Same-Level Fracture of the Tibial Metal Tray and Polyethylene Insert After Total Knee Arthroplasty.

Jong Yeal Kang, Yong Seuk Lee.   

Abstract

The authors report a case of failure fracture of the tibial metal tray and polyethylene insert at the same level in a 73-year-old woman 10 years after total knee arthroplasty using the AMK Total Knee System (DePuy, Warsaw, Indiana). Causes of this fracture are analyzed and discussed, with the focus on the importance of component design, position, and size. The overall aim of this case report is for orthopedic surgeons to avoid this complication in total knee arthroplasty by paying attention to these controllable factors. [Orthopedics. 2016; 39(4):e787-e789.]. Copyright 2016, SLACK Incorporated.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27203415     DOI: 10.3928/01477447-20160513-03

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  1 in total

1.  Tibial tray fracture in a modern prosthesis with retrieval analysis.

Authors:  John J Callaghan; David E DeMik; Nicholas A Bedard; Andrew N Odland; William M Kane; Steven M Kurtz
Journal:  Arthroplast Today       Date:  2018-02-10
  1 in total