Literature DB >> 27235513

Pseudotumour formation and subsequent resolution in metal-on-metal total hip arthroplasty following revision: Instructional review and an illustrative case report with revision using a dual mobility design.

A A Sassoon1, R L Barrack2.   

Abstract

The use of large-diameter metal-on-metal (MoM) components in total hip arthroplasty (THA) is associated with an increased risk of early failure due to adverse local tissue reaction to metal debris (ARMD) in response to the release of metal ions from the bearing couple and/or head-neck taper corrosion. The aim of this paper was to present a review of the incidence and natural history of ARMD and the forms of treatment, with a focus on the need for and extent of resection or debulking of the pseudotumour. An illustrative case report is presented of a patient with an intra-pelvic pseudotumour associated with a large diameter MoM THA, which was treated successfully with revision of the bearing surface to a dual mobility couple and retention of the well-fixed acetabular and femoral components. The pseudotumour was left in situ Resolution of the intra-pelvic mass and normalisation of metal ion levels was observed seven months post-operatively. Cite this article: Bone Joint J 2016;98-B:736-40. ©2016 The British Editorial Society of Bone & Joint Surgery.

Entities:  

Keywords:  Corrosion; Metal-on-Metal; Pseudotumour; Revision Total Hip Arthroplasty

Mesh:

Substances:

Year:  2016        PMID: 27235513     DOI: 10.1302/0301-620X.98B6.36908

Source DB:  PubMed          Journal:  Bone Joint J        ISSN: 2049-4394            Impact factor:   5.082


  5 in total

1.  Salvage of a monoblock metal-on-metal cup using a dual mobility liner: a two-year MRI follow-up study.

Authors:  Maximilian F Kasparek; Lisa Renner; Martin Faschingbauer; Wenzel Waldstein; Kilian Rueckl; Friedrich Boettner
Journal:  Int Orthop       Date:  2017-09-19       Impact factor: 3.075

Review 2.  Diagnosis and management of implant debris-associated inflammation.

Authors:  Stuart B Goodman; Jiri Gallo; Emmanuel Gibon; Michiaki Takagi
Journal:  Expert Rev Med Devices       Date:  2019-12-17       Impact factor: 3.166

Review 3.  Revision of a Monoblock Metal-on-Metal Cup Using a Dual Mobility Component: Is It a Reasonable Option?

Authors:  Saverio Affatato; Emanuela Castiello; Luca Amendola; Saverio Comitini; Jean Louis Prudhon; Domenico Tigani
Journal:  Materials (Basel)       Date:  2020-04-27       Impact factor: 3.623

4.  Haemorrhagic Pseudotumour Following Metal-on-Metal Hip Replacement.

Authors:  Joshua L Filer; James Berstock; Ynyr Hughes-Roberts; Julian Foote; Harvey Sandhu
Journal:  Cureus       Date:  2021-06-09

5.  A new strategy to reconstruct type III acetabular bone defect associated with inflammatory pseudotumor: combined medial and lateral acetabular bone grafting: A case report.

Authors:  Jing Wang; Dejiu Wang; Jiang Du; Zhen Lin
Journal:  Medicine (Baltimore)       Date:  2017-12       Impact factor: 1.817

  5 in total