Safety of hydrophilic guidewires used for side-branch protection during stenting and proximal optimization technique in coronary bifurcation lesions.

Arka Chatterjee1, Brigitta C Brott2, Robin Foley3, Oluseun Alli1, Mark Sasse1, Mustafa Ahmed1, Firas Al Solaiman1, Gautam Reddy1, Sameer Ather1, Massoud A Leesar4.   

Abstract

BACKGROUND AND PROPOSE: In coronary bifurcation lesions (CBL), hydrophilic guidewires used for side-branch (SB) protection can be withdrawn from underneath the stent easier than other wires. However, the safety of which has not been investigated. METHODS/MATERIALS: We performed scanning electron microscopic (SEM) examination of hydrophilic wires - the Whisper and Runthrough wires - used for SB protection during stenting and proximal optimization technique (POT) in 30 patients with CBL. The distal 15cm of the wire was examined every 1mm by SEM and 4500 segments were analyzed to investigate for wire fracture, polymer shearing (PS), and its correlations with post-stenting creatine kinase (CK)-MB release.
RESULTS: SEM examination showed no evidence for wire fracture. The total area of PS and the largest defect on the wire were significantly larger with the Whisper wire versus the Runthrough wire (0.15±0.04mm2 vs. 0.026±0.01mm2 and 0.04±0.05mm2 vs. 0.01±0.01mm2; P<0.05, respectively). The total length of PS and the longest defect on the wire were significantly longer with the Whisper wire vs. the Runthrough wire (12.1±14.5mm vs. 2.7±3.0mm and 2.9±4.2mm vs. 1.0±1.2mm; P<0.05, respectively), but there were weak correlations between the extents of PS with CK-MB release.
CONCLUSIONS: Hydrophilic guidewires may be safely used for SB protection during stenting and POT in CBLs. The extent of PS was significantly greater with the Whisper wire than with the Runthrough wire, but its correlation with post-stenting CK-MB release was weak.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Coronary bifurcation lesions; Hydrophilic guidewire; Polymer shearing; Proximal optimization

Mesh:

Substances:

Year:  2016        PMID: 27210866     DOI: 10.1016/j.carrev.2016.04.006

Source DB:  PubMed          Journal:  Cardiovasc Revasc Med        ISSN: 1878-0938


  2 in total

1.  Jailing polymer jacketed guide-wires during bifurcation coronary interventions is associated with procedural myocardial infarction.

Authors:  Arka Chatterjee; Jeremy S White; Taimoor Hashim; Massoud A Leesar
Journal:  World J Cardiol       Date:  2017-05-26

Review 2.  Fundamentals of percutaneous coronary bifurcation interventions.

Authors:  Tamer Kırat
Journal:  World J Cardiol       Date:  2022-03-26
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.