| Literature DB >> 27226972 |
Jack K Tung1, Ken Berglund2, Claire-Anne Gutekunst2, Ute Hochgeschwender3, Robert E Gross1.
Abstract
The use of bioluminescent reporters in neuroscience research continues to grow at a rapid pace as their applications and unique advantages over conventional fluorescent reporters become more appreciated. Here, we describe practical methods and principles for detecting and imaging bioluminescence from live cells and animals. We systematically tested various components of our conventional fluorescence microscope to optimize it for long-term bioluminescence imaging. High-resolution bioluminescence images from live neurons were obtained with our microscope setup, which could be continuously captured for several hours with no signs of phototoxicity. Bioluminescence from the mouse brain was also imaged noninvasively through the intact skull with a conventional luminescence imager. These methods demonstrate how bioluminescence can be routinely detected and measured from live cells and animals in a cost-effective way with common reagents and equipment.Entities:
Keywords: bioluminescence; imaging; in vivo; live cell; luciferase
Year: 2016 PMID: 27226972 PMCID: PMC4874058 DOI: 10.1117/1.NPh.3.2.025001
Source DB: PubMed Journal: Neurophotonics ISSN: 2329-423X Impact factor: 3.593