Literature DB >> 27247748

New Technique for Removal of the Ulnar Intramedullary Nail Failed to Remove: Broken Assembly Piece in the Nail.

Jin Woong Yi1, Byung Hak Oh1, Sang-Bum Kim1, Youn Moo Heo1, Tae Gyun Kim1, Doo Hyun Kim1, You Sun Jung1.   

Abstract

Methods about removal of intramedullary nail in complicated cases were reported in some literatures but there are no reports about nail removal in the ulna. The authors would like to report such a case and the technique. We removed bone of the inlet site and created another bony window using an osteotome to expose the interlocking screw holes. Only a bony window the size of 2 inter-interlocking holes at the most proximal part of the nail can be used to remove the nail with minimal damage of the triceps brachii tendon and soft tissue.

Entities:  

Keywords:  Device removal; Intramedullary nailing

Mesh:

Year:  2016        PMID: 27247748      PMCID: PMC4870326          DOI: 10.4055/cios.2016.8.2.210

Source DB:  PubMed          Journal:  Clin Orthop Surg        ISSN: 2005-291X


  3 in total

1.  Technique for removal of intramedullary nails when there is failure of the proximal extraction device: a report of three cases.

Authors:  G M Georgiadis; B E Heck; N A Ebraheim
Journal:  J Orthop Trauma       Date:  1997 Feb-Mar       Impact factor: 2.512

2.  Retrieval of a broken intramedullary nail.

Authors:  S J Incavo; T K Kristiansen
Journal:  Clin Orthop Relat Res       Date:  1986-09       Impact factor: 4.176

3.  A simple method for removal of a fractured intramedullary nail.

Authors:  O Levy; Y Amit; S Velkes; H Horoszowski
Journal:  J Bone Joint Surg Br       Date:  1994-05