Billowing of Endologix Powerlink Stent Graft in Post Endovascular Repair of Abdominal Aortic Aneurysm: Simultaneous Findings on CT and Ultrasound Imaging.

Naseem Helo1, Arthur C Chang2, Christine Hyun2, Kenneth S Chon2, Alex C Yi2.   

Abstract

Endoleaks remain a main concern in endovascular aneurysm repair. A potential false positive in the diagnosis of endoleak surveillance includes "billowing," a phenomenon of the polytetrafluoroethylene plastic outer fabric of the stent graft is seen separated from the struts of the metallic endoskeleton. Contrast presents beyond the endoskeleton but is still contained within the graft. This is secondary to intermittent attachment of the graft to the endoskeleton, and is only found in the Endologix Powerlink. The finding of billowing has been previously reported and is a common knowledge by those who deploy the stent graft. We report a case where there is contemporaneous imaging on computed tomography angiography as well as on ultrasound.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27263812     DOI: 10.1016/j.avsg.2016.01.053

Source DB:  PubMed          Journal:  Ann Vasc Surg        ISSN: 0890-5096            Impact factor:   1.466


  1 in total

1.  A Unique Indication for the Altura Endograft to Repair Bilateral Common Iliac Artery Aneurysms Associated with a Short-Infrarenal Aorta.

Authors:  Stavros K Kakkos; Chrysanthi P Papageorgopoulou; Konstantinos Katsanos; Peter Zampakis; Athina Siampalioti; Ioannis Ntouvas; Chrysanthi-Helen Loizou; Eleni Tsamantioti; Spyros Papadoulas; Konstantinos M Nikolakopoulos; Anastasia Kouri
Journal:  Aorta (Stamford)       Date:  2020-07-31
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.