The aim of this study was to evaluate the effectiveness of "Protex" (Parker; Fairfield, NJ, USA) for disinfection of ultrasound probes. We examined bacterial contamination on ultrasound probes that were wiped with a plain paper towel, with a plain and an ethanol-soaked paper towel, or with a plain and Protex-soaked paper towel. The plain paper towel was used to remove the gel, and was contaminated by large numbers of bacteria, but the use of ethanol-soaked paper towels and that of paper towels soaked in Protex™ broad-spectrum disinfectant (Parker: Fairfield, NJ, USA) reduced those numbers markedly.

Entities:  

Keywords:  Disinfection; Probe; Protex

Year:  2012        PMID: 27277109     DOI: 10.1007/s10396-012-0404-5

Source DB:  PubMed          Journal:  J Med Ultrason (2001)        ISSN: 1346-4523            Impact factor:   1.314


  5 in total

1.  An outbreak of pyodermas among neonates caused by ultrasound gel contaminated with methicillin-susceptible Staphylococcus aureus.

Authors:  K Weist; C Wendt; L R Petersen; H Versmold; H Rüden
Journal:  Infect Control Hosp Epidemiol       Date:  2000-12       Impact factor: 3.254

2.  Degradation of ultrasound probes caused by disinfection with alcohol.

Authors:  Harumi Koibuchi; Yasutomo Fujii; Kazuhiko Kotani; Kei Konno; Hiroaki Matsunaga; Michiaki Miyamoto; Nobuyuki Taniguchi
Journal:  J Med Ultrason (2001)       Date:  2010-12-23       Impact factor: 1.314

3.  Evaluation of procedures for decontaminating ultrasound probes.

Authors:  Shunji Hayashi; Harumi Koibuchi; Nobuyuki Taniguchi; Yoshikazu Hirai
Journal:  J Med Ultrason (2001)       Date:  2011-10-26       Impact factor: 1.314

4.  Comparison of methods for evaluating bacterial contamination of ultrasound probes.

Authors:  Harumi Koibuchi; Shunji Hayashi; Kazuhiko Kotani; Yasutomo Fujii; Kei Konno; Yoshikazu Hirai; Nobuyuki Taniguchi
Journal:  J Med Ultrason (2001)       Date:  2009-08-20       Impact factor: 1.314

5.  Nosocomial outbreak of Klebsiella pneumoniae producing SHV-5 extended-spectrum beta-lactamase, originating from a contaminated ultrasonography coupling gel.

Authors:  O Gaillot; C Maruéjouls; E Abachin; F Lecuru; G Arlet; M Simonet; P Berche
Journal:  J Clin Microbiol       Date:  1998-05       Impact factor: 5.948

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.