Literature DB >> 27187983

Safety and Effectiveness of the Nav-6 Filter in Preventing Distal Embolization During Jetstream Atherectomy of Infrainguinal Peripheral Artery Lesions.

Avantika Banerjee, Karan Sarode, Atif Mohammad, Emmanouil S Brilakis, Subhash Banerjee, Gail A Shammas, Nicolas W Shammas1.   

Abstract

BACKGROUND: The risk of distal embolization (DE) during infrainguinal peripheral artery interventions (PAI) is often mitigated by the use of embolic protection devices. There are limited data on the use of filters with the Jetstream (JS) atherectomy device, a rotational cutter with aspiration capacity. The Nav-6 filter is uniquely suited for use with the JS due to its wire compatibility and detachment from the filter; however, data on the off-label use of this combination have not been reported.
METHODS: Consecutive patients between October 2008 and April 2015 undergoing endovascular infrainguinal PAI with JS were analyzed as part of the Excellence in Peripheral Artery Disease (XL-PAD) registry (NCT01904851). Patients were divided into two subgroups with Nav-6 filter use vs no filter use. Descriptive and univariate analyses were performed.
RESULTS: Among 141 patients (mean age, 67.8 ± 10.8 years; 169 lesions) included in this study, the Nav-6 filter was used in 82 (59%). Use of a filter was more frequent in longer lesions (146 ± 106 mm vs 91 ± 72 mm; P=.01), in more severe stenoses (95% vs 87%; P=.04), and in chronic total occlusions (33% vs 8.3%; P=.01). Patients receiving filters had longer procedure duration (102 ± 51 min vs 66 ± 41 min; P=.01) and longer fluoroscopy times (31 ± 16 min vs 21 ± 10 min; P<.001). Use of the Nav-6 filter with the JS during PAI was associated with numerically lower rates of DE (1.8% vs 8%; P=.10) and similar rates of death and amputation. At 12 months, the target-lesion revascularization rate was higher in the filter group (22% vs 2.7%; P=.02), likely secondary to use of the filter in more complex lesions.
CONCLUSION: Nav-6 filter during JS atherectomy was predominantly used during complex infrainguinal PAI and was associated with less occurrence of DE.

Entities:  

Mesh:

Year:  2016        PMID: 27187983

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  4 in total

1.  Embolic Protection using the WIRION Embolic Protection System with JetStream Atherectomy in a Patient with Superficial Femoral Artery Chronic total Occlusion.

Authors:  Nicolas W Shammas
Journal:  Int J Angiol       Date:  2016-06-28

Review 2.  Review of the Latest Percutaneous Devices in Critical Limb Ischemia.

Authors:  Leila Haghighat; Sophia Elissa Altin; Robert R Attaran; Carlos Mena-Hurtado; Christopher J Regan
Journal:  J Clin Med       Date:  2018-04-14       Impact factor: 4.241

Review 3.  [Atherectomy in Peripheral Artery Disease: Current and Future].

Authors:  Yohan Kwon; Jinoo Kim; Je-Hwan Won; Seong Ho Kim; Jeong-Eun Kim; Sung-Joon Park
Journal:  Taehan Yongsang Uihakhoe Chi       Date:  2021-05-20

4.  Comparison and Analysis between the NAV6 Embolic Protection Filter and SpiderFX EPD Filter in Superficial Femoral Artery Lesions.

Authors:  Prakash Krishnan; Arthur Tarricone; Allen Gee; Serdar Farhan; Haroon Kamran; Annapoorna Kini; Samin Sharma
Journal:  J Interv Cardiol       Date:  2021-06-01       Impact factor: 2.279