Adriana Calderaro1, Maddalena Piergianni1, Giovanna Piccolo1, Sabina Rossi1, Sara Montecchini1, Mirko Buttrini1, Maria Cristina Arcangeletti1, Maria Cristina Medici1, Carlo Chezzi1, Flora De Conto1.
Abstract
This study evaluated the immunochromatographic (IC) assay "TECHLAB(®) E. HISTOLYTICA QUIK CHEK™" analysing 36 faecal samples and 7 cultured strains. This assay was compared to the methods performed in our laboratory for the diagnosis of amoebiasis. The IC assay revealed a detection limit of 103 trophozoites/g faeces and no cross-reactivity with other parasites and failed to detect E. histolytica antigen in frozen faeces. In our laboratory located in a non-endemic setting this assay could not replace the methods currently used for the diagnosis of amoebiasis.Entities:
Keywords: Antigen detection; Diagnosis; Entamoeba histolytica; Real-time PCR
Mesh:
Substances:
Year: 2016 PMID: 27196557
Source DB: PubMed Journal: New Microbiol ISSN: 1121-7138 Impact factor: 2.479