Inadvertent permanent ventricular pacing from the coronary vein: an electrocardiographic, roentgenographic, and echocardiographic assessment.

U R Shettigar1, R R Loungani, C A Smith.   

Abstract

The incidence of inadvertent permanent ventricular pacing from the coronary vein is not known. In a retrospective analysis of 69 patients in whom transvenous pacemakers were implanted between 1979 and 1986, 12 patients were discovered to have right bundle-branch block pattern to the paced complexes on electrocardiogram. In this group, three patients were considered to have inadvertent placement of pacing lead in the coronary vein by two-dimensional echocardiographic criteria. No complications were noted in follow-up of 2-79 months. Monitoring of surface electrocardiogram, frontal and lateral fluoroscopy, and pacing threshold and sensing parameters (during implantation of pacemaker) were found to offer no absolute protection against malplacement of the lead. Placement of the lead into the lung field via the main pulmonary artery and then withdrawing with eventual positioning into the right ventricular apex will avoid malposition into the coronary vein. Two-dimensional echocardiography is useful for the diagnosis of pacing lead malplacement and should be performed in any patient with right bundle-branch block pattern in the surface electrocardiogram following pacemaker implantation.

Entities:  

Mesh:

Year:  1989        PMID: 2721040     DOI: 10.1002/clc.4960120508

Source DB:  PubMed          Journal:  Clin Cardiol        ISSN: 0160-9289            Impact factor:   2.882


  8 in total

1.  Transvenous implantation of a ventricular pacing lead in a patient with an artificial tricuspid valve prosthesis.

Authors:  M Scheffer; B van Gelder
Journal:  Neth Heart J       Date:  2003-09       Impact factor: 2.380

2.  Inadvertent placement of pacemaker lead into the middle cardiac vein.

Authors:  S Topaloglu; F Bayraktar; S Okten; S Cetin; O Ozeke; S Cay; D Aras
Journal:  Herz       Date:  2014-07-19       Impact factor: 1.443

3.  Electrocardiography findings in right ventricular apical pacing.

Authors:  Xiayan Shen; Ching-Hui Sia; Kian-Keong Poh; Weiting Huang; Kah Leng Ho
Journal:  Singapore Med J       Date:  2020-10       Impact factor: 1.858

4.  Electrocardiographic follow-up of biventricular pacemakers.

Authors:  S Serge Barold; Bengt Herweg; Michael Giudici
Journal:  Ann Noninvasive Electrocardiol       Date:  2005-04       Impact factor: 1.468

Review 5.  From lysosome to proteasome: the power of yeast in the dissection of proteinase function in cellular regulation and waste disposal.

Authors:  D H Wolf
Journal:  Cell Mol Life Sci       Date:  2004-07       Impact factor: 9.261

6.  Inadvertent malposition of a permanent ventricular lead into the middle cardiac vein was misdiagnosed as lead perforation.

Authors:  Mingxian Chen; Zhihong Wu; Zhenjiang Liu; Lin Hu; Xuping Li; Qiming Liu; Shenghua Zhou
Journal:  Ann Noninvasive Electrocardiol       Date:  2022-04-23       Impact factor: 1.485

7.  Sinus venosus atrial septal defect: a rare cause of misplacement of pacemaker leads.

Authors:  Malick Bodian; Fatou Aw; Mouhamadou Ndiaye Bamba; Adama Kane; Modou Jobe; Alioune Tabane; Alassane Mbaye; Simon Antoine Sarr; Maboury Diao; Moustapha Sarr; Serigne Abdou Bâ
Journal:  Int Med Case Rep J       Date:  2013-07-05

Review 8.  Complications of electrotherapy - the dark side of treatment with cardiac implantable electronic devices.

Authors:  Szymon Domagała; Michał Domagała; Jakub Chyła; Celina Wojciechowska; Marianna Janion; Anna Polewczyk
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-03-22       Impact factor: 1.426

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.