Abstract
A maintenance program generated through the consideration of characteristics and failures of medical equipment is an important component of technology management. However, older technology devices and newer high-tech devices cannot be efficiently managed using the same strategies because of their different characteristics. This study aimed to generate a maintenance program comprising two different strategies to increase the efficiency of device management: preventive maintenance for older technology devices and predictive maintenance for newer high-tech devices. For preventive maintenance development, 589 older technology devices were subjected to performance verification and safety testing (PVST). For predictive maintenance development, the manufacturers' recommendations were used for 134 high-tech devices. These strategies were evaluated in terms of device reliability. This study recommends the use of two different maintenance strategies for old and new devices at hospitals in developing countries. Thus, older technology devices that applied only corrective maintenance will be included in maintenance like high-tech devices.Entities:
Mesh:
Year: 2016 PMID: 27195666 PMCID: PMC5058563 DOI: 10.1155/2016/7267983
Source DB: PubMed Journal: J Healthc Eng ISSN: 2040-2295 Impact factor: 2.682
Figure 1Medical devices in the old technology and new high-tech device groups.
Scores used to calculate the Equipment Management Number.
| Score | Function | Risk | Maintenance requirement |
|---|---|---|---|
| 10 | Life recovery | — | — |
| 9 | Surgical and intensive care | — | — |
| 8 | Physical therapy | — | — |
| 7 | Surgical and intensive care | — | — |
| 6 | Other physiological monitors | — | — |
| 5 | Analytical laboratory | Patient death | Very important |
| 4 | Laboratory equipment | Patient-staff injury | Moderately important |
| 3 | Computers | Wrong diagnosis | Less important |
| 2 | Belong to the patients | Treatment delays | The least important |
| 1 | Other equipment pieces | Risk not important | Minimally important |
PVST parameters and PVST intervals of medical devices (FP: function point, RP: risk point, MR: maintenance requirement, and EMN: equipment management number).
| Medical | PVST | Simulator | FP | RP | MR | EMN | Test interval |
|---|---|---|---|---|---|---|---|
| Electrocardiography | Linearity | Patient | 7 | 3 | 4 | 14 | 12 months |
|
| |||||||
| Pulse | ECG BPM test | SPO2 | 7 | 3 | 4 | 14 | 12 months |
|
| |||||||
| Sphygmomanometer | Pressure leakage | NIBP Simulator (Fluke | 6 | 3 | 4 | 13 | 12 months |
|
| |||||||
| Infant | Temperature test | Patient | 10 | 5 | 5 | 20 | 6 months |
|
| |||||||
| Phototherapy | Intensity | Phototherapy | 8 | 4 | 5 | 17 | 6 months |
|
| |||||||
| Defibrillator | ECG BPM test | Defibrillator Analyzer | 10 | 5 | 5 | 20 | 6 months |
|
| |||||||
| Aspirator | Max. free flow | Flow Analyzer | 9 | 3 | 4 | 16 | 12 months |
|
| |||||||
| Electrosurgical | Cutting-power test | Electrosurgical | 9 | 5 | 5 | 19 | 6 months |
Data list screen (8 records; total: 589 records).
| Location | Device name | Brand code | Serial number | Status | Error |
|---|---|---|---|---|---|
| ECG room | ECG | Nihon Kohden | 4971 | Passed | No problem |
| Service | SPO2 | Massimo | N43378 | Passed | No problem |
| Emergency | Sphygmomanometer | Riester | 091250123 | Failed | High pressure leakage |
| Infant intensive care unit | Infant incubator | Fanem | CI1649 | Failed | High temperature |
| Infant intensive care unit | Phototherapy unit | Medix | 560-09 | Passed | No problem |
| Emergency | Defibrillator | Nihon Kohden | 07728 | Failed | Low battery |
| Operation room | Aspirator | Bıçakcılar | 1598 | Failed | Low vacuum |
| Operation | Electrosurgical | Martin | BO 88 | Failed | Power circuit error |
In total, the data of 589 medical devices were listed.
Predictive maintenance time schedule including newer high-tech devices.
| Predictive maintenance time schedule | ||||||||
|---|---|---|---|---|---|---|---|---|
| Device | Brand | Model | Daily | Every | Every | Every | Every | Every |
| 1 m | 3 m | 4 m | 6 m | 12 m | ||||
| CT | Philips | Brilliance CT 16-Slice | × | × | ||||
| Siemens | Somatom Sensation 4 | × | × | |||||
| Toshiba | Aquilion 64 | × | × | |||||
|
| ||||||||
| Angio | Philips | MultiDiagnost Eleva | × | × | ||||
| Siemens | Axiom Artis dtA | × | × | |||||
|
| ||||||||
| Mammo | IMS | Giotto SDL | × | × | ||||
|
| ||||||||
| C-arm | Siemens | Arcadis Varic | × | × | ||||
| Siemens | Arcadis Varic Gen 2 | × | × | |||||
| Siemens | Siremobil Compact | × | × | |||||
| Siemens | Siremobil Compact L | × | × | |||||
| Philips | BV Endura | × | × | |||||
|
| ||||||||
| MR | Siemens | Magnetom Symphony | × | × | ||||
| Philips | Achieva 1,5 T | × | × | |||||
| Philips | Achieva 3,0 T | × | × | |||||
|
| ||||||||
| Pet-CT | Siemens | Biograph 6 TruePoint | × | × | ||||
|
| ||||||||
| Gamma | Siemens | E-Cam Extended Gantry | × | × | × | |||
| Camera | Mediso | Nucline DHV-2 Sprit | × | × | × | × | × | |
| Mediso | Nucline TH-22 | × | × | × | × | × | ||
|
| ||||||||
| Ventilator | Maquet | Servo-s | × | × | ||||
| Maquet | Servo-i | × | × | |||||
| Draeger | Babylog 8000 plus | × | × | |||||
| Draeger | Evita 4 Neoflow | × | × | |||||
| GE | Engström | × | × | |||||
|
| ||||||||
| Anesthesia unit | Draeger | Fabius | × | × | ||||
| Draeger | Fabius GC | × | × | |||||
| Draeger | Julian | × | × | |||||
| Draeger | Primus | × | × | |||||
| GE Datex | Avannce S5 | × | × | |||||
| GE Datex | Aestiva 5 | × | × | |||||
Istanbul University Hospitals.
