| Literature DB >> 27279503 |
Dmitry V Petrov1, Ivan I Matrosov2.
Abstract
In the present work, an improved model of the Raman gas analyzer (RGA) of natural gas (NG) developed by us is described together with its operating principle. The sensitivity has been improved and the number of measurable gases has been expanded. Results of its approbation on a real NG sample are presented for different measurement times. A comparison of the data obtained with the results of chromatographic analysis demonstrates their good agreement. The time stability of the results obtained using this model is analyzed. It is experimentally established that the given RGA can reliably determine the content of all molecular NG components whose content exceeds 0.005% for 100 s; moreover, in this case the limiting sensitivity for some NG components is equal to 0.002%.Keywords: Raman spectroscopy; gas analysis; methane; natural gas
Year: 2016 PMID: 27279503 DOI: 10.1177/0003702816644611
Source DB: PubMed Journal: Appl Spectrosc ISSN: 0003-7028 Impact factor: 2.388