Delivering RapidArc®: A comprehensive study on accuracy and long term stability.

A Scaggion1, A Negri2, M A Rossato3, A Roggio4, F Simonato5, S Bacco6, M Paiusco7.   

Abstract

PURPOSE: To establish the reliability and accuracy of a UNIQUE Linac in delivering RapidArc treatments and assess its long term stability.
MATERIALS AND METHODS: UNIQUE performance was monitored and analyzed for a period of nearly two years. 2280 Dynalog files, related to 179 clinical RapidArc treatments were collected. Different tumor sites and dose scheduling were included, covering the full range of our treatment plans. Statistical distributions of MLC motion error, gantry rotation error and MU delivery error were evaluated. The stochastic and systematic nature of each error was investigated together with their variation in time.
RESULTS: All the delivery errors are found to be small and more stringent tolerances than those proposed by TG142 are suggested. Unlike MLC positional errors, where a linear relationship with leaf speed holds, other Volumetric Modulated Arc Therapy (VMAT) parameters reveal a random nature and, consequently, a reduced clinical relevance. MLC errors are linearly related only to leaf speed no matter the shape of the MLC apertures. Gantry rotation and MU delivery are as accurate as major competing Linacs. UNIQUE was found to be reliable and accurate throughout the investigation period, regardless of the specific tumor sites and fractionation schemes.
CONCLUSIONS: The accuracy of RapidArc treatments delivered with UNIQUE has been established. The stochastic nature of delivery errors is proven. Long term statistics of the delivery parameter errors do not show significant variations, confirming the reliability of the VMAT delivery system.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Delivery accuracy; Delivery error; Dynamic log files; Long-term stability; VMAT

Mesh:

Year:  2016        PMID: 27262921     DOI: 10.1016/j.ejmp.2016.05.056

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy.

Authors:  Hideaki Hirashima; Mitsuhiro Nakamura; Yuki Miyabe; Megumi Uto; Kiyonao Nakamura; Takashi Mizowaki
Journal:  Radiat Oncol       Date:  2018-02-14       Impact factor: 3.481

2.  Quality assurance of VMAT on flattened and flattening filter-free accelerators using a high spatial resolution detector.

Authors:  F S Matar; D Wilkinson; J Davis; G Biasi; T Causer; I Fuduli; O Brace; N Stansook; M Carolan; A B Rosenfeld; Marco Petasecca
Journal:  J Appl Clin Med Phys       Date:  2020-04-11       Impact factor: 2.102

3.  Plan quality comparison for cervical carcinoma treated with Halcyon and Trilogy intensity-modulated radiotherapy.

Authors:  Chengqiang Li; Jinhu Chen; Jian Zhu; Guanzhong Gong; Cheng Tao; Zhenjiang Li; Jie Lu; Yong Yin
Journal:  J Cancer       Date:  2019-10-15       Impact factor: 4.207

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.