Intraocular pressure study using monitored forced-infusion system phacoemulsification technology.

Jason D Jensen1, Tyler Boulter1, Nathan G Lambert1, Brian Zaugg1, Brian C Stagg1, Jeff H Pettey1, Randall J Olson2.   

Abstract

PURPOSE: To optimize phacoemulsification efficiency by varying the intraocular pressure (IOP) using the Centurion Vision System, a monitored forced-infusion system.
SETTING: John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, USA.
DESIGN: Laboratory investigation.
METHODS: Porcine lenses were soaked in formalin and divided into 2.0 mm cubes; 0.9 mm balanced tips were used. Intraocular pressure levels were tested from 30 to 110 mm Hg at 10 mm Hg intervals. Vacuum was set at 500 mm Hg, aspiration at 35 mL/min, and torsional power at 60%. Efficiency (time to lens removal) and chatter (number of lens fragment repulsions from the tip) were measured.
RESULTS: There was a linear relationship between efficiency and IOP. Increasing IOP led to an increase in efficiency (ie, to a decrease in time necessary to remove the lens fragment) (R(2) = .5769, P = .0176). Chatter correlated with IOP in the range of 30 to 50 mm Hg (R(2) = .4506, P = .0448) and was minimal at higher IOPs.
CONCLUSIONS: Increasing IOP in the monitored forced-infusion system improved efficiency. Chatter was also decreased or eliminated with increasing IOP. Additional studies will be required to understand the reasons for these findings. FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27255254     DOI: 10.1016/j.jcrs.2016.01.045

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  The effect of increasing power when grooving using phacoemulsification.

Authors:  Rhett S Thomson; Brian A Bird; Lance A Stutz; Joshua B Heczko; Ashlie A Bernhisel; William R Barlow; Brian Zaugg; Randall J Olson; Jeff H Pettey
Journal:  Clin Ophthalmol       Date:  2019-04-12

2.  Optimizing Tip Diameter in Phacoemulsification of Varying Lens Sizes: An in vitro Study.

Authors:  Aniket Ramshekar; Joshua Heczko; Ashlie Bernhisel; William Barlow; Brian Zaugg; Randall Olson; Jeff Pettey
Journal:  Clin Ophthalmol       Date:  2021-11-17

3.  A prospective randomized clinical trial of active-fluidics versus gravity-fluidics system in phacoemulsification for age-related cataract (AGSPC).

Authors:  Yu Luo; Hongyu Li; Wenqian Chen; Yi Gao; Tianju Ma; Zi Ye; Zhaohui Li
Journal:  Ann Med       Date:  2022-12       Impact factor: 5.348

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.