Literature DB >> 27215098

High-Sensitivity Cardiac Troponin T: a Preanalytical Evaluation.

Moïse Michel, Fouzi Mestari, Rana Alkouri, Sylvie Dever, Catherine Devilliers, Françoise Imbert-Bismut, Dominique Bonnefont-Rousselot, Denis Monneret.   

Abstract

BACKGROUND: Little is known about the effects of preanalytical conditions on high-sensitivity cardiac troponin T (hs-cTnT) concentrations.
METHODS: Variations of hs-cTnT concentrations were evaluated under the following preanalytical conditions: 1) serum vs. lithium-heparin (Li-Hep) plasma, with or without separator gel; 2) centrifugation time (15-minutes vs. 10-minutes) and speed (1467 to 3756 g); 3) stability in Li-Hep plasma at room temperature and +4 degrees C for 4 days and at -80 degrees C for up to 12 months, for three concentrations; 4) four freeze-thaw cycles at -20 degrees C and -80 degrees C, for three concentrations.
RESULTS: No significant changes were found regarding the type of blood collection tube, the centrifugation, and storage conditions. Minor decreases were observed after four freeze-thaw cycles at -20 degrees C (< 6.5%) and -80 degrees C (< 3.4%).
CONCLUSIONS: High-sensitivity cardiac troponin T may be considered as not impacted by usual preanalytical conditions, thus strengthening its reliability in laboratory practice and clinical research.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27215098     DOI: 10.7754/clin.lab.2016.160123

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  3 in total

1.  STABILITY OF SERUM BONE-MINERAL, KIDNEY, AND CARDIAC BIOMARKERS AFTER A FREEZE-THAW CYCLE: THE ARIC STUDY.

Authors:  Junichi Ishigami; Amy B Karger; Pamela L Lutsey; Josef Coresh; Kunihiro Matsushita
Journal:  Am J Epidemiol       Date:  2022-02-19       Impact factor: 5.363

2.  Serial cardiac biomarkers for risk stratification of patients with COVID-19.

Authors:  Kwaku Tawiah; Laurel Jackson; Catherine Omosule; Claire Ballman; Bobby Shahideh; Mitchell G Scott; Gillian Murtagh; Christopher W Farnsworth
Journal:  Clin Biochem       Date:  2022-06-09       Impact factor: 3.625

3.  Measuring high-sensitivity cardiac troponin T blood concentration in population surveys.

Authors:  Antonio Ivan Lazzarino; Jennifer S Mindell
Journal:  PLoS One       Date:  2017-01-31       Impact factor: 3.240

  3 in total