A Sankiewicz1, L Romanowicz2, P Laudanski3, B Zelazowska-Rutkowska4, B Puzan1, B Cylwik4, E Gorodkiewicz5.
Abstract
A new method for the selective determination of laminin-5 concentration using a biosensor and surface plasmon resonance imaging (SPRI) technique is presented. A biosensor based on the specific interaction of laminin-5 with rabbit polyclonal antibody was constructed. The analytically useful dynamic response range of the biosensor is between 0.014 and 0.1 ng mL(-1). The detection limit is 4 pg mL(-1). The potential influence of interferences on the SPRI signal was investigated, and the high selectivity of the biosensor was confirmed. In order to demonstrate the potential application of the biosensor, laminin-5 concentration in blood plasma was determined. The results were compared with the laminin-5 concentration obtained by the commercial enzyme-linked immunosorbent assay (ELISA) kit. A comparison of results from healthy donors obtained by SPRI measurement and ELISA indicates that they are close and shows good agreement with the data reported in the literature. The plasma samples of bladder cancer patients gave higher concentration measured with specific biosensor than by ELISA assay. The study shows the clear difference in concentration of laminin-5 in healthy humans and patients with bladder cancer. Extensive clinical studies using the newly developed method can result in an increase in the use of laminin-5 as a potential cancer marker.Entities:
Keywords: Biosensors; Cancer marker; Laminin-5; Surface plasmon resonance imaging
Mesh:
Substances:
Year: 2016 PMID: 27209594 PMCID: PMC4925691 DOI: 10.1007/s00216-016-9621-x
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142
Fig. 1a Picture of chip (A photopolymer, B free gold surface, C hydrophobic paint). b The SPR image of the chip obtained using a CCD camera. c The schematic illustration of the sensor active part
Fig. 2Dependence of SPRI signal (arbitrary units) of antibody-laminin-5 complex on antibody concentration. The laminin-5 concentration is 1 ng mL−1. The pH of laminin-5 solution was 7.4. Error bars are calculated for 12 independent measurements for each concentration at a 95 % confidence level
Fig. 3Dependence of SPRI signal (arbitrary units) of antibody-laminin-5 complex on pH. The antibody concentration is 5.0 μg mL−1. The laminin-5 concentration is 1.0 ng mL−1. Error bars are calculated for 12 independent measurements for each antibody solution pH point at a 95 % confidence level
Fig. 4Dependence of the SPRI signal (arbitrary units) on laminin-5 concentration. Linear ranges are on a separate sub-figure. The antibody concentration is 5.0 ng mL−1. The pH of laminin-5 solution is 7.4. Error bars are calculated for 12 independent measurements for each concentration at a 95 % confidence level
Influence of human albumin, glycoprotein IIb/IIIa, fibronectin, and collagen on the determination of laminin-5 concentration by an SPRI method
| Protein |
| Added | Found | Recovery (%) |
|---|---|---|---|---|
| Albumin | 1:1 | 0.050 | 0.050 ± 0.003 | 100 |
| 1:100 | 0.052 ± 0.004 | 104 | ||
| 1:1000 | 0.054 ± 0.003 | 108 | ||
| Glycoprotein GPIIBb/IIIa | 1:1 | 0.050 | 0.050 ± 0.005 | 100 |
| 1:100 | 0.054 ± 0.002 | 108 | ||
| 1:1000 | 0.055 ± 0.004 | 110 | ||
| Fibronectin | 1:1 | 0.050 | 0.050 ± 0.002 | 100 |
| 1:100 | 0.054 ± 0.007 | 108 | ||
| 1:1000 | 0.050 ± 0.002 | 100 | ||
| Collagen type IV | 1:1 | 0.050 | 0.051 ± 0.002 | 102 |
| 1:100 | 0.055 ± 0.004 | 110 | ||
| 1:1000 | 0.055 ± 0.004 | 110 |
Precision of measurement of laminin-5 concentration
| No. of measurements | Added | Found | Recovery (%) | SD | Confidence limit (95 %) (ng mL−1) |
|---|---|---|---|---|---|
| 36 | 0.100 | 0.106 | 106 | 0.016 | 0.005 |
| 0.050 | 0.052 | 104 | 0.012 | 0.004 | |
| 0.014 | 0.014 | 100 | 0.002 | 0.001 |
Recoveries of laminin-5 in the plasma
| No. of samples | Laminin-5 concentration without spike (ng mL−1) | Added standard concentration (ng mL−1) | Determined (ng mL−1) | Recovery (%) |
|---|---|---|---|---|
| 1 | 46.7 ± 1.2 | 50.0 | 94.8 ± 1.8 | 98 |
| 2 | 51.1 ± 2.0 | 50.0 | 107.2 ± 2.2 | 106 |
| 3 | 51.9 ± 1.5 | 50.0 | 101.9 ± 2.1 | 100 |
| 4 | 52.9 ± 1.9 | 50.0 | 101.9 ± 1.5 | 99 |
| 5 | 62.6 ± 1.1 | 50.0 | 110.3 ± 1.7 | 98 |
| 6 | 60.2 ± 1.7 | 50.0 | 111.3 ± 1.7 | 101 |
| 7 | 53.3 ± 1.5 | 50.0 | 105.4 ± 2.1 | 102 |
| 8 | 57.7 ± 1.5 | 50.0 | 113.1 ± 1.4 | 105 |
| 9 | 37.1 ± 2.1 | 50.0 | 91.5 ± 1.2 | 105 |
Concentration of laminin-5 in the human blood plasma measured by SPRI method and ELISA
| Laminin-5 concentration (ng mL−1) | ||||
|---|---|---|---|---|
| Healthy donors | Patients with bladder cancer | |||
| SPRI biosensor | ELISA | SPRI biosensor | ELISA | |
| 1 | 53.0 ± 1.5 | 57.9 ± 1.8 | 173.0 ± 2.5 | 138.0 ± 1.1 |
| 2 | 59.6 ± 1.1 | 61.1 ± 2.0 | 172.5 ± 3.1 | 109.8 ± 2.4 |
| 3 | 63.6 ± 0.9 | 63.7 ± 1.6 | 135.7 ± 1.6 | 121.2 ± 2.0 |
| 4 | 62.9 ± 0.7 | 15.5 ± 1.7 | 146.7 ± 2.8 | 109.1 ± 2.1 |
| 5 | 37.1 ± 1.2 | 40.9 ± 0.8 | 209.9 ± 1.9 | 151.7 ± 1.3 |
| 6 | 53.3 ± 0.8 | 52.2 ± 1.2 | 125.6 ± 1.9 | 91.0 ± 1.5 |
| 7 | 67.3 ± 0.9 | 65.8 ± 1.2 | 72.4 ± 2.7 | 63.5 ± 2.0 |
| 8 | 60.2 ± 0.6 | 61.1 ± 2.1 | 129.0 ± 3.0 | 112.6 ± 1.9 |
| 9 | 46.7 ± 1.5 | 44.4 ± 1.5 | ||
| Mean value ± confidence limit | 56.0 ± 6.2 | 51.4 ± 10.4 | 145.6 ± 26.5 | 112.1 ± 17.6 |
| Range | 37.1–67.3 | 15.5–65.8 | 72.4–209.9 | 63.5–138.0 |
| Literature range | 46.2 ± 10.2 [ | |||