.  Self-Monitoring Symptoms in Glaucoma: A Feasibility Study of a Web-Based Diary Tool.

Authors:  Leanne McDonald; Fiona C Glen; Deanna J Taylor; David P Crabb
Journal:  J Ophthalmol       Date:  2017-05-04       Impact factor: 1.909

8.  Self-monitoring of intraocular pressure using Icare HOME tonometry in clinical practice.

Authors:  Barbara Cvenkel; Makedonka Atanasovska Velkovska
Journal:  Clin Ophthalmol       Date:  2019-05-10

9.  Diurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial.

Authors:  Jessie Huang; Paula Katalinic; Michael Kalloniatis; Michael P Hennessy; Barbara Zangerl
Journal:  Optom Vis Sci       Date:  2018-02       Impact factor: 1.973

Review 10.  Icare® rebound tonometers: review of their characteristics and ease of use.

Authors:  Shunsuke Nakakura
Journal:  Clin Ophthalmol       Date:  2018-07-12