Measuring temperature rise during orthopaedic surgical procedures.

Sarah Manoogian1, Adam K Lee2, James C Widmaier2.   

Abstract

A reliable means for measuring temperatures generated during surgical procedures is needed to recommend best practices for inserting fixation devices and minimizing the risk of osteonecrosis. Twenty four screw tests for three surgical procedures were conducted using the four thermocouples in the bone and one thermocouple in the screw. The maximum temperature rise recorded from the thermocouple in the screw (92.7±8.9°C, 158.7±20.9°C, 204.4±35.2°C) was consistently higher than the average temperature rise recorded in the bone (31.8±9.3°C, 44.9±12.4°C, 77.3±12.7°C). The same overall trend between the temperatures that resulted from three screw insertion procedures was recorded with significant statistical analyses using either the thermocouple in the screw or the average of several in-bone thermocouples. Placing a single thermocouple in the bone was determined to have limitations in accurately comparing temperatures from different external fixation screw insertion procedures. Using the preferred measurement techniques, a standard screw with a predrilled hole was found to have the lowest maximum temperatures for the shortest duration compared to the other two insertion procedures. Future studies evaluating bone temperature increase need to use reliable temperature measurements for recommending best practices to surgeons.
Copyright © 2016 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bone; Fracture; Orthopaedic; Osteonecrosis; Schanz

Mesh:

Year:  2016        PMID: 27246667     DOI: 10.1016/j.medengphy.2016.05.003

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  1 in total

1.  Unicortical self-drilling external fixator pins reduce thermal effects during pin insertion.

Authors:  Markus Greinwald; Patrick A Varady; Peter Augat
Journal:  Eur J Trauma Emerg Surg       Date:  2017-12-14       Impact factor: 3.693

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.