| Literature DB >> 27247461 |
Jessica Burn1, Nigel A Caulkett1, Marta Gunn1, Claire Cooney1, Susan J Kutz1, Søren R Boysen1.
Abstract
This study evaluated the ability of a portable oxygen concentrator (POC) to provide fresh gas to an anesthetic machine via an Ayre's T-piece or a Bain circuit. Fraction of inspired oxygen (FiO2) was compared at flows of 0.5 to 3.0 L/min. Measured FiO2 was 96% at flow rates ≥ 1 L/min. Mean battery life at 1.0, 2.0, and 3.0 L/min was 4.21 ± 0.45, 2.62 ± 0.37 and 1.5 ± 0.07 hours, respectively. The POC proved to be useful and effective during 2 power outages. The POC was sufficient to prevent rebreathing in 70% of dogs using a T-piece circuit and 20% of dogs with a Bain circuit. A significant negative correlation between inspired CO2 and O2 flow rates was noted. A significant positive correlation between inspired CO2 and ETCO2 was documented. The occurrence of hypercarbia was associated with low O2 flow. Battery back-up was essential during power outages. The POC can be effectively used for delivery of anesthesia.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27247461 PMCID: PMC4866666