Predictive Maintenance Program 2014.
Medical device failures.
| Medical device | Total # | Number | Errors | Error code |
|---|---|---|---|---|
| Electrocardiogram | 50 | 2 | Not working | ECG01 |
| 1 | Power circuit error | ECG02 | ||
| 1 | Paper speed error | ECG03 | ||
| 3 | Electrode error | ECG04 | ||
| 1 | Sensitivity error | ECG05 | ||
|
| ||||
| Pulse oximeter | 46 | 3 | Not working | SPO01 |
| 1 | Low BPM | SPO02 | ||
| 1 | High BPM | SPO03 | ||
| 2 | Low oxygen saturation | SPO04 | ||
| 2 | High oxygen saturation | SPO05 | ||
| 5 | Probe error | SPO06 | ||
|
| ||||
| Sphygmomanometer | 200 | 12 | Not working | SPG01 |
| 15 | High pressure leakage | SPG02 | ||
| 8 | Cuff error | SPG03 | ||
| 4 | Broken manometer | SPG04 | ||
| 5 | Missing piece | SPG05 | ||
|
| ||||
| Infant incubator | 28 | 3 | Not working | INC01 |
| 2 | Over temperature | INC02 | ||
| 1 | Display error | INC03 | ||
| 1 | Baby probe error | INC04 | ||
| 2 | Broken cover | INC05 | ||
|
| ||||
| Phototherapy unit | 30 | 3 | Not working | PHT01 |
| 2 | Low intensity | PHT02 | ||
|
| ||||
| Defibrillator | 86 | 3 | Not working | DEF01 |
| 1 | Low/high energy | DEF02 | ||
| 3 | Low battery | DEF03 | ||
| 2 | Electrode error | DEF04 | ||
| 1 | Paddle error | DEF05 | ||
| 2 | BPM error | DEF06 | ||
| 2 | Synchronization error | DEF07 | ||
|
| ||||
| Aspirator | 97 | 6 | Not working | ASP01 |
| 4 | High vacuum | ASP02 | ||
| 3 | Low vacuum | ASP03 | ||
| 3 | Maximum vacuum | ASP04 | ||
| 2 | Vacuum rise error | ASP05 | ||
|
| ||||
| Electrosurgical unit | 52 | 1 | Not working | ESU01 |
| 3 | Power circuit error | ESU02 | ||
| 3 | High/low cut power | ESU03 | ||
| 2 | High/low coag. power | ESU04 | ||
| 1 | High/low bipolar power | ESU05 | ||
| 2 | Foot switching error | ESU06 | ||
| 2 | Patient electrode error | ESU07 | ||
Figure 2Failures specific for each type of older technology medical device.
Preventive maintenance time schedule including older technology devices.
| Preventive maintenance time schedule | |||||||
|---|---|---|---|---|---|---|---|
| Device | Daily | Weekly | Every | Every | Every 12 m | As needed | |
| 3 m | 6 m | ||||||
| Infant incubator | Baby temperature probe | × | |||||
| Air filter | × | ||||||
| Electrical fuse | × | ||||||
| Switch | × | ||||||
| Breaker relay | × | ||||||
| Fan motor | × | ||||||
| Vacuum compressor motor | × | ||||||
| Pressure sensor | × | ||||||
| Noise sensor | × | ||||||
| Air circulation sensor | × | ||||||
| Humidity sensor | × | ||||||
| Temperature sensor | × | ||||||
| Accumulator | × | ||||||
| Gasket | × | ||||||
|
| |||||||
| Aspirator | Air input filter | × | |||||
| Pump motor | × | ||||||
| Fan motor | × | ||||||
| Air hose | × | ||||||
| Fluid suction hose | × | ||||||
|
| |||||||
| Defibrillator | Battery | × | |||||
| Spoon connection cable | × | ||||||
| Fibrillation detection sensor | × | ||||||
| Charging transformer | × | ||||||
| Heart beat sensor | × | ||||||
| ECG sensor | × | ||||||
| Leakage relay | × | ||||||
| Electrical fuse | × | ||||||
| Defibrillation time sensor | × | ||||||
|
| |||||||
| Pulse oximeter | Probes | × | |||||
| Optic sensor | × | ||||||
| Battery | × | ||||||
| Connector | × | ||||||
Istanbul University Hospitals
Preventive Maintenance Program 2014.
Figure 3(a) Failures of the defibrillators and (b) of the electrosurgical units before and after preventive maintenance.
Figure 4Failures of the gamma camera before and after predictive maintenance. (a) Failures reported during daily checks. (b) Failures reported during work.
Figure 5Failures of the ventilator before and after predictive maintenance. (a) Failures reported during daily checks. (b) Failures reported during work